A new HP M2477B 4.0 amp-hour battery will provide 75 360-joule
NOTE
charge-shock cycles or a minimum of 3 hours of monitoring time. Replace the
battery when it fails to provide the minimum3 hours of monitoring time or 10
minutes of Low Battery warning time (starting from a fully charged battery).
Properly dispose of depleted batteries according to local regulations. Do
WARNING
not disassemble, puncture or incinerate the disposed batteries. Be careful
not to short the battery terminals, because this can cause a fire hazard.

New Batteries

To assure optimal performance of a new battery, all newly purchased
batteries should be capacity tested in the M2480B charger prior to being put
into service. Accomplish this by placing the battery in the charger and
performing the "Battery Capacity Test" on page 13-15.

Battery Life

Recommended battery pack replacement:
• After two years of service.
• After failure of battery capacity test.
Replace the battery only with one of the battery packs listed in Table 1-2 on
page 1-24.
The operating time of the battery depends on how much the battery pack is
charged. Press the capacity gauge button on the battery pack to determine
the charge level.
CodeMaster 100
13-13