9.5 Message list
Data and operating errors
(code 5000 to 8999)
5003 Cannot exit service mode
5004 Command or data transmission only
available in service mode
5006 Command or data transmission not possi-
ble because of BUSY
5007 Command or data transmission not possi-
ble because module is faulty or SIMATIC CPU
stop
5102 Command not possible because there is
no standstill
5103 Command not possible because legal
trade display is missing
5104 Command not possible because range is
exceeded
5105 Load cell parameters not plausible
5107 Characteristic cannot be moved
5108 ID does not exist
5199 Error in command to DI
6002 Logging not possible because weight is
too small
7000 Permitted number range violated
7001 Regulation code not known
7002 Specifications of string lengths not plausi-
ble
7003 Specification of date / time not plausible
7004 Assignment of DIs/DOs incorrect
7006 Reserved
7007 The calibration weights or calibration digits
are not plausible
7008 Set to zero or tare parameter not plausible 7008
7009 Standstill range / standstill wait time
7010 Scale interval / rounding
7011 Filter parameters
7013 Interface assignment for calibratable HMI
not plausible
7014 Specified time not plausible
7016 Parameter assignment of analog output
not plausible
7017 MAC address cannot be changed
7018 Error in IP mask
7019 RS485 parameter error
136
Error code
Description and remedy
5003
Cannot exit calibration mode; calibration incomplete
5004
Activation of service mode is required to execute command or trans-
mit data
5006
Command can currently not be executed because module is BUSY
(data record or command transmission already active, ...)
5007
Command can currently not be executed because of a problem or
SIMATIC CPU stop
5102
Scale command (set to zero, tare, log, ... ) cannot be executed be-
cause standstill is missing.
5103
Command not possible because legal trade display is missing or not
visible.
5104
Command (e.g. set to zero, tare, calibrate command) cannot be
executed because the permitted range has been exceeded. The
ranges are defined in DR 3.
5105
Load cell parameters in data record DR 10 are not plausible (number,
support points, load specifications, etc.).
5107
Characteristic cannot be moved due to possibility of range violation.
5108
Requested logbook ID not present in memory.
5199
Processing of a command triggered at the DIs is not possible. The
cause can be determined in data record DR 32.
6002
Logging is not possible because the limits for the minimum weight or
maximum weight were not observed.
7000
The permitted number range, such as for weight values, was violat-
ed.
7001
Regulation code for application requiring official calibration not
known.
7002
The string header in a specified string variable is not plausible.
7003
Specifications for date and time are not plausible.
7004
An error occurred while assigning the digital inputs or digital outputs.
7006
Reserved
7007
Specifications for calibration weights or digits in data record DR 3 are
incorrect (minimum distance, reversal of incline).
The specifications for set to zero (data record DR 3) or tare specifica-
tions (data record DR 15) are not plausible.
7009
Standstill range or standstill wait time are not plausible.
7010
Scale interval or selection for rounding to decimal places is not plau-
sible.
7011
Specification of filter parameters is not plausible.
7013
The assignment of the interface to the calibratable HMI is incorrect.
7014
Specified time value is not plausible or may be signaled in connection
with additional errors
7016
The parameters for the analog output (data record DR 7) are not
plausible.
7017
7018
The specified IP addresses (DR12) are not plausible.
7019
The specified RS485 interface parameters (DR13) are not plausible.
SIWAREX WP231
Manual, 07/2014, A5E31238908A-02