Sony Tektronix VX4792 User Manual page 115

Arbitrary waveform generator
Table of Contents

Advertisement

Command Descriptions
DIAG?
Group
Related Commands
Syntax
Arguments
Responses
Examples
3-32
The DIAG? query returns the selected self-test routine(s), runs the routine, and
returns the results.
CALIBRATION and DIAGNOSTIC
DIAG:SELect, DIAG:STATe, DIAG:RESUlt?
DIAG?
DIAG
?
None
:DIAG:SELECT <Self-test Routine>; [RESULT],<Result>[,<Result>]...
<Self test Routine>::= <label>
where <label> is one of following routines:
ALL
all routines
CPU
CPU unit check routine
SETUp
setup related unit check routine
WMEMory
waveform memory check routine
and where <Result>::=<NR1> is one of following responses:
0
terminated without error
100
detected an error in the CPU unit
600
detected an error in the setup-related unit
700
detected an error in the waveform memory unit
NOTE. The waveform generator does not respond to any commands or queries
issued during the self-test. The self-test takes up to 90 seconds to complete.
DIAG?
might return :DIAG:SELECT ALL;RESULT 0.