GE Measurement & Control Druck DPI 620 Genii User Manual

GE Measurement & Control Druck DPI 620 Genii User Manual

Advanced modular calibrator
Hide thumbs Also See for Measurement & Control Druck DPI 620 Genii:
Table of Contents

Advertisement

Quick Links

User Manual
GE Measurement & Control
Druck DPI 620 Genii
Advanced Modular Calibrator
© 2013 General Electric Company. All Rights Reserved. Specifications are
subject to change without notice. GE is a registered trademark of General
Electric Company. Other company or product names mentioned in this
document may be trademarks or registered trademarks of their respective
companies, which are not affiliated with GE
Druck DPI620 Genii

Advertisement

Table of Contents
loading

Summary of Contents for GE Measurement & Control Druck DPI 620 Genii

  • Page 1 Advanced Modular Calibrator © 2013 General Electric Company. All Rights Reserved. Specifications are subject to change without notice. GE is a registered trademark of General Electric Company. Other company or product names mentioned in this document may be trademarks or registered trademarks of their respective...
  • Page 2: Table Of Contents

    User Manual Druck DPI620 Genii Contents Overview ......................9 Equipment in the Box ................. 9 Optional Items ..................10 Observance of the User Manual .............. 10 General Safety Precautions ..............11 General Warnings ..................12 Electrical warnings ..................13 Pressure Warnings ..................16 Preparing the Instrument .................
  • Page 3 User Manual Druck DPI620 Genii 1.13 Display Icons ..................... 23 1.14 Software and Firmware Upgrades ............24 1.14.1 Viewing Software Revision .............. 24 1.14.2 Upgrading the Software ..............24 1.15 Maintenance ..................... 29 1.15.1 Cleaning .................... 29 1.16 Instrument Return ..................30 1.16.1 Returned Material Procedure for USA ..........
  • Page 4 User Manual Druck DPI620 Genii Example Procedure: Measure or Source Current with 24V Loop ..... 50 2.10 Example Procedure: Measure or Source Frequency Signals ..... 52 2.11 Example Procedure: Measure/Simulate a Resistance Temperature Detector (RTD) ...................... 54 2.12 Example Procedure: Measure or Simulate a Thermocouple (TC) .... 56 2.13 Example Procedure: Switch Test ..............
  • Page 5 User Manual Druck DPI620 Genii 4.4.1 Transfer ....................78 4.4.2 Erase ..................... 79 4.4.3 Memory Status ..................79 Data Format ....................80 Documentation ....................81 Analysis ..................... 81 Set-up ....................... 84 5.2.1 Define the Reference Channel ............84 5.2.2 Define each Input Channel ..............85 Analysis Function ..................
  • Page 6 User Manual Druck DPI620 Genii 6.12.2 Method Example - D/A Trim ............99 6.13 Preferences ..................... 100 6.14 Failed to find Device ................101 Foundation™ Fieldbus ..................102 Introduction .................... 102 Startup ....................102 FOUNDATION™ Fieldbus Toolbar ............105 Scanning For Devices ................107 7.4.1 Context Sensitive Menu ..............
  • Page 7 User Manual Druck DPI620 Genii 7.11.3 Advanced ..................132 Calibration Procedures ................... 133 Before Starting ..................133 Procedures (CH1/CH2): Current (measure) ..........136 Procedures (CH1/CH2): Current (Source) ..........138 Procedures (CH1/CH2): DC mV/Volts (measure) ........140 Procedures (CH1): DC mV/Volts (source) ..........142 Procedures (CH1): Frequency (measure/source) ........
  • Page 8 User Manual Druck DPI620 Genii 06 June 2013 Page 8 of 179...
  • Page 9: Overview

    User Manual Druck DPI620 Genii Overview The Druck DPI620 Genii is a battery-powered instrument for electrical measure and source operations and HART® communications. The Druck DPI620 Genii also supplies the power and user interface functions for all optional items. The touch-screen displays up to six different parameters Equipment in the Box The following items are supplied with the Druck DPI 620 Genii:...
  • Page 10: Optional Items

    User Manual Druck DPI620 Genii Optional Items The items that follow are optional items which can be used with the Druck DPI 620 Genii: • Pressure Module Carrier, MC 620, this attaches directly to the Druck DPI 620 Genii to make a fully integrated pressure instrument •...
  • Page 11: General Safety Precautions

    User Manual Druck DPI620 Genii General Safety Precautions Read and obey all the operator's local Health and Safety regulations and Safe Working Procedures or Practices. When doing a procedure or task: • Use only the approved tools, consumable materials and spares to operate and maintain the equipment •...
  • Page 12: General Warnings

    User Manual Druck DPI620 Genii General Warnings • It is dangerous to ignore the specified limits for the instrument or its related accessories. This can cause injuries. • If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired •...
  • Page 13: Electrical Warnings

    To prevent an explosion or fire, do not short circuit, do not disassemble, and keep it safe from damage • To prevent an explosion or fire, use only the GE specified battery (Part: 191-356), power supply (Part: 191-339) and battery charger (Part: IO620-CHARGER) •...
  • Page 14 User Manual Druck DPI620 Genii • The power supply input range is 100 – 240Vac, 50 to 60Hz, 250mA, installation category CAT I • Position the power supply so not to obstruct the supply disconnecting device. • Note that the operating and storage temperature range of the mains PSU does not match that of the DPI620.
  • Page 15 User Manual Druck DPI620 Genii The following summary of installation and measurement overvoltage categories are derived from IEC61010-1. The overvoltage categories indicate the severity of overvoltage transients Overvoltage Description Category Overvoltage category I has the least severe overvoltage transients. Generally CAT I CAT I equipment is not designed to be directly connected to the mains supply.
  • Page 16: Pressure Warnings

    User Manual Druck DPI620 Genii Pressure Warnings • Some liquid and gas mixtures are dangerous. This includes mixtures that occur because of contamination. Make sure that the equipment is safe to use with the necessary media • To prevent a dangerous release of pressure, isolate and bleed the system before disconnecting a pressure connection •...
  • Page 17: Preparing The Instrument

    User Manual Druck DPI620 Genii • Connect only clean equipment to the instrument. To avoid any contamination, an external filter is recommended Preparing the Instrument On receipt of the instrument check the contents in the box, listed in section 1.1. It is recommended to retain the box and packaging for future use Install the Battery Remove the five Pozidriv screws (A) (Ref: Figure 1-1)
  • Page 18 User Manual Druck DPI620 Genii Figure 1-1 06 June 2013 Page 18 of 179...
  • Page 19: Charge The Battery

    User Manual Druck DPI620 Genii 1.10 Charge the Battery Connect the DC power supply/battery charger unit in to the +5V DC connection on the side of the unit (Ref: Figure 1-2) The battery can also be charged using the USB connections (Ref: Figure 1-2) The unit can be On or Off when charging.
  • Page 20: Basic Modes

    User Manual Druck DPI620 Genii 1.11 Basic Modes 1.11.1 Power On From OFF – momentarily press the power button until the display flashes (Ref: Figure 1-2) 1.11.2 Power Off Press and hold power button until the screen is blank 1.11.3 Sleep Mode Press and release power button for sleep mode 1.11.4...
  • Page 21: Druck Dpi 620 Genii, Modes

