ALARM ID
DESCRIPTION
Water not heated within time
AF:01
allowed.
The temperature difference
between the two probes, "TL1"
AF:02
and "TC", is more than 2°C (only
active on WD series)
Probe "TL1" (chamber
temperature) reads a higher
AF:04
temperature than the effective
one.
Working temperature probe "TL1"
AF:05
is generating an abnormal signal
(probe "open").
Drying temperature probe TA1 is
AF:07
generating an abnormal signal
(probe "open").
Control temperature probe TLC is
generating an abnormal signal
AF:10
(probe "open").
(only active on WD series)
AF:11
No cold water during filling.
AF:13
No demineralised water.
AF:17
Cold water filling time not correct.
Demi filling time too long. The
AF:19
appliance is taking too long to take
in demineralised water.
Not enough water. Chamber water
AF:23
level too low.
Anomaly related to the circulating
AF:25
pump "ML". Circulating pump
pressure too low.
User Manual
For the user: follow the DEFAULT PROCEDURE described above.
If the alarm does not disappear, follow the RESET PROCEDURE.
For the user: follow the DEFAULT PROCEDURE described above.
If the alarm does not disappear, follow the RESET PROCEDURE.
For the user: follow the DEFAULT PROCEDURE described above.
If the alarm does not disappear, follow the RESET PROCEDURE.
For the user: follow the DEFAULT PROCEDURE described above.
If the alarm does not disappear, follow the RESET PROCEDURE.
For the user: follow the DEFAULT PROCEDURE described above.
If the alarm does not disappear, follow the RESET PROCEDURE.
For the user: follow the DEFAULT PROCEDURE described above.
If the alarm does not disappear, follow the RESET PROCEDURE.
Check the water supply:
1. Check that the intake tap is open.
2. Check the water supply pressure.
3. Check that the filler hoses are correctly positioned.
Follow the DEFAULT PROCEDURE described above. If the alarm
does not disappear, follow the RESET PROCEDURE.
Check the water and demi water supply:
1. Check that the intake tap is open.
2. If using water from a tank (with PAD accessory), check
that the tank is not empty or placed too low down.
3. Check the water supply pressure.
4. Check that the display settings actually correspond to
the water connections (demi water present or not).
Follow the DEFAULT PROCEDURE described above. If the alarm
does not disappear, follow the RESET PROCEDURE.
Checks and procedures as for AF:11
Checks and procedures as for AF:13
Checks and procedures as for AF:11
Also check that there are no leaks from the appliance (no water
around it).
If there are leaks: turn off all water supply taps at once and
contact the After-Sales Service.
There may be foam in the chamber.
Check the type of detergent used.
For the user: follow the DEFAULT PROCEDURE described above.
If the alarm does not disappear, follow the RESET PROCEDURE.
GW1160 - GW4060 Series
USER ACTION
Page 53 - 75