Authorized Repair Service - GE AED Pro Operator's Manual

Hide thumbs Also See for AED Pro:
Table of Contents

Advertisement

ANNUAL MAINTENANCE
Perform the following tests annually to confirm that the diagnostics are functioning properly and to verify the integrity of
the case.
Check the Integrity of the Pads and Circuitry
1.
Open the AED Pro lid.
2.
Remove the pads.
3.
Close the lid.
4.
Confirm that the STATUS INDICATOR turns RED.
5.
Open the lid and confirm that the PAD indicator is lit.
6.
Reconnect the pads and close the lid.
7.
Make sure the expiration date is visible through the clear window of the lid.
8.
Check to make sure that the STATUS INDICATOR is GREEN.
9.
Open the lid and confirm that no diagnostic indicators are lit.
10. Check the expiration date of the pads; if expired, replace them.
11. Check the pad's packaging integrity.
12. Close the lid.
Check the Integrity of the Service Indicator (LED) and Circuitry
1.
Immediately after opening the AED Pro lid, press and hold the SHOCK button and confirm that the
SERVICE LED is lit.
2.
Release the SHOCK button.
3.
Close the lid.
4.
Verify that the STATUS INDICATOR remains RED.
5.
Open the lid and confirm that no diagnostic indicators are lit.
6.
Close the lid.
7.
Verify the STATUS INDICATOR turns GREEN
Check the Integrity of the Case
Examine the molded case of the AED Pro for any visible signs of stress. If the case shows signs of stress, contact
Customer Service or contact your local distributor.
Cleaning the AED Pro Case
Gently clean the surface of the AED Pro case with a damp sponge or with a cloth and mild soap.
CAUTION: Case Cleaning Solutions
When disinfecting the case, use a non-oxidizing disinfectant, such as ammonium salts or a
glutaraldehyde based cleaning solution, to avoid damage to the metal connectors.
No periodic safety analysis tests referred to by the IEC 60601-1 international standard are required.

AUTHORIZED REPAIR SERVICE

The AED Pro has no user-serviceable internal components. Try to resolve any maintenance issues with the AED Pro by
using the Troubleshooting Table presented in this chapter. If you are unable to resolve the problem, contact Customer
Service.
WARNING: Shock Hazard
Do not disassemble the AED Pro! Failure to observe this warning can result in personal injury or death. Refer
maintenance issues to authorized service personnel.
Note: The warranty will be void upon unauthorized disassembly or service of the AED Pro.
2023486-201 Rev C
Section 5: Maintenance and Troubleshooting
© 2007 General Electric Company
Page 37 of 53

Advertisement