Quick Pacer Functionality Test - HP Codemaster 100 User Manual

Portable defibrillator/monitor
Table of Contents

Advertisement

Notify Service Personnel if the ECG strip does not print TEST 100J
NOTE
PASSED or if any of the shock button tests fail.
If you see the CHECK SETUP message on the screen, call service to replace
CAUTION
the backup battery.
10 Disconnect the test load (M1781A) from the electrode adapter cable.

Quick Pacer Functionality Test

1 Connect the electrode adapter cable to the defibrillator and the HP
2 Press Pacer On. (During the test only, ignore the LEADS OFF warning
3 Put the pacer into the Fixed Mode of operation by pressing Mode.
4 Adjust the current to 30 mA by pressing Output. Adjust the pacer rate to
5 Start the pacer by pressing Start/Stop on the pacer keypad, and start the
6 Verify that pacer pulses are shown on the recorder strip approximately
If the recorder is in delay mode it will take several seconds before pulses
NOTE
appear.
7 Turn the pacer off by pressing Pacer On and stop the recorder printing
M1781A test load to the adapter cable. Turn the defibrillator on by turning
the Energy Select switch to the Monitor On position.
message if it occurs.)
60 pulses per minute by pressing Rate.
recorder by pressing Record on the recorder keypad.
every 5 large boxes.
Allow the pacer to run for 10-12 seconds.
by pressing Record. Disconnect the test load from the pads adapter
cable.
CodeMaster 100
13-5