    User Manual Druck DPI620 Genii 1.12 Druck DPI 620 Genii, Modes The Druck DPI 620 Genii can be used as follows: • Calibrator (with independent functions on each of six channels) - Data logging capabilities - Documenting capabilities • HART® Communicator •...
  • Page 22: Dashboard Navigation

    User Manual Druck DPI620 Genii 1.12.1 Dashboard navigation The Dashboard is navigated by swiping a finger from top to bottom while touching the screen. Functions screens are navigated by swiping a finger from right to left while touching the screen Figure 1-3 Dashboard Note: Fieldbus is not installed on all units 06 June 2013...
  • Page 23: Set Date, Time And Language

    User Manual Druck DPI620 Genii 1.12.2 Set Date, Time and Language To access Date, Time and Language menus select: DASHBOARD >> SETTINGS >> DATE >> TIME >> LANGUAGE 1.12.3 Themes Two themes are available: Dark and Light; select the correct theme for the light level.
  • Page 24: Software And Firmware Upgrades

    Note: If the software revision number is highlighted red then an upgrade is available 1.14.2 Upgrading the Software Follow the website instructions to download the files onto a USB flash memory drive. www.ge-mcs.com 1. Select: DASHBOARD >> ADVANCED 2. Enter the calibration PIN: 5487 3. Select the button.
  • Page 25 User Manual Druck DPI620 Genii • Upgrade the Application software and SDC625 Application. 1. Copy the ‘AMC’ application folder into the root of a USB flash memory device 2. Put the USB flash memory drive in the USB type A connector 3.
  • Page 26 User Manual Druck DPI620 Genii • Upgrade the HART processor Application and Boot Loader 1. Create a folder named ‘HART’ in the root of a USB flash memory device 2. Copy the files ‘DK416.s19’ and ‘DK417.s19’ into the ‘HART’ folder 3.
  • Page 27 User Manual Druck DPI620 Genii • Upgrade the HART Device Library By default the HART device library is stored on the micro SD Card. 1. Set the DPI620 Genii USB client port to Storage Device mode by selecting: DASHBOARD >> DEVICES >>...
  • Page 28 User Manual Druck DPI620 Genii Figure 1-4 Hart DD Directory Structure Note: • If a mistake is made during upgrade and there are no files to upload, follow the on-screen instructions and complete the procedure. • When an upgrade completes normally, the initial operation of the touch screen may be slower (a period of approximately 30 seconds).
  • Page 29: Maintenance

    User Manual Druck DPI620 Genii 1.15 Maintenance The DPI620 Genii instrument contains no user serviceable parts and should be returned to a GE service center for repair 1.15.1 Cleaning CAUTION Do not use solvents or abrasive materials. Clean the case and display with a lint free cloth and a weak detergent solution.
  • Page 30: Instrument Return

    GE Service Center Approved Service Agents Web site: www.ge-mcs.com Contact the GE Service Center, either by 'phone, fax or e-mail to obtain a Returned Material Authorization (RMA) number, providing the following information: Product (i.e. Druck DPI 620 Genii)
  • Page 31: Returned Goods Procedure For Europe

    If the instrument is unserviceable and requires a repair return to a GE Service Centre Approved Service Agents Website www.ge-mcs.com Contact the GE Service Centre, either by 'phone, fax or e- mail to obtain a Returned Goods Authorization (RGA) number, providing the following information: Product (i.e. DPI 620)
  • Page 32: Instrument Disposal In The European Union

    Do not dispose of this product or its battery as household waste. (Ref: mark) Use an approved organization that collects and/or recycles the applicable item For more information contact • GE Sensing customer service department: (www.ge-mcs.com) • Local government office 06 June 2013 Page 32 of 179...
  • Page 33: Packaging For Storage Or Transportation

    User Manual Druck DPI620 Genii 1.17 Packaging for Storage or Transportation To store the unit or to return the unit for calibration or repair carry out the following procedures: Pack the Instrument To return the instrument for calibration or repair, complete the return goods procedure (Ref: 1.15) Return the instrument to the manufacturer or an approved service agent for all repairs...
  • Page 34: Electrical Operations

    User Manual Druck DPI620 Genii Electrical Operations Basic Calibrator Operation Select: DASHBOARD >> CALIBRATOR Select the channel by performing the following tasks • Swipe to the TASK MENU by swiping the display from right to left. Figure 2-1 Task Menu 06 June 2013 Page 34 of 179...
  • Page 35 User Manual Druck DPI620 Genii • Select the CUSTOM TASK option. This will allow the user to set up Channels 1 & 2 in addition to the Pressure Channels, USB (IDOS) and Communications (HART or Foundation Fieldbus) Figure 2-2 Task Settings Menu 06 June 2013 Page 35 of 179...
  • Page 36 User Manual Druck DPI620 Genii • Select CH1 or CH2 to enter the CHANNEL SETTINGS menu Figure 2-3 Channel Settings Menu 06 June 2013 Page 36 of 179...
  • Page 37 User Manual Druck DPI620 Genii Setup a channel for measurement • FUNCTION selects the function required (eg Current or Voltage). For more options scroll down the menu by swiping the display from bottom to top. • UNITS selects the type of unit required, (eg Hz, kHz), please note that there may only be 1 type of unit available in particular Functions.
  • Page 38 User Manual Druck DPI620 Genii Other Task Menu Options available are: • DEFAULT – Returns the settings to the default settings The remaining options require a Pressure Module to be attached, please refer to Section 3. • (P1-P2) TO V CH2 – Read differential output between both channels using CH2.
  • Page 39: Set The Function Utility Options

    User Manual Druck DPI620 Genii Set the Function Utility Options For each function only one utility may be active. Not all source and measure functions have associated utilities For all options the button resets the additional readings 2.2.1 Max/Min.Avg This utility is only available with measure functions. The additional values displayed show the minimum, maximum, and average values of the input signal Figure 2-4 Max/Min Example...
  • Page 40: Switch Test

    User Manual Druck DPI620 Genii 2.2.2 Switch Test This utility is available with measure or source functions. The additional values displayed show signal values (measure or source) when the instrument detects a switch opening and closing. The difference between the two values is displayed as hysteresis value for the switch.
  • Page 41: Relief Valve

    User Manual Druck DPI620 Genii 2.2.3 Relief Valve This utility is only available with measure functions. This utility tests circuits or mechanisms that have a cut-out response when an input reaches a defined threshold value. The utility allows the user to select a mode of operation which can be rising or falling.
  • Page 42 User Manual Druck DPI620 Genii Falling Rising Figure 2-7 Relief Value Utility 06 June 2013 Page 42 of 179...
  • Page 43: Measurement Display Options

    User Manual Druck DPI620 Genii Measurement Display Options There are 2 display views in the CALIBRATOR screen when multiple channels are in use: • Figure 2-8 displays a reduced view of all the selected channels. Figure 2-8 Calibration Window – Reduced View 06 June 2013 Page 43 of 179...
  • Page 44 User Manual Druck DPI620 Genii • Figure 2-9 displays an expanded view of the selected channel and minimizes the remaining channels. Figure 2-9 Calibration Window - Expanded View The displays options can be changed by pressing the channel the user wants to display in expanded view. Selecting displays all channels in the reduced view 06 June 2013...
  • Page 45: Example Procedure: Measure Or Source Current

