5.8 Replacement of bulbs

The Image eraser contains four warm-white compact fluorescent bulbs. When any one
of the bulbs fails, replace all four bulbs to provide uniform light intensity across the
screen. To order replacement bulbs, contact GE Healthcare Technical Support or your
local distributor (see the manufacturer part number on the bulb). See back cover for
contact information.
To replace the bulbs, follow the instruction below.
1
2
3
4
5
6
7
Typhoon Operating Instructions 28-9623-56 AA
Reconnect the power cord to a grounded mains outlet and turn on the Image
eraser.
NOTICE
Before you turn on the Imager eraser after you change the fuse, make
sure that the correct operating voltage appears in the fuse window.
Selecting the wrong voltage can severely damage the eraser.
NOTICE
If you are replacing the bulbs shipped with the Image eraser, you must
order replacement bulbs and fixtures from GE Healthcare Technical
Support. The bulbs shipped with the Image eraser are epoxied to the
fixture to avoid damage during shipment. The replacement fixture and
bulbs are not epoxied.
Turn off the Image eraser and disconnect the power cord.
Unscrew the two thumbscrews (Fig 5-3 A) that hold the clip support (Fig 5-3 B) and
diffuser screen (Fig 5-3 C) in position.
Lift off the clip support and diffuser screen.
Unscrew the bulb and fixture from each receptacle (Fig 5-3 D).
Screw a bulb and fixture into each receptacle.
Replace the front diffuser screen and clip support, and screw the two
thumbscrews.
Verify in the fuse window that the fuse voltage is correct.
Maintenance 5
Replacement of bulbs 5.8