M3500B HeartStart XLT Defibrillator/Monitor
Events Recorded
The following events and related information are stored in the Event
Summary:
Table 9-1 Event Information
Event
Power Change
Pads Change
AED Mode Analysis
Mode Change
Rhythm Monitoring
Charging
Shock
Shock Failed
Disarm
ECG Monitoring
Heart Rate Alarm Violation
Related Information Stored
Power on, Power off, Continued use, Battery
low.
Pads on, Pads off.
Analyzing, Analysis Stopped, Artifact
Detected, Cannot Analyze, Shock Advised,
No Shock Advised.
AED Mode or Manual Mode.
Check Patient, Pause, Resume.
ECG waveform, Energy charged to.
ECG waveform, Shock#, Delivered energy,
Peak current, and Patient impedance.
No Shock Delivered.
ECG waveform.
Leads on, Leads off, Lead change, Gain
change.
Lead, Heart Rate, and Heart Rate alarm limits.
Events Recorded
9
9-3