    User Manual Druck DPI620 Genii Example Procedure: Measure or Source Current Figure 2-10 shows CH1 set-up to measure or source a current with external loop power Note: When using the CH2 connectors, set CH 2 to measure or source these ranges with internal or external loop power. Set the appropriate function by selecting mA or mA +24V.
  • Page 46: Example Procedure: Measure Dc Voltage

    User Manual Druck DPI620 Genii Example Procedure: Measure DC Voltage Figure 2-11 shows CH1 set-up to measure a DC voltage (0 to 30 V) or DC mV (0 to 2000 mV) Note: When using the CH2 connectors, set-up CH2 to measure these range. Figure 2-11 Measure DC Volts or DC mV on CH1.
  • Page 47: Example Procedure: Measure Ac Voltage (Ch1), 0 To 20 Vrms Only

    Example Procedure: Measure AC Voltage (CH1), 0 to 20 Vrms Only WARNING To prevent electrical shocks, use only the GE specified AC probe (Part: IO620-AC) to measure AC voltages that are more than 20 Vrms (maximum: 300 Vrms). Refer to Section 2.7...
  • Page 48: Example Procedure: Measure Ac Voltage (Ch1) With The Ac Probe

    Example Procedure: Measure AC Voltage (CH1) with the AC Probe WARNING To prevent electrical shocks, use only the GE specified AC probe (Part: IO620-AC) to measure AC voltages that are more than 20 Vrms (maximum: 300 Vrms). Attach it to the specified connections only...
  • Page 49: Example Procedure: Source Dc Voltage (Ch1)

    User Manual Druck DPI620 Genii Example Procedure: Source DC Voltage (CH1) Figure 2-14shows CH1 Set-up to source a DC voltage on CH1 (0 to 20V) Figure 2-14 Source Voltage on CH1. Range 0 to 20V. 1. Set the applicable software options 2.
  • Page 50: Example Procedure: Measure Or Source Current With 24V Loop

    User Manual Druck DPI620 Genii Example Procedure: Measure or Source Current with 24V Loop Figure 2-15 and Figure 2-16 show CH2 set-up to measure (±55mA) or source (0 to 24mA) a current with internal loop power (Selectable to 24 V or 28V). Figure 2-15 Measure current on CH2 with internal loop power (Range: ±55mA) Figure 2-16 Source current on CH2 with internal...
  • Page 51 User Manual Druck DPI620 Genii 1. Set the applicable software options 2. Complete the electrical connections and continue with the measure or source operation 3. Source only (Automation): Set the applicable output value 06 June 2013 Page 51 of 179...
  • Page 52: Example Procedure: Measure Or Source Frequency Signals

    User Manual Druck DPI620 Genii 2.10 Example Procedure: Measure or Source Frequency Signals Figure 2-17 and Figure 2-18 show CH1 set-up to measure or source a frequency. The units could be Hz, kHz or counts (cpm or cph) Figure 2-17 Example A - Measure Frequency on CH1 (Range 0 to 50 kHz) Figure 2-18 Example B - Source Frequency CH1 (Range 0 to 50 kHz)
  • Page 53 User Manual Druck DPI620 Genii Example A Example B Measure frequency on Source frequency on CH1 Range: 0 to 50 kHz Range: 0 to 50 kHz Waveform: Square Amplitude:5.0 V Trigger level: 2.5V Note (If in Auto Trigger, trigger value shows current auto trigger level) 1.
  • Page 54: Example Procedure: Measure/Simulate A Resistance Temperature Detector (Rtd)

    User Manual Druck DPI620 Genii 2.11 Example Procedure: Measure/Simulate a Resistance Temperature Detector (RTD) Figure 2-19 and Figure 2-20 show CH1 set-up to measure or simulate an RTD. A 4-wire configuration gives the best accuracy; a 2-wire configuration has the lowest accuracy (4- wire RTD shown) Figure 2-19 PT100 RTD Measure CH1 4-Wire (Range 0 to 850 °C)
  • Page 55 User Manual Druck DPI620 Genii Note: To measure or simulate pure resistance Ω, select the Resistance function(Range 0 to 4000 Ω) 1. Set the applicable software options 2. Complete the electrical connections 3. If necessary change the RTD Type SETTINGS >> RTD TYPE 06 June 2013 Page 55 of 179...
  • Page 56: Example Procedure: Measure Or Simulate A Thermocouple (Tc)

    User Manual Druck DPI620 Genii 2.12 Example Procedure: Measure or Simulate a Thermocouple (TC) Figure 2-21 and Figure 2-22 show CH1 set-up to measure or simulate a TC temperature. To measure or simulate TC millivolts, set the TC mV function Figure 2-21 K-TypeThermocouple Measure CH1 (Range 0 to 1372 °C) Figure 2-22 Figure 2 21K-TypeThermocouple Source...
  • Page 57 User Manual Druck DPI620 Genii Note: To measure or simulate TC millivolts, set the TC mV function 1. Set the applicable software options 2. Complete the electrical connections 3. If necessary change the thermocouple type SETTINGS >> TC TYPE 4. Set CJ Compensation mode SETTINGS >>...
  • Page 58: Example Procedure: Switch Test

    User Manual Druck DPI620 Genii 2.13 Example Procedure: Switch Test CH1, P1, P2 and IDOS functions use the CH2 switch connections. CH2 functions use the CH1 switch connections Switch operation When setting the Switch Test utility on one channel, the software automatically sets-up the other channel for the switch connections Note: If there is a measure or source function on the...
  • Page 59 User Manual Druck DPI620 Genii Figure 2-23 CH1 Thermocouple Source CH2 Switch Test Figure 2-23 shows a thermocouple switch test: 1. Set the applicable software options • The TC is set to source a temperature • The UTILITY is set to SWITCH TEST. The AUTOMATION is set to RAMP 2.
  • Page 60 User Manual Druck DPI620 Genii to start the “Ramp” cycle 6. Use to stop the “Ramp” cycle 7. Use 8. If necessary, supply the output values in the opposite direction until the switch changes condition again 9. The display will show the following: •...
  • Page 61: Measure Pressure: Idos Option

    User Manual Druck DPI620 Genii 2.14 Measure Pressure: IDOS Option Optional item. An IDOS Universal Pressure Module (UPM) uses Intelligent Digital Output Sensor (IDOS) technology to measure the applied pressure and supply the data to an IDOS instrument. Before using an IDOS module, (Ref: User Manual: K0378, Druck IDOS UPM) Note: To attach an IDOS module to the Druck DPI 620 Genii calibrator use an IO620-IDOS-USB adaptor...
  • Page 62: Idos Option Assembly Instructions

    User Manual Druck DPI620 Genii 2.14.1 IDOS Option Assembly Instructions Step Procedure Attach one end of the adaptor IO620- IDOS-USB to the applicable IDOS module Push the Type A end of USB Cable into the USB socket on the instrument and the type B end into the adaptor (IO620- IDOS-USB) Set the instrument power on...
  • Page 63: Idos Function Procedures

