How to Use
To stop operation
See page 22
To check the settings during operation
Press the
The settings are indicated when this button is
pressed.

10. Operation is completed.

• Liquid sterilization cycle or equipment sterilization cycle
The buzzer sounds and the completion of cycle is indicated on the
digital display II.
⋅ In the liquid sterilization cycle, the completion of cycle is indicated
at 64°C
⋅ In the equipment sterilization cycle, the completion of cycle is indicated
at 74°C
• Sterilization/Heat-retention cycle or melting/heat-retention cycle
When the heat-retention cycle starts, the buzzer sounds and the bar
sign of "Heat-retention" of the process indicator lamps flickers. The
heat-retention elapsed time is indicated on the digital display II.
After the heat-retention has been continued for 72 hours, the
operation is completed, the buzzer sounds and the completion of
cycle is indicated on the digital display II.

11. Open the cover.

1) Be sure to check the following points before opening the cover.
Forcible opening of the door may cause trouble.
⋅ Pressure is lowered to 0MPa.
⋅ COVER LOCK lamp is lit.
When the COVER LOCK lamp is lit, the temperature or pressure
inside the chamber rises to make it impossible to pull the movable
handle toward you.
2) While bringing down the handle,
3) Pull the movable handle toward you to unlock the cover.
4) Pull up the handle to slightly open the cover and wait until no
steam comes out.
Since high-temperature steam will escape when the cover is
opened, wait until the steam dissipates. During this time, keep
your head away from the product.
5) Open the cover slowly.
Sudden opening of the door may break the container of
sterilization objects and cause the culture medium to spill over.
21
button.
CAUTION
To change the settings during operation
The settings of sterilization time, melting time, heat-
retention temperature, exhaust temperature, exhaust
rate, exhaust fan operation start/stop can be changed.
However, the changes will not be saved in memory.
When changing the settings, see page 18.
* The sterilization temperature can not be changed
during operation.
Indication of completion
1)
Handle
Movable
handle
4)
5)
Make sure that the
COVER LOCK lamp
goes out.
Make sure of
0
indication on the
pressure gauge.
3)
2)