9. Press the
button.
In the case when the operation is started after selection of liquid sterilization cycle,
sterilization/heat-retention cycle or equipment sterilization cycle
• When the sterilization setting time is longer than
the heat-retention time shown in the table below,
1) Press the
button.
The process starts.
The cooling fan is operated automatically after
the completion of sterilization.
If the cooling fan operation not needed, press the
cooling fan button to turn OFF.
• On sterilization setting time
The
sterilization
time
Pharmacopeia and the standards of JIS T 7324-2005
Small-sized High-pressure Steam Sterilizer for Medical
Use and it does not guarantee the sterilization. Be sure
to confirm the sterilization by sterilization indicator.
In the case when the operation is started after selection of melting/heat-retention cycle
1) Press the
button.
The melting setting temperature appears on the digital
display I and the sterilization setting time and cycle
appear alternately on the digital display II.
2) Press the
button.
The process starts.
• When the cover is opened after the completion of
process, the cycle/program is set automatically at
the 1-1 program of the liquid sterilization cycle.
When the temperature inside the chamber reaches
the melting setting temperature, the temperature
inside the chamber appears on the digital display I
and the sterilization remaining time or melting
remaining time appears on the digital display II.
Advisory note
If the cover is not closed completely,
message) appears on the digital display II. In this case,
bring down the handle again and press the
when the buzzer sounds.
• The COVER LOCK lamp lights up after starting.
When the lamp is lit, the movable handle is locked to
make it impossible to open the cover.
Beep
is
based
on
Japanese
Beep
Beep
(cautionary
button
• When the sterilization setting time is shorter
than the heat-retention time shown in the table
below,
1) Press the
button.
The sterilization setting temperature appears on
the digital display I and the sterilization setting
time and cycle appear alternately on the digital
display II.
Confirm the setting.
2) Press the
button.
The process starts.
Sterilization Setting
Liquid Sterilization
Temperature (°C)
Heat-retention Time
115
121
126
132
135
Temperature inside chamber
Sterilization remaining time or
melting remaining time
(after the temperature reaches the
sterilization temperature or melting
temperature)
Cover is not closed.
Beep
Beep
Equipment Sterilization
Heat-retention Time
37Min.
30Min.
27Min.
20Min.
22Min.
15Min.
12Min.
5Min.
10Min.
3Min.