M Mode Softmenu
Table 4-6
Step
1
Adjust Sweep Speed
2
Adjust Edge Enhance
3
Activate Grey Map
4
Activate Colorize
5
Activate Full Timeline
6
Select Display Format
7
Adjust Dynamic Range
8
Power output
9
Anatomical M
10
Adjust Rejection
4 - 14
9
EVISION
M Mode Softmenu
Task
Section 4-3 - General Procedure
LOGIQ C S
Expected Result(s)
Changes the speed at which the time line is
swept. The following speed values are
available, 1, 2, 3, 4, 6, 8, 12, 16.
Edge Enhance brings out subtle tissue
differences and boundaries by enhancing the
gray scale differences corresponding to the
edges of structures. Adjustments to M Mode's
edge enhancement affects the M Mode only.
Determines how the echo intensity levels
received are presented as shades of gray.
Enables gray scale image colorization. To
deactivate, reselect a Gray Map.
Displays only timeline screen. Press the Full
Timescreen to activate.
Select the format to display B image and M
image on the LCD. Press Display Format,
and select from the pop up menu.
Dynamic Range controls how echo
intensities are converted to shades of gray,
thereby increasing the adjustable range of
contrast.
Optimizes image quality and allows user to
reduce beam intensity. 10% increments
between 0-100%. Values greater than 0.1 are
displayed.
Enter into AMM Mode (Option).
Selects a level below which echoes will not be
amplified (an echo must have a certain
minimum amplitude before it will be
processed).