    User Manual Druck DPI620 Genii 2.14.2 IDOS Function Procedures Set the calibrator functions for use on the display. This includes: • IDOS function (to measure the pressure) • Maximum: 6 functions • If necessary, change the Units for the function •...
  • Page 64: Error Indications

    User Manual Druck DPI620 Genii 2.15 Error Indications Under range: The display shows this symbol for this condition: <<<<< Reading < Negative FS - (2% of negative FS) Over range: The display shows this symbol for this condition: >>>>> Reading > Positive FS + (2% of positive FS) If the display shows <<<<...
  • Page 65: Pressure Indicator Operation (Mc620)

    User Manual Druck DPI620 Genii Pressure Indicator Operation (MC620) This section gives examples of how to connect and use the instrument to measure pressure with the module carrier (MC 620) and the applicable pressure modules (PM 620). To make a fully integrated pressure calibrator instrument with one of the three Pressure Stations, refer to the User Manual for the PV62x Series of Pressure Stations, K0457 Do not use a damaged instrument...
  • Page 66 User Manual Druck DPI620 Genii The label includes: Pressure range: - g: gauge; a: absolute, Serial Number (S/N); manufacturer: name, address, website This section shows the parts of the Module Carrier (MC 620) and Pressure Module (PM 620) CAUTION To prevent damage to the PM 620 module, only use it within the specified pressure limit on the label When the items are attached to the Druck DPI 620 Genii it is a fully integrated pressure indicator, measuring pneumatic or...
  • Page 67: Assembly Instructions

    User Manual Druck DPI620 Genii Assembly Instructions 3.1.1 Pressure Connections WARNING Pressurized gases and fluids are dangerous. Before attaching or disconnecting pressure equipment, safely release all the pressure. The pressure ports for external equipment use “Quick fit” pressure adaptors Procedure (Attaching External Equipment) Remove the adaptor from the pressure port Use an applicable seal for the pressure connection NPT type: Use an applicable sealant on the...
  • Page 68 User Manual Druck DPI620 Genii menus to set-up the necessary operations 1. Align the two slots on the calibrator with the two posts on the module carrier. 2. When the posts are fully engaged in the slots, tighten the two screws hand-tight. 3.
  • Page 69: Procedure Overview

    User Manual Druck DPI620 Genii 3.3.1 Procedure Overview To use the pressure indicator, complete the procedures that follow: Pressure function (P1 and/or P2) Channel 1 (CH 1) electrical function (measure or source) Channel 2 (CH 2) electrical function (measure or source) 06 June 2013 Page 69 of 179...
  • Page 70 User Manual Druck DPI620 Genii If required, change the Units or the function. If necessary, set a Utility for the function: • Max/Min/Avg • Switch Test • Relief valve • Leak Test 06 June 2013 Page 70 of 179...
  • Page 71 User Manual Druck DPI620 Genii If required, change the Settings for the pressure function: • Process (Tare, Alarm, Filter, Flow, Scaling • Leak Test (Only when the Utility is set) • Relief valve (Only when the Utility is set) • Zero. Zero the gauge sensor before use When all the software selections are complete, make the applicable pressure and electrical connections 06 June 2013...
  • Page 72: Set Up A Leak Test

    User Manual Druck DPI620 Genii Set up a Leak Test Set the Utility to Leak Test After setting the Utility to Leak Test, set the options that follow: Wait Time: The time before the test starts in hours:minutes:seconds (hh:mm:ss). Test Time: The period of the leak test in hours:minutes:seconds (hh:mm:ss) Note: To set the Leak Test options, a pressure module must...
  • Page 73: Set The Pressure Module To Zero

    User Manual Druck DPI620 Genii Set the Pressure Module to Zero Use this option to write a new zero pressure value to the pressure module in use. The sensor adjustment is permitted if it obeys the condition that follows: • Adjustment ≤10% FS positive pressure value (for the Sensor) Note: To make a temporary adjustment for zero, use the Tare function...
  • Page 74: Error Indications

    User Manual Druck DPI620 Genii Error Indications If the display shows <<<< (under range) or >>>> (over range) • Make sure that the range is correct • Make sure that all the related equipment and connections are serviceable. Under range: The display shows this symbol for this condition •...
  • Page 75: Data Logging Operation

    User Manual Druck DPI620 Genii Data Logging Operation Select the Data Logging option on the Dashboard. The Data Logging function records instrument readings so they can be reviewed or analyzed. The data file can be reviewed by using the following: •...
  • Page 76: Set-Up

    User Manual Druck DPI620 Genii Set-up Before starting set all channels to the correct functions. To access the Data logging function do the following: • Dashboard > Data Logging > Setup. Use the menu that follows to set the Data Logging functions •...
  • Page 77: Operation

    User Manual Druck DPI620 Genii Operation To begin data logging tap ‘Start logging’ button. In ‘Periodic’ mode an icon flashes in the status bar each time a data point is logged. In ‘Key Press’ mode tap the Log Button to log a data point. To stop data logging the tap Stop.
  • Page 78: File Management

    User Manual Druck DPI620 Genii File Management The data log file management options are as follows: • Transfer: Upload data log files to another computer • Erase: delete data log files • Memory Status: Displays amount of free memory 4.4.1 Transfer Data may be transferred as follows: •...
  • Page 79: Erase

    User Manual Druck DPI620 Genii 4.4.2 Erase The Erase options are as follows: • Erase One File: Select file and tap tick bottom right on the screen to erase • Clear Internal: Clears all internal files 4.4.3 Memory Status The memory status button will show the amount of available memory in the areas that follow: •...
  • Page 80: Data Format

    User Manual Druck DPI620 Genii Data Format The data files are produced in a Comma Separated Variable (csv) format (refer to Figure 4-1).. This allows the data to be imported into a spreadsheet (e.g. Microsoft® Excel). The first section of the data file contains the following: •...
  • Page 81: Documentation

    User Manual Druck DPI620 Genii Documentation This chapter describes the Documenting functions available with the Druck DPI 620 Genii calibrator and are as follows: • Analysis • Run Procedure Analysis The Analysis function takes readings from two or more DPI 620 channels to calibrate the transfer data of the device being tested.
  • Page 82 User Manual Druck DPI620 Genii The other channel is the Input channel and is used as follows: • Measures the output signal from the device • If calibrating a process transmitter it could be CH2 in Current Measure mode A second input channel can also be used, to transfer data between three points in the signal path and may be calibrated at the same time, as in the example that follows.
  • Page 83 User Manual Druck DPI620 Genii When the DPI620 is set-up, do the following: 1. Set the Reference signal to each calibration signal • At each value the Analysis function calculates the difference of each Input channel to the ideal transfer characteristic and compares this to a tolerance limit •...
  • Page 84: Set-Up

    User Manual Druck DPI620 Genii Set-up Set the Druck DPI 620 Genii channels in the Calibrator function Connect the calibrator to the device Enter the Documenting function Tap the Analysis button 5.2.1 Define the Reference Channel Tap the channel button that is to be used as the Reference channel for the analysis Set the channel type to Reference All other channel settings for that channel...
  • Page 85: Define Each Input Channel

    User Manual Druck DPI620 Genii 5.2.2 Define each Input Channel 1. Tap each Input channel button to set the options that follow: • Scaling - the scaling values are four set values • The maximum and minimum Reference signal values (Reference High and Reference Low) •...
  • Page 86: Analysis Function

