Honeywell GA-SP01 User Manual
Honeywell GA-SP01 User Manual

Honeywell GA-SP01 User Manual

Motorized sampling pump

Advertisement

Quick Links

optimumstores
www.
.com/bw
Shop for BW products online at:

Advertisement

Table of Contents
loading

Summary of Contents for Honeywell GA-SP01

  • Page 1 optimumstores www. .com/bw Shop for BW products online at:...
  • Page 2 Limited Warranty & Limitation of Liability BW Technologies Ltd. (BW) warrants this product to be free from defects in material and workmanship under normal use and service for a period of two years, beginning on the date of shipment to the buyer. This warranty extends only to the sale of new and unused products to the original buyer. BW’s warranty obligation is limited, at BW’s option, to refund of the purchase price, repair, or replacement of a defective product that is returned to a BW authorized service center within the warranty period.
  • Page 3: Table Of Contents

    Table of Contents Title Page Introduction............................1 Contacting BW Technologies........................ 2 Safety Information - Read First ......................2 Getting Started ............................5 Activating the Sampler .......................... 9 Normal Operation .......................... 9 Operation............................... 9 Guidelines............................9 Standard Hose Configuration...................... 10 Sample Tube Connections......................10 Block the Airflow ..........................
  • Page 4 The Sampler Users Manual optimumstores www. .com/bw Shop for BW products online at:...
  • Page 5 List of Tables Table Title Page The Sampler International Symbols ......................4 The Sampler ........................6 LED Descriptions ....................... 7 Pushbuttons ........................8 Standard Hose Configuration ..................10 Replacing the Batteries....................16 Replacing a Filter ......................17 Draining the Water Trap ....................18 Troubleshooting Tips .......................
  • Page 6 The Sampler Users Manual optimumstores www. .com/bw Shop for BW products online at:...
  • Page 7 List of Figures Figure Title Page The Sampler ........................6 LED Descriptions ....................... 7 Standard Hose Configuration ..................10 Sample Tube Connections ....................10 Replacing the Batteries....................16 Replacing a Filter ......................17 Draining the Water Trap ....................18 Rear Assembly Screws....................20 optimumstores www.
  • Page 8 PERSONNEL ONLY. READ AND UNDERSTAND THE USER MANUAL COMPLETELY BEFORE OPERATING OR SERVICING. The Sampler Order Number Description Motorized sampling pump with rigid GA-SP01 probe and sample tubing Motorized sampling pump with rigid GA-SP02 probe and sample tubing (Europe) optimumstores www.
  • Page 9: Introduction

    The Sampler Introduction a Warning To ensure your personal safety, read Safety Information before you use the sampler. The Motorized Sampling Pump (“the sampler”) is a spot sampling device that can be used with most BW single and multiple gas detectors. An internal motorized pump transports potentially hazardous gasses from a sample area to a detector.
  • Page 10: Contacting Bw Technologies

    The Sampler Users Manual Contacting BW Technologies Safety Information - Read First To contact BW Technologies, call: Use the sampler only as specified in this manual, otherwise the protection provided by the sampler may be impaired. USA: 1-888-749-8878 Canada: 1-800-663-4164 International symbols used on the sampler and in this Europe: +44 (0) 1869 233004 manual are explained in Table 1.
  • Page 11 The Sampler Safety Information - Read First a Cautions ⇒ Warning: Substitution of components may impair Intrinsic Safety. ⇒ This product is not a gas detector. ⇒ Do not use the sampler if it is damaged. Before you use the sampler, inspect the case. Look for cracks or missing parts.
  • Page 12 The Sampler Users Manual Table 1. International Symbols Symbol Meaning Approved to both U.S. and Canadian Standards by the Canadian Standards Association ATEX Conforms to European ATEX directives European Explosive Protection Conforms to European Union directives optimumstores www. .com/bw Shop for BW products online at:...
  • Page 13: Getting Started

    The Sampler Getting Started To order replacement parts, see the section Replacement Getting Started Parts and Accessories. Airflow is continuously monitored and controlled by the The sampler is shipped with the short sample tube installed. sampler. It occasionally adjusts motor speed to compensate The Operation section describes how to install another for changes in environmental and sampling conditions and sample tube.
  • Page 14 The Sampler Users Manual Table 2. The Sampler Item Function Short sample tube (inlet) Block trigger Water Trap Drain plug Activation button C LEDs Audible alarm Cowl Tube fitting (outlet) Filter plug Figure 1. The Sampler optimumstores www. .com/bw Shop for BW products online at:...
  • Page 15 The Sampler Getting Started Table 3. LED Descriptions Item Function The green ON LED indicates the sampler is operating properly. The yellow BLOCK LED prompts you to block the airflow. The red FAIL LED indicates that the sampler is in an error condition. Figure 2.
  • Page 16 The Sampler Users Manual Table 4. Pushbuttons Pushbutton Description • To turn on the sampler, press C. • To initiate a calibration, confirm the sampler is on and then press C and hold for 6 seconds. There will be a beep, the sampler will turn off and upon the next startup the sampler will enter ON/OFF/CALIBRATION calibration mode.
  • Page 17: Activating The Sampler

