Table - Canon OMNERA 500A Operator's Manual

Hide thumbs Also See for OMNERA 500A:
Table of Contents

Advertisement

Safety
Mechanical Safety

2.12.4 Table

WARNING!
Squeezing hazard can occur between the:
• table top (1) and the top of the imaging unit (4); when the table top (1) is in the outer
position (Y-direction) or moving in a longitudinal direction (X-direction).
• table top (1) and the imaging unit rail (6); when the table top (1) is in the outer
position (Y-direction).
• imaging unit (4) and the cover (5); when the imaging unit is moving in a longitudinal
direction (X-direction).
• column (7) and the footplate (8); when the column (7) is moving downwards (Z-
direction).
• cover (2) and the column cover foot (3); when the column (7) is moving downwards
(Z-direction).
Possible squeezing hazard areas and placement of warning labels:
0
1. Table top (X/Y/Z)
2. Cover
3. Column cover foot
4. Imaging unit (X)
OMNERA 500A
Z
Y
X
6
7
8
Fig. 2-10 Squeeze hazard
1000-095-071-EN Rev. 2.1
1
2
3
4
5. Cover
6. Imaging unit rail
7. Column (Z)
8. Footplate
Page 32
5
002010.eps
Operator´s Manual

Advertisement

Table of Contents
loading