    User Manual Druck DPI620 Genii Analysis Function Set Input channel parameters, and tap Start. The Analysis window displays the following: • The deviation of each Input channel from the ideal transfer characteristic • A tolerance limit test icon • Pass [Pass] (within tolerance test limit) •...
  • Page 87: Run Procedure

    • A copy of the 4 Sight Calibration Software • Standard USB lead (as supplied) • A Druck DPI 620 Genii calibrator device driver available as a download from the website www.ge-mcs.com 06 June 2013 Page 87 of 179...
  • Page 88 User Manual Druck DPI620 Genii 5.4.1 Sequence to Upload and Download File Step Procedure 1. Connect Standard USB lead (as supplied) to the Druck. DPI 620 Genii calibrator. 2. Connect the lead to the USB port on the computer that has the calibrator manager installed.
  • Page 89: Hart® Operations

    User Manual Druck DPI620 Genii HART® Operations The Druck DPI 620 Genii can communicate with devices that use the HART® protocol • The Universal and Common Practice commands specified in HART®revision 5 to 7 • Devices that support Device Descriptions (DD). This section includes procedures to use the HART functions available in the calibrator HART®...
  • Page 90: Start-Up

    User Manual Druck DPI620 Genii The Druck DPI 620 Genii can be used to communicate with other devices that use a HART®application as follows: • One HART device (a master device) starts and controls the communications • The field device (a slave device) uses each command from the master device to make a change and/or send data back •...
  • Page 91: Screen Format

    User Manual Druck DPI620 Genii Screen Format The Druck DPI 620 Genii displays the HART®application screen in light or dark mode Figure 6-1 HART® Window The HART®window shows the following: • Primary Variable Lower Range Value (PV LRV) • Primary Variable Upper Range Value (PV URV) •...
  • Page 92: Hart® Connections

    User Manual Druck DPI620 Genii HART® Connections Before setting-up the electrical connections between the HART® device and the Druck DPI 620 Genii get the correct connection scheme (Ref: Help menu on Dashboard) Power Supply from the Calibrator A 24 V or 28 V loop drive can be supplied using the CH2 mA (24V) measure function In the example that follows, the Druck DPI 620 Genii supplies the loop power and a 250 Ω...
  • Page 93: External Loop Power

    User Manual Druck DPI620 Genii External Loop Power In the example that follows, there is an external power supply Figure 6-3 Measure current on CH2 without 24 V loop power. HART® function is enabled and 250 Ω resistor is enabled 06 June 2013 Page 93 of 179...
  • Page 94: Communicator Attached To A Network

    User Manual Druck DPI620 Genii Communicator Attached to a Network In the example that follows, the calibrator connects directly to a network. There must be a 250 Ω resistor in series with the loop power supply and the HART® device Figure 6-4 No function on CH2.
  • Page 95: Data Display

    User Manual Druck DPI620 Genii 6.10 Data Display The display data is color coded as follows: • - HART® transmitter data • Blue - Druck DPI 620 Genii channel data • Black - can be edited For data display identification icons (Ref: Section 11) Common acronyms used are as follows: •...
  • Page 96: Editing Values

    User Manual Druck DPI620 Genii 6.11 Editing Values Any value displayed in black with a [V] or [E] icon can be edited. Edit variables as follows: 1. Tap the variable 2. If a selection window is opened tap the variable 3.
  • Page 97: Executing Methods

    User Manual Druck DPI620 Genii 6.12 Executing Methods Not all HART® devices have the same methods available. The function, purpose and execution of each method can be different. Methods may include the following: • Self-test • Loop test • Sensor trim •...
  • Page 98: Method Example - Self-Test

    User Manual Druck DPI620 Genii Some methods need entry values, use the alpha/numeric keypads 3. A drop down menu can be used for method selection options 4. Some methods require input from the DPI 620 instrument channels. A drop down menu displays the channels as follows: •...
  • Page 99: Method Example - D/A Trim

    User Manual Druck DPI620 Genii 6.12.2 Method Example - D/A Trim The Druck DPI 620 Genii can perform an analogue trim on the 4 to 20 mA loop without connecting to any external reference meters 1. Navigate to the calibrator folder 2.
  • Page 100: Preferences

    User Manual Druck DPI620 Genii 6.13 Preferences Select the applicable icon to change the poll address, short tag or long tag • Poll Address - when each transmitter has a unique address • Short tag - if the transmitter supports 8 character tags •...
  • Page 101: Failed To Find Device

    User Manual Druck DPI620 Genii 6.14 Failed to find Device Fault Possible Action Cause Failed Power Check device is turned on to find Supply Check applicable fuses device Check the supply voltage is within limits Check the loop current is in the range 3.5mA to 24mA Device under Check device is HART®...
  • Page 102: Foundation™ Fieldbus

    User Manual Druck DPI620 Genii Foundation™ Fieldbus Introduction FOUNDATION™ Fieldbus (FF) is a device application for configuration of FF enabled field devices. On-line connection is via the integrated H1 modem. Devices directly connected to an H1 field segment can be configured and supported by Startup FOUNDATION™...
  • Page 103 User Manual Druck DPI620 Genii To start the FOUNDATION™ Fieldbus application and connect to a device Connect DPI620 Genii to a H1 FOUNDATION™ Field bus device. Figure 7-1 Example Connection Diagram Foundation Fieldbus Select the FOUNDATION™ Fieldbus channel so that it is in its expanded view (refer to Section 2.3) 06 June 2013 Page 103 of 179...
  • Page 104 User Manual Druck DPI620 Genii Select button Figure 7-2 Calibrator Fieldbus Channel 4. The Main FOUNDATION™ Fieldbus application will now open Note: Updating CH1, CH2, P1, P2 tasks is only possible when the FOUNDATION™ Fieldbus application is not running – to close FOUNDATION™ Field bus application select : (refer to Section 7.3) HOME >>...
  • Page 105: Foundation™ Fieldbus Toolbar

    User Manual Druck DPI620 Genii FOUNDATION™ Fieldbus Toolbar Figure 7-3 FOUNDATION™ Fieldbus Toolbar On entering the FOUNDATION™ Fieldbus application, the toolbar is displayed. The ICONs are greyed out when they are not active. The ICON functions are described below: OPEN CONNECTION – Only available when awaiting an open connection.
  • Page 106 User Manual Druck DPI620 Genii ABORT – Abort Update of parameters, reverts to previous values (refer to Section 7.7.3) STATUS – Provides Device profile of currently connected device (refer to Section 7.5) FUNCTION FINDER – Search FF variables and device functions HOME –...
  • Page 107: Scanning For Devices

    User Manual Druck DPI620 Genii Scanning For Devices The following steps describe how to scan for FOUNDATION™ Fieldbus devices over a FOUNDATION™ Fieldbus H1 connection: Connect DPI620 Genii to H1 bus (refer to section 7.2) Tap the OPEN CONNECTION Icon on the Toolbar to enter device scan screen Figure 7-4 Device Scan Screen...
  • Page 108 User Manual Druck DPI620 Genii Select the SCAN button The ‘scanning’ progress dialog view will open. Any devices found in the selected range will appear in the bus tree window list - All scanned devices are shown as a bold icon with an associated tag. Previous scan results are indicated in grey.
  • Page 109 User Manual Druck DPI620 Genii Selecting any device in the search results and tapping OK will initiate connection to the ‘Device Focus view.’ To Re-scan use the context sensitive menu on the Genii Modem (Refer to section 7.4.1) 06 June 2013 Page 109 of 179...
  • Page 110: Context Sensitive Menu