    The Sampler Activating the Sampler Operation Activating the Sampler Guidelines ⇒ To activate the sampler, press C. When operating the sampler, adhere to the following All 3 LEDs will flash and the audible alarm will emit 4 beeps. guidelines: When the sampler is activated for the first time, it will enter calibration mode.
  • Page 18: Standard Hose Configuration

    The Sampler Users Manual Standard Hose Configuration Sample Tube Connections Table 5. Standard Hose Configuration The sampler is shipped with the short sample tube installed. Item Description To remove the installed sample tube, grasp the end of the tube and pull straight out. Insert the new sample tube Attachment hose through the enclosure and into the gasket opening.
  • Page 19: Block The Airflow

    The Sampler Deactivating the Sampler Block the Airflow Deactivating the Sampler When you block the airflow to the sampler you are confirming ⇒ To turn off the sampler, press C and hold for 2 seconds. that the sampler is working correctly and there are no leaks The audible alarm beeps 3 times, all LEDs light up and then in the system.
  • Page 20: Alarms

    The Sampler Users Manual Alarms Low Flow Alarm If the sampler detects a low flow situation, the red FAIL LED will flash and the audible alarm will beep twice. If the sampler detects the expected airflow it returns to normal operation. If the sampler does not detect the expected airflow it returns to the low flow alarm state until the airflow recovers.
  • Page 21: Low Battery Alarm

    The Sampler Alarms Low Battery Alarm Note: Once the Low Battery Alarm has been activated you have approximately 20 minutes before the sampler turns off. The sampler tests the batteries on activation and continuously thereafter. If the battery voltage is low, the sampler activates the Low Battery Alarm.
  • Page 22: Calibration

    The Sampler Users Manual Calibration Calibration Pass When the sampler passes calibration the audible alarm emits Upon startup, the audible alarm emits three beeps, the pump 3 beeps and the sampler enters normal operation mode. activates at a high speed and the green ON LED flashes. Calibration Fail If the sampler fails calibration, the audible alarm emits 6 beeps, the red FAIL LED lights and the sampler turns off.
  • Page 23: Maintenance

    The Sampler Maintenance Maintenance Replacing the Batteries a Warning To keep the sampler in good operating condition, perform the To avoid personal injury: following basic maintenance as required: ⇒ Replace the batteries as soon as the sampler • Keep an operations log of all maintenance and fault indicates a Low Battery Alarm.
  • Page 24 The Sampler Users Manual If the sampler is on, turn off the sampler before replacing the batteries. 1. Loosen but do not remove the #4-40 x 3/8” button head screw on the side of the sampler. The screw may not be included in certain regions.
  • Page 25: Replacing A Filter

    The Sampler Maintenance Replacing a Filter Table 7. Replacing a Filter Item Description The filter will need to be changed when it becomes clogged with dirt. Filter plug 1. Remove the filter plug assembly from the sampler. Filter Open end of filter 2.
  • Page 26: Draining The Water Trap

    The Sampler Users Manual Draining the Water Trap Periodically, the water trap will need to be emptied. Pull open the drain plug and shake any water out of the water trap. Replace the plug when complete. Table 8. Draining the Water Trap Item Function Water trap...
  • Page 27: If The Sampler Does Not Work

    The Sampler If the Sampler Does Not Work If the Sampler Does Not Work If you encounter a problem, calibrate the sampler. If this does not solve the problem follow the solutions listed in Table 9. If you still are unable to correct the problem, contact BW Technologies using one of the numbers on page 2 Table 9.
  • Page 28 The Sampler Users Manual Problem Possible Cause Solution A blocked state was not triggered during → Activate the sampler. During Sampler turns off with red FAIL LED calibration. flashing. calibration block the airflow of the sampler when prompted to by the yellow BLOCK LED.
  • Page 29: Replacement Parts And Accessories

    The Sampler Replacement Parts and Accessories Model No. Description Replacement Parts and Accessories M1806 Flexible sampling hose a Warning /0.3m To avoid personal injury or damage to the sampler, use SP-PF-1 Sampler particulate filters only the specified replacement parts. #4-40 x 3/8” button head M2437K To order parts or accessories listed in Table 10, contact screw...
  • Page 30: Specifications

    The Sampler Users Manual Maha Powerex 1700 mAh MH-AA170 192°C Specifications Maha Powerex 1800 mAh MH-AA180 201.1°C Operating Humidity: 5% to 95% relative humidity Yuasa Delta 1300 mAh DHA1400AA 209°C (non-condensing) Yuasa Delta 1500 mAh DHA1600AAC 204.4°C Sample Flow Rate: Typical 300 ml/min Worst Case Temperature Code: Sample Distance: 50ft/15.2m maximum recommended.
  • Page 31 optimumstores www. .com/bw Shop for BW products online at:...