Sony Tektronix VX4792 User Manual page 252

Arbitrary waveform generator
Table of Contents

Advertisement

Check Waveform Advance
Mode
VX4792 User Manual
Ch 1 From VX4792
Ch 2 From Function
Generator
Figure F-8: Waveform Output Signal with Negative Gate Signal
5. End procedure: Turn the function generator output off and disconnect the
function generator.
This procedure checks the operation of Waveform Advance mode.
Electrical Characteristic Checked: Operating mode, Waveform Advance, on
page E–4.
Equipment Required: A 50 W coaxial cable and an oscilloscope.
Prerequisites: The instrument must meet the prerequisites listed on page F–3.
Procedure:
1. Install test setup and set test equipment controls (see Figure F–9):
a. Connect the oscilloscope: Connect the WAVEFORM OUTPUT
connector through the coaxial cable to the CH1 vertical input connector
on the oscilloscope.
Performance Verification
F-19