Thermal Array Printer (Cont.); Battery And Battery Power Module - Philips HEARTSTART XLT M3500B Instructions For Use Manual

Defibrillator/monitor
Table of Contents

Advertisement

Specifications
13-6

Thermal Array Printer (cont.)

Auto Printing: The printer can be configured to automatically print on Mark,
Charge, Shock and Alarm.
Delayed Printing: The printer can be configured to run real time or with a six
second delay.
Reports: The following can be printed: Event Summary, Configuration,
Extended Self Test, System Log.
Speed: 25 mm/s with an accuracy of
Amplitude Accuracy:
Paper Size: 50 mm by 30 m (100 ft).

Battery and Battery Power Module

Battery Type: 2 Ah, 12V, rechargeable, Sealed Lead Acid.
Battery Dimensions:
Weight: 1.4 lb (0.65 kg).
Charge Time:
Approximately 14.5 hours to 100%, indicated by LED on power module.
Approximately 2.5 hours to 90%, indicated by LED on power module.
Capacity: 2.7 hours ECG monitoring or 50 full-energy discharges or 2.1
hours ECG monitoring while pacing (with a new, fully charged battery and
temperature above 20
Battery Indicators: LOW BATTERY message appears on monitor when at
least 10 minutes of monitoring time and 6 maximum-energy discharges
remain (with a new battery and temperature above 20
Battery Storage: Should not be stored above 40
time.
Charger Output: Unit can be operated using Class I AC power module alone
with no battery installed.
+ _
5%.
+ _
+ _
10% or
50 uV, whichever is greater.
2.4" (H) x 0.94" (W) x 7.2" (D).
61.7 mm (H) x 23.9 mm(W) x 182 mm (D).
o
C).
o
C).
o
C for extended periods of