Replace A Sensor; Review Logs - Honeywell BW Ultra User Manual

Portable five-gas detector with internal pump
Hide thumbs Also See for BW Ultra:
Table of Contents

Advertisement

4. Confirm that you want to apply Nitrogen to zero CO
5. When Apply calibration gas now is displayed, apply the gas and wait for a maximum of five
minutes. The detector first tests for a specific type of gas. When it detects enough of that
gas for sensor calibration, a check box will be displayed next to that gas. Calibration then
begins. Gas values will adjust on the screen during the calibration.
6. When Turn gas off is displayed, disconnect the device from gas. Check marks will be
displayed next to the calibrated sensors. These sensors reset to the number of days until
the next calibration is due (for example, 180 days). The calibration cycle will take about two
minutes after which the user will be prompted to Press button to continue.
7. If the calibration was successful, Calibration Passed will be displayed. Press the button to
exit calibration. If the calibration failed for some or all of the gases, either a Cal Error All
gases applied mixed results message (if the detector was not successfully
NOTE: After a Chlorine (Cl
device, a manual calibration for Cl
in the gas detector. Further calibration cycles of Cl
sticky gases CL
calibration is available. However, it requires a wait period of 15 minutes between calibrations
and all the nests used in gang should be for BW Ultra. If it is not the case, it is highly
recommended to perform calibration for NH
Ammonia is set on bays 2 to 5 a warning will be displayed as
mentioned here are not available to be calibrated for BW Ultra using IntelliDoX.
NOTE: In 1.024 Firmware Version the CO
using a 5 gas combination. If calibration is skipped once, all gases will be skipped, and
calibration failure will be flagged.
NOTE: SO
Calibration and / or quad gas cannot be individually skipped.
2

Replace a Sensor

Use only sensors designed by Honeywell for the Honeywell BW Ultra detectors. Replace the
sensors in a non-hazardous location.
The Honeywell BW Ultra detector can be configured for a maximum of 5 gases and may contain
dummy sensors.
If an HCN is set up on the 5th, after cal/bump, the device enters to high alarm (HCN) this is
due to high cross sensitivity. The type of sensor can only be configured at factory during
assembly process.
If the user requests a BW Ultra with a dummy sensor placed in the analog sensor position, then
this sensor won't appear in the configuration options.

Review Logs

Many of the detector events are logged and can be reviewed via IntelliDoX or BLE. Typical
logged events would be:
BW Ultra
) Sensor is installed in the BW Ultra and it is configured in the
2
2
, NO
, and HCN calibration is available in all bays. For Ammonia (NH
2
2
must be performed in order to complete the set-up process
can be done using IntelliDoX. For BW Ultra
2
only on the first bay. For this reason, when NH
3
can be calibrated along with the quad gas mixture
2
34
.
2
. Other sticky gases not
) the
3
,
3
User Manual

Advertisement

Table of Contents
loading

Table of Contents