TRIGGER
INPUT
WAVEFORM
OUTPUT
Figure F-4: Triggered Mode Initial Test Setup
VX4792 User Manual
VX4792
a. Connect the oscilloscope: Connect the WAVEFORM OUTPUT
connector through the coaxial cable to the CH1 vertical input connector
on the oscilloscope.
b. Connect the function generator: Connect the TRIGGER INPUT
connector though a coaxial cable to the function generator output
connector.
c. Set the oscilloscope controls:
Vertical:
CH1 coupling:
CH1 scale
CH1 input impedance:
Horizontal
Sweep
Trigger
Source
Coupling
Slope
Level
Mode
Performance Verification
Oscilloscope
Ch 1 Input
CH1
DC
0.2 V/div
50 W
10 ms/div
CH1
DC
Positive
–100 mV
Auto
Function
Generator
Output
F-13