Safety

Precautions, Safety

2.2 Precautions, Safety
WARNING!
No modification of this equipment is allowed.
WARNING!
The equipment is intended for use in radiographic examinations under the guidance
of trained health care professionals. Operating personnel must be familiar with the
equipment and the instructions given in this manual before using the equipment.
WARNING!
Safety devices must not be removed or modified. Any modification or removal will
immediately impair the safety.
WARNING!
All motorized movements shall be supervised by trained personnel.
WARNING!
Only medical-approved products shall be in the X-ray room.
Risk of electric shock to patient or user.
• No non-medical electrical devices shall be used in the x-ray room.
• Note that the monitor and the PC for the Image system, are non-medical approved
products.
WARNING!
Risk of electrical hazard or damage to the system
• Before cleaning or disinfection, switch off the system to prevent electric shocks,
for exceptions see
6.1.1 Cleaning and Disinfection Permitted with System Switched ON.
• Do not spray or pour cleaning liquid on any part of the system.
Use a lint-free cloth moistened with a moderate amount of liquid to avoid that
cleaning liquids seep into the openings of the system, e.g., air openings, gaps
between covers.
• Do not restart the system if cleaning liquids have leaked in.
OMNERA 500A
1000-095-071-EN Rev. 2.1
Page 12
Operator´s Manual