Honeywell BW Clip Real Time Quick Reference Manual

Honeywell BW Clip Real Time Quick Reference Manual

2 or 3 year h2s, co, o2, so2 single-gas detector

Advertisement

Quick Links

BW Clip Real Time
2 OR 3 YEAR H
S, CO, O
, SO
2
2
BW Clip Real Time is a portable, single-gas detector manufactured by Honeywell. BW
Clip Real Time continuously displays the concentration of a specific gas in the ambient
environment, and activates an alarm when concentrations exceed alarm setpoints. It is
your responsibility to respond appropriately to the alarms.
BW Clip Real Time is a member of the GasAlert family of products. It is compatible
with IntelliDoX automatic test and calibration stations, as well as Safety Suite
Device Configurator software. For more information, visit the product website at
www.honeywellanalytics.com
Safety Information: Read First
WARNING
a
1. Substitution of components may impair Intrinsic Safety.
2. Honeywell recommends performing a bump test prior to each day's use to confirm
sensor response and alarm activation by exposing the detector to a concentration
of target gas that exceeds the low alarm set point. Honeywell also recommends
performing a bump test if the detector has been subjected to physical impact, liquid
immersion, an Over Limit alarm event, custody changes, or anytime the detector's
performance is in doubt. Calibrate the detector after a failed bump test. Always bump
test and calibrate in a fresh air environment.
3. Detectors in hibernation mode are not functional.
CAUTION
a
1. Activate the detector before the activation date on the package.
2. This product is a gas detector, not a measurement device.
3. Ensure that the sensor grill is free of dirt, debris, and is not obstructed.
4. Clean the exterior with a soft, damp cloth.
5. For