A success or failure message displays and the screen returns to the main
menu option.

TestMode Menu

The TestMode menu performs a test of all functions on the monitor. All
outputs are activated according to their failsafe mode, relay setting, and
buzzer setting. All functions are activated simultaneously for a maximum
of 3 minutes (or until the user cancels the test) and the 4-20mA output is
saturated at 20 mA to ensure working order. The monitor does not detect
gas during this TestMode procedure. Press enter to access the TestMode.
Use the arrow keys to scroll to Yes (or No) and press any key to
begin the test sequence.
To cancel the test, press enter at anytime when Stop displays.

Memory Menu

This menu is reserved for Honeywell Analytics authorized technicians' use
only. It contains no modifiable values or information for the user. Changing
values may damage the monitor.

Quit Menu

The Quit menu exits the programming functions and returns the screen to
its normal display. Press enter to exit the programming function.