4.4 Output Instructions

The CA0 can output DC current.
There are two output modes.
SOURCE mode:
SIMULATE mode (SINK):
There are two configuration modes.
Constant current output:
Current sweep output:
WARNING
Do not apply a voltage of 8 V or greater to the output terminals. Doing so may lead to
electric shock.
Also, keep the voltage between the circuit and the ground below 8 V.
Always use the attached lead cables and lead cables.
(Check that they are appropriate for the measurement category.)
CAUTION
In modes other than SIMULATE mode, do not apply voltages to the output terminals.
Mistakenly applying voltage to the output terminals could damage the internal circuitry.
If the
mark appears while you are using the current output function, replace the
batteries.
Current is supplied from the CA0.
T he CA0 sinks current from an external voltage
source.
The specified current is output continuously.
T he output current is increased and decreased
within the specified span.