Philips HEARTSTART XLT M3500B Instructions For Use Manual page 77

Defibrillator/monitor
Table of Contents

Advertisement

WARNING
M3500B HeartStart XLT Defibrillator/Monitor
Figure 6-4 Charging Display
LEAD
HR ALARM
SELECT
78
Pads
150 J
ENERGY
DISARM
SELECT
If desired, you may increase or decrease the selected energy level after
pressing the
button. To do so, press
desired energy level is displayed. The defibrillator charges to the selected
energy automatically. Wait until the current charge reaches the selected
energy level before proceeding.
3. Shock
Confirm that a shock is still indicated. Make sure no one is touching the
patient or anything connected to the patient. Call out "Clear." Then press
to deliver a shock to the patient.
To disarm the defibrillator, press
onds, the defibrillator disarms automatically.
If additional shocks are indicated, repeat the defibrillation process
Defibrillation current can cause operator or bystander injury. Do not touch the
patient, or equipment connected to the patient, during defibrillation.
SpO2
87 J
Current Charge
Shocks:0
00:00:8
repeatedly until the
. If
is not pressed in 30 sec-
Defibrillating
6
6-5