    User Manual Druck DPI620 Genii 7.4.1 Context Sensitive Menu Context sensitive menus are available for each device in the scan Menu by select and hold. Modem This provides access to the following functions: • Properties (Displays modem profile) • Re-Scan Figure 7-6 Scan Menu - Context Sensitive - Modem...
  • Page 111 User Manual Druck DPI620 Genii Devices This provides access to the following functions: • Tag / Address Change Set Tag Set Address Clear Tag Clear Address • Boot Operational Function (BOF) class change Set Basic Set Link Master Set Bridge •...
  • Page 112: Troubleshooting

    User Manual Druck DPI620 Genii 7.4.2 Troubleshooting • If no device is found when performing scan check: Field wiring. Verify electrical segment connections are in accordance with the specific manual supplied with the field device and segment coupler / power supply The loop is not suffering from interference due to an unstable voltage supply from the segment power supply and / or electrical interference.
  • Page 113: Device Focus View

    User Manual Druck DPI620 Genii Device Focus View In this view the specific information of the device is shown. – PD tag – Device Id – Block List with Target / Actual Mode On entering ‘Device Focus view’ the software will load the Blocks of the target field device and make them available for parameterization.
  • Page 114 User Manual Druck DPI620 Genii Selecting the Scan button returns device scan view (reference Figure 7-5) Selecting the desired block with a single key press will open the Navigation Tree for the block (Refer to section 7.6) 06 June 2013 Page 114 of 179...
  • Page 115: The Navigation Menu Tree

    User Manual Druck DPI620 Genii The Navigation Menu Tree This is the overview of the entire loaded block of the device (this is not the entire field device only one aspect of it) showing all the menus available according to the configured access level.
  • Page 116: Block Header Bar

    User Manual Druck DPI620 Genii 7.6.1 Block Header bar The block header bar indicates the target and actual mode of the block. Figure 7-11 Block Header Bar The Highlighted text indicates the actual mode of the device block. The highlight is shown in green if the target mode is equal to the actual mode of the device block.
  • Page 117: Functional Group View

    User Manual Druck DPI620 Genii Functional Group View The functional groups show all the variables or settings in that menu group along with the current value. Figure 7-12 Function Group View The left side hosts the 'Variable Description Area' and access to the context based adjustment functions.
  • Page 118 User Manual Druck DPI620 Genii Black variables are open for editing under the appropriate access conditions, such as access code or PIN that may be required in a different Functional Group The Navigation Tree bar shows the hierarchy of menus and groups above the current Function Group View.
  • Page 119: Displaying Parameter Help

    User Manual Druck DPI620 Genii 7.7.1 Displaying Parameter Help • The yellow triangle in the corner of the Variable Description area indicates that help is available for that parameter • Context menu opened up with touch and hold action • Selecting Display Help will show the help attributes Figure 7-14 Parameter Help 06 June 2013 Page 119 of 179...
  • Page 120: Refreshing Data

    User Manual Druck DPI620 Genii 7.7.2 Refreshing Data When a refresh is in progress the variable description goes grey and in the right area of the Variable Editing area the pending icon appears. When the read request is fulfilled the variable description goes from grey to black again and the pending icon disappears Refreshing Data can be selected from the context menu.
  • Page 121: Editing Values

    User Manual Druck DPI620 Genii 7.7.3 Editing Values Values which may be editable are shown in black in the “variable editing area” of the function group view (Refer to Figure 7-12). Select the parameter to open for editing. Once Edit is complete, the variable description is highlighted in bold and the commit and cancel icons in the toolbar become active.
  • Page 122: Methods

    User Manual Druck DPI620 Genii If an invalid value is entered the variable will be shown in red and the error icon appears 7.7.4 Methods These are viewed as a button or selection menu. In the case of a button, a momentary touch activates the execution of the associated functionality.
  • Page 123: Function Finder

    User Manual Druck DPI620 Genii Function Finder Function Finder is a way of searching for FF variables and device functions in the on-line device. In complex devices with multiple menus this allows the user to navigate a device without a manual, greatly simplifying the on-line experience, even with an unfamiliar device.
  • Page 124 User Manual Druck DPI620 Genii Figure 7-17 Function Finder 06 June 2013 Page 124 of 179...
  • Page 125: Exporting Data To Main Genii Application

    User Manual Druck DPI620 Genii Exporting Data to Main Genii Application The FOUNDATION™ Fieldbus application allows selected parameters to displayed in the Communications channel windows for the Calibrator screen. The selected parameters are defined by the Export List Menu Figure 7-18 Export List 06 June 2013 Page 125 of 179...
  • Page 126 User Manual Druck DPI620 Genii Parameters are added to the Export List using the context menu in the Variable Description area (refer to Section 7.7) and selecting ‘Add to ‘Export List’’ Note: Only Parameters that return a value can be added to the Export List A Maximum of 6 parameters can be added to the Export List...
  • Page 127: Viewing Exported Variables In Channel Window

    User Manual Druck DPI620 Genii 7.9.1 Viewing Exported Variables in Channel Window Return to Main Application by Minimizing the FOUNDATION™ Fieldbus application >> HOME Minimize Expand the FOUNDATION™ Fieldbus window and select Settings >> PRIMARY VALUE The Export List of selected parameters will be displayed. The selected parameter will be displayed in the FOUNDATION™...
  • Page 128: My Block

    User Manual Druck DPI620 Genii 7.10 My Block My Block allows the user to create Menus of commonly used parameters for easy recall Additional Menus can be created under ‘My Block’ using the context menu accessed with touch and hold gesture. Figure 7-21 Add 'My Block' Menu 06 June 2013 Page 128 of 179...
  • Page 129 User Manual Druck DPI620 Genii Parameters are added to ‘My Block’ (or the created Menu) using the context menu in the Variable Description area (refer to Section 7.7) Figure 7-22 Adding Parameters to 'My Block' 06 June 2013 Page 129 of 179...
  • Page 130: Application Settings

    DPI620 Genii. Allows the user to browse for a specific device to determine that there is support for it To request support for a non-registered DD please contact your local GE Service center (refer to Section 1.16) 06 June 2013 Page 130 of 179...
  • Page 131: Options

    User Manual Druck DPI620 Genii 7.11.2 Options • Poll header every - Sets the refresh rate for the device parameters displayed in the header • Poll all dynamic every - Sets the refresh rate for FF dynamic variables in the function group view (note this setting only becomes valid if the function group option Refresh Vars On is active) (refer to Section 7.7.2) •...
  • Page 132: Advanced

    User Manual Druck DPI620 Genii 7.11.3 Advanced These settings are for advanced users only and it is recommended to retain Default Values 06 June 2013 Page 132 of 179...
  • Page 133: Calibration Procedures

