Install The Exhaust Tank - Sanyo MLS-3751 Instruction Manual

High-pressure steam sterilizer
Hide thumbs Also See for MLS-3751:
Table of Contents

Advertisement

How to Use

1. Install the exhaust tank.

1) Take out the exhaust tank from the product and remove the
packings.
Advisory note
Do not pull out the exhaust tank by force from the product with the
exhaust hose being connected. Otherwise, trouble may be
caused.
2) Pour water up to the LOW (minimum) level mark, insert the packings
to the holes of the product and connect the exhaust hoses.
Steam may leak during operation to cause scald unless the
packings are inserted completely and the exhaust hoses are
connected completely.
Memo
The water in the exhaust tank increases in each sterilizing
operation. If it will increase up to the HIGH (maximum) level mark,
once discard the water and fill the tank with fresh water up to the
LOW level mark.
3) Make sure that the exhaust hose packings have been
inserted completely in the holes of the exhaust tank.
When the packings are removed from the holes of the exhaust
tank, insert them completely back in the holes of the exhaust tank.
Advisory note
Insert the packings up to the hilt while turning them clockwise and
counterclockwise. If they are turned in only one direction, the
exhaust hoses will be twisted to cause trouble.
4) Stow the exhaust tank inside the product.
Make sure that the exhaust hoses are not bent or twisted. If they
are so, air in the chamber is hard to escape and an error message
may appear.
15
CAUTION
CAUTION
HIGH level mark
LOW level mark
Exhaust hoses
Packing
Hole
Hose inside
the tank
Exhaust tank