3.3.7 Xe lamp energy check

* If ESC key is pressed prior to ENT key, or if this status
continues for 5 minutes or longer, the system will return
automatically to the 1
(5)
The result of emission wavelength (Em) check will now be
displayed.
<Em side check result>
Em SIDE ACCURACY (254 nm)
GOOD -1nm
Press the ENT or ESC key.
* If the error is within ±3 nm, "GOOD" will appear, and if the
error exceeds ±3 nm, "FAIL" will appear on the display.
* If 5 minutes or more elapses without pressing the ENT or
ESC key, the system will return automatically to the 1st
monitor screen.
(6)
When the wavelength accuracy check is finished, the
system returns to the 1
Energy level of the xenon lamp can be checked using the xenon
lamp beam at 254 nm.
GLP
(1)
Press the
(2)
The following display will be presented.
<GLP initial screen>
GLP
(LOGBOOK=1, KEY LOCK=2,
4
WL CHECK=3, Xe CHECK=4)
Select Xe CHECK.
Press numeric key 4 and then the ENT key.
* If the Xe lamp is turned off, the system will return to the
st
1
monitor screen without performing an energy check.
(3)
The Xe lamp energy check screen appears and an energy
check will be performed.
3 - 25
st
monitor screen.
st
monitor screen.
key.