    User Manual Druck DPI620 Genii Calibration Procedures Note: GE Measurement and Control can provide a calibration service that is traceable to international standards Note: GE Measurement and Control recommend returning the instrument to the manufacturer or an approved service agent for calibration. If using an alternative calibration facility, check that it uses the standards that follow.
  • Page 134 User Manual Druck DPI620 Genii Table 8-1 Calibration equipment Function (ppm = parts per million) Current (mA) calibrator. Current (CH1 or CH2) Accuracy - Current measure/source, refer to Table 8-2 or Table 8-3 Volts calibrator. Voltage (CH1 or CH2) Accuracy - Voltage measure/source, refer to mV calibrator.
  • Page 135 User Manual Druck DPI620 Genii Calibration equipment Function (ppm = parts per million) Thermocouple temperature reference unit (0°C) Accuracy: 30 mK AC mV calibrator. AC mV (CH1) Accuracy - AC mV measure, Refer to Table 8-15 AC Volts calibrator. AC Volts (CH1) Accuracy - AC Volts measure, Refer to Table 8-16...
  • Page 136: Procedures (Ch1/Ch2): Current (Measure)

    User Manual Druck DPI620 Genii To do a calibration on a measure or source function, use the advanced menu option. 1. Select: DASHBOARD >> ADVANCED 2. Enter the calibration PIN: 4321 3. Select the button.  4. Select: PERFORM CALIBRATION Then select a function and start the calibration: 1.
  • Page 137 User Manual Druck DPI620 Genii 3. Use the calibration menu (Ref: Section 8.1) to do a three- point calibration (-FS, Zero and +FS) for each range: 20 mA and 55 mA 4. Check the calibration is correct • Select the applicable Current (measure) function, (Ref: Section 2.4, 3).
  • Page 138: Procedures (Ch1/Ch2): Current (Source)

    User Manual Druck DPI620 Genii Procedures (CH1/CH2): Current (Source) Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref: Table 8-1): • CH1/CH2 (24 mA range): (Ref: Section 2.4 and Section 2.9) • CH2 (-24 mA range only): ): (Ref: Section 2.4 and Section 2.9) 2.
  • Page 139 User Manual Druck DPI620 Genii 4. Check the calibration is correct: • Select the applicable Current (source) function(Ref: Section 2.4 and Section 2.9) • Apply the values that follow: CH1/CH2: 0.2, 6, 12, 18, 24 • Check the error is within limits (Ref: Table 8-3) Table 8-3 Current (source) error limits Source mA Calibrator...
  • Page 140: Procedures (Ch1/Ch2): Dc Mv/Volts (Measure)

    User Manual Druck DPI620 Genii Procedures (CH1/CH2): DC mV/Volts (measure) Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref: Table 8-1) 2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on) 3.
  • Page 141 User Manual Druck DPI620 Genii Table 8-4 Millivolts (measure) error limits Applied Calibrator Permitted DPI 620 error uncertainty (mV) (mV) ±2000 0.051 0.14 ±1000 0.040 ±200 0.0051 0.017 ±100 0.0040 0.0125 0 (short 0.0036 0.008 circuit) Table 8-5 Voltage (measure) error limits Applied Calibrator Permitted...
  • Page 142: Procedures (Ch1): Dc Mv/Volts (Source)

    User Manual Druck DPI620 Genii Procedures (CH1): DC mV/Volts (source) Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref: Table 8-1) 2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on) 3.
  • Page 143 User Manual Druck DPI620 Genii Table 8-6 millivolts (source) error limits Calibrator Permitted Source mV uncertainty DPI 620 error (mV) (mV) 0.0001 0.008 0.00046 0.0125 0.0009 0.017 1000 0.003 2000 0.006 0.14 Table 8-7 Voltage (source) error limits Source Calibrator Permitted uncertainty DPI 620 error...
  • Page 144: Procedures (Ch1): Frequency (Measure/Source)

    User Manual Druck DPI620 Genii Procedures (CH1): Frequency (measure/source) Only do one frequency calibration, use either the measure function or the source function. Frequency calibration (measure function) Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref: Table 8-1) 2.
  • Page 145 User Manual Druck DPI620 Genii Frequency calibration (source function) Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref: Table 8-1). 2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on) 3.
  • Page 146 User Manual Druck DPI620 Genii Frequency calibration check • Frequency (measure) calibration check Signal generator: Output = 10 V Unipolar Square wave Druck DPI 620 Genii: Input trigger level = 5 V Units: Hz or kHz as specified in Table or Table 8-9 •...
  • Page 147 User Manual Druck DPI620 Genii • Check the error is within limits (Ref: • Table or Table 8-9) Table 8-8 Hz error limits (measure/source) Measure/ Calibrator Permitted uncertainty source Hz DPI 620 error (Hz) (Hz) (measure) (source) 0.0002 0.0023 0.0026 0.0005 0.0050 0.0053...
  • Page 148: Procedures (Ch1): Frequency Amplitude (Source)

    User Manual Druck DPI620 Genii Procedures (CH1): Frequency Amplitude (source) Do the procedure as follows: Note: The procedure that follows calibrates the “mark” value of the square ware frequency output. The “space” value is fixed and is approximately -120 mV 1.
  • Page 149 User Manual Druck DPI620 Genii 5. Check the calibration is correct. • Set-up the equipment as follows: Source frequency = 0 (For DC output); Waveform = Square • Apply the amplitude values that are applicable to the calibration (Ref: Table 8-10) •...
  • Page 150: Procedures (Ch1): Resistance Measure)

    User Manual Druck DPI620 Genii Procedures (CH1): Resistance measure) Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref: Table 8-1) 2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on) 3.
  • Page 151 User Manual Druck DPI620 Genii 4. Check the calibration is correct • Select the applicable Resistance (measure) function (Ref Section 54) • Make a 4-wire connection to the applicable standard resistor (Ref:Table 8-11) and measure the value • Check the error is within limits (Ref: Table 8-11) Table 8-11 Resistance (measure) error limits Resistor Standard...
  • Page 152: Procedures (Ch1): True Ohms (Measure)

    User Manual Druck DPI620 Genii Procedures (CH1): True Ohms (measure) Do the procedure as follows: 1. Repeat procedure 8.8; in step 3 and 4 select True Ohms 2. Check the error is within limits (Ref: Table 8-1) Table 8-12 True Ohms (measure) error limits Resistor Standard Permitted...
  • Page 153: Procedures (Ch1): Resistance (Source)

    User Manual Druck DPI620 Genii 8.10 Procedures (CH1): Resistance (source) Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref: Table 8-1) 2. Let the equipment get to a stable temperature (minimum: 5 minutes since the last power on) 3.
  • Page 154 User Manual Druck DPI620 Genii 4. Check the calibration is correct: • Select the Resistance (source) function (ref section 2.11) • Apply the resistance values that are applicable to the calibration (Ref: Table 8-13) • Check the error is within limits (Ref: Table 8-13) Table 8-13 Resistance (source) error limits Calibrator Ohms...
  • Page 155: Procedures (Ch1): Tc Mv (Measure Or Source)

    User Manual Druck DPI620 Genii 8.11 Procedures (CH1): TC mV (measure or source) Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref: Table 8-1) 2. Let the equipment get to a stable temperature (minimum: 5 minutes `since the last power on) 3.
  • Page 156: Procedures (Ch1): Cold Junction (Tc Method) And Cj (Measure)

