Motor Drive Failure Alarm; Power Failure (Led Indicator) - GE Giraffe OmniBed Carestation CS1 Service Manual

Table of Contents

Advertisement

Chapter 6: Alarms and System Failure Codes
Alarms

6.1.18 Motor Drive Failure Alarm

Cause or Circumstances
The software has detected that
the elevating base motor or the
canopy motor was running
when not turned on.

6.1.19 Power Failure (LED Indicator)

Cause or Circumstances
No AC power into device
Mains power is present but the
5V power supply is not present.
6-10
Failure
Defective relay board
Failure
On/Off switch is not turned on.
Defective fuse F1 or F2 on relay
board
Defective power supply
Action
Replace the relay board.
Note:
Humidifier will not operate
during this failure.
Action
Verify cord is plugged into a live
wall outlet. Verify power switch/
circuit breaker on back of device is
turned on.
Check the 5 volt supply. If no 5 volts,
possible defective fuses on mains
to the power supply, check F1 and
F2 on relay board.
Check input to power supply for
mains voltage. If present and no 5
volts at output, then power supply
is defective.