9.21.2 Replace
bellows base latch
assembly
M1046983 04/08
To replace the latch assembly, you must disassemble the bellows base assembly to the
point where you can remove the guide (A) and latch assembly (B) as a unit.
1. Remove the Bag/Vent switch cartridge (Section 9.21.1).
2. Remove the two remaining screws (C) that hold the APL/BTV manifold to the bellows
base assembly. Remove the APL/BTV manifol.
3. To remove the guide/latch assembly, remove two mounting screws (D) from the
underside. Remove two additional mounting screws from the topside. Remove the
guide/latch assembly from the bellows base assembly.
4. Separate the latch assembly from the guide assembly.
5. To install the new latch assembly, put the spring (E) into place in the guide assembly
(long leg down).
6. Place the latch assembly on the guide assembly so that the latch engages the short
leg of the spring. Secure the latch assembly (F) to the guide assembly.
7. Mount the guide/latch assembly into the bellows base assembly.
• Extend the latch (G) while placing the assembly into the base.
8. Reassemble the breathing system in reverse order.
9. Perform the checkout procedure (Section 3).
D
A
E
F
9 Repair Procedures
B
C
G
9-53