    User Manual Druck DPI620 Genii 8.12 Procedures (CH1): Cold Junction (TC method) and CJ (measure) Note: Do the TC mV (measure) calibration before the cold junction calibration. The conditions for CJ calibration are as follows: • Battery mode (with DC charger disconnected) •...
  • Page 157 User Manual Druck DPI620 Genii CJ (TC Method) Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref:Table 8-1) 2. Set the reference unit temperature: 0°C 3. Let the equipment get to a stable temperature (minimum: 1 hour since the last power on) 4.
  • Page 158 User Manual Druck DPI620 Genii Do the procedure as follows: 1. Connect the applicable calibration equipment (Ref: Table 8-1) 2. Set-up the equipment • Function = TC (measure) • TC Type = K Type • CJ Compensation, Mode = Automatic 3.
  • Page 159: Procedures (Ch1): Ac Mv/Volts (Measure)

    User Manual Druck DPI620 Genii 7. Check the calibration is correct • Select the TC (measure) function 8. Check the equipment gives a TC temperature that is the same as the temperature on the reference unit ±0.1°C (0.2°F) 8.13 Procedures (CH1): AC mV/Volts (measure) Do the procedure as follows: 1.
  • Page 160 User Manual Druck DPI620 Genii 4. Check the calibration is correct • Select the applicable AC mV or AC Volts (measure) function (ref Section 0) 5. Apply the input values that are applicable to the Calibration • AC mV: 10, 500, 1000, 2000 •...
  • Page 161: Procedures: Pressure Indicator Modules (Pm 620)

    User Manual Druck DPI620 Genii 8.14 Procedures: Pressure Indicator Modules (PM 620) Do the procedure as follows: 1. Assemble the pressure indicator with the necessary PM620 modules 2. Connect the instrument to the pressure standard 3. Let the equipment get to a stable temperature (minimum: 60 minutes since the last power on) 4.
  • Page 162 User Manual Druck DPI620 Genii Table 8-17 Calibration pressures Nominal applied pressure Ranges: gage mbar (psi) -FS † Zero < 700 mbar (10.0 psi) > 700 mbar (10.0 psi) -900 (-13.1) †For a three-point calibration, do not apply more than - 90% of the specified FS for the unit.
  • Page 163 User Manual Druck DPI620 Genii 5. Check the calibration is correct: • Select the applicable pressure function • Apply the following pressure values (absolute sensors) • 0, 20, 40, 60, 80, 100 (%FS) • Go back to 0 in the same steps •...
  • Page 164: Procedures: Idos Upm

    User Manual Druck DPI620 Genii 8.15 Procedures: IDOS UPM (Ref: User manual IDOS UPM) When the calibration is complete, the instrument automatically sets a new calibration date in the UPM 06 June 2013 Page 164 of 179...
  • Page 165: General Specification

    User Manual Druck DPI620 Genii General Specification Introduction For a full specification of the Druck DPI 620 Genii calibrator and its related accessories (MC 620 carrier, PM 620 module and PV 62x pressure stations) refer to the relevant product datasheet. The DPI 620 is suitable for indoor use with the following environmental requirements.
  • Page 166 Pressure Equipment Directive - Class: Sound Pressure safety Engineering Practice (SEP) CE Marked Approved Lithium-Polymer battery (GE Part number: 191- 356) Capacity: 5040 mAh (minimum), 5280 mAh Battery power (typical), Nominal voltage: 3.7 V. Charge temperature: 0 to 40°C (32 to 104°F) outside this range, charging stops.
  • Page 167 User Manual Druck DPI620 Genii Note 1: The DPI 620 has been assessed to the European IEC60529 standard as having an ingress protection rating of IP55, but this is for reliability purposes and not for safety reasons Note 2: To meet the immunity requirements of annex A of EN61326-1:2006, when used in an industrial environment, the unit must be battery powered to guarantee measurement specification...
  • Page 168: Manufacturer

    User Manual Druck DPI620 Genii Manufacturer Druck Limited Fir Tree Lane Groby Leicester LE6 0FH England Tel: 0116 231 7100 06 June 2013 Page 168 of 179...
  • Page 169: Display Icons

    User Manual Druck DPI620 Genii Display Icons Table 11-1 Dashboard Icons Advanced Hart® Calibrator Help Data Logging LED (shows status) Blue – Active Red – Alarm Green - Connected Touchscreen Multimeter Documenting Scope Foundation Settings Fieldbus Files Status Applications Devices 06 June 2013 Page 169 of 179...
  • Page 170 User Manual Druck DPI620 Genii Table 11-2: Status Icons Battery Pressure Base Battery Charging Wireless HART® Table 11-3: Button/Footer Icons Auto Rotate Save File Back Back Space Graph Mode Home Exit Log 06 June 2013 Page 170 of 179...
  • Page 171 User Manual Druck DPI620 Genii Table 11-4: Button/Footer Icons Previous Log Next Log Nudge Down Nudge Up More Reset Start Stop Table 11-5: Documenting Icons Analysis Parameter Text Entry Channel Error Channel Linearity Channel Scaling Channel Tolerance 06 June 2013 Page 171 of 179...
  • Page 172 User Manual Druck DPI620 Genii Table 11-6: Documenting Icons Result as Found Result as Left Reference High Reference Low Start Store Reading Table 11-7: List/Channel Icons Alarm Channel USB Channels Channels Channel P1 Channel P2 06 June 2013 Page 172 of 179...
  • Page 173 User Manual Druck DPI 620 Genii Doc Adjust Doc Cal Doc Fail Doc in Limits Doc Linear Doc Root Falling Relief Valve Documenting Filter Light Flow Light Gate Closed Gate Open Measure Complete Max/Min 06 June 2013 Page 173 of 179...
  • Page 174 User Manual Druck DPI 620 Genii Leak Test Measure Next Previous Channel Nudge Play/Pause Percent Step Ramp Relief Valve Repeat Resize Rising Relief Valve Sine Wave Source 06 June 2013 Page 174 of 179...
  • Page 175 User Manual Druck DPI 620 Genii Table 11-8: List/Channel Icons Span Check Square Wave Bullet Switch Actuate Switch Switch Initial State De-actuate Switch Closed Switch Open Switch Test TARE Task Triangular Wave 06 June 2013 Page 175 of 179...
  • Page 176 User Manual Druck DPI 620 Genii Table 11-9: HART® Icons Abort Commit New File Close Comms Failed Comms Off Comms On Comms Start Down Left Right Tree Toggle Navigation Toggle 06 June 2013 Page 176 of 179...
  • Page 177 User Manual Druck DPI 620 Genii Table 11-9: HART® Icons Home Settings Status Dashboard Folder Folder Selected Image Online Method Value Value List 06 June 2013 Page 177 of 179...
  • Page 178 User Manual Druck DPI 620 Genii Table 11-10: Fieldbus Icons Add Connect Bridge Blue tooth Pending Go to Desktop Function Gateway Generic H1 Device Method Mobile Information Open Closed 06 June 2013 Page 178 of 179...
  • Page 179 User Manual Druck DPI 620 Genii Table 11-10: Fieldbus Icons People Error Resource Search Unrecognized 06 June 2013 Page 179 of 179...