Section 3-9
Connectivity Setup - Software v7.x (BT'08), v6.x, v5.x and v4.x
NOTE:
If connected to a stand-alone network (Peer-to-Peer network with a Vivid 7 scanner, an
EchoPAC PC workstation and an optional network printer), you should use default delivery
settings.
3-9-1
Introduction
To be able to use the network functions when connected to a hospital network, the scanner must have
a proper network address.
•
Before you can set up the scanner, you need to collect some information.
•
The
this information.
•
Typical source for this information is the network administrator.
NOTE:
The remote use of Common Service Platform is not compatible with use of DHCP. If DHCP is
turned on, remote diagnostics will be disabled.
3-9-2
Contents in this Section
•
3-9-3 "Compatibility" on page 3-43
•
3-9-4 "Select TCP/IP Set-up Screen" on page 3-44
•
3-9-5 "Changing the Computer Name, AE Title and/or Port Number (Port No.)" on page 3-45
•
3-9-6 "Set the Scanner's Network Information - S/W v4.2/v5.2/v6.1/v7.x and later" on page 3-48
•
3-9-7 "Set the Scanner's Network Information - S/W v4.0/v4.1/v5.0/v5.1/v6.0" on page 3-50
•
3-9-8 "Advanced Settings" on page 3-51
•
3-9-9 "DHCP Configuration" on page 3-52
•
3-9-10 "Set the Remote Archive's Network Information" on page 3-55
•
3-9-11 "Save the New Settings" on page 3-55
•
3-9-14 "Set Up Connection to a DICOM Server in a Network" on page 3-58
•
3-9-15 "Export Configuration" on page 3-70
•
3-9-16 "Create a New Dataflow" on page 3-75
•
3-9-17 "Query/Retrieve (Q/R) Setup for Software v7.x/v6.x/v5.x/v4.x" on page 3-78
•
3-9-18 "Mapping of Parameters from Vivid 7 to DICOM" on page 3-85
3 - 42
Section 3-9 - Connectivity Setup - Software v7.x (BT'08), v6.x, v5.x and v4.x
11
"Pre-quote Worksheet for DICOM Network Information" on page 2-14
V
7 S
M
IVID
ERVICE
ANUAL
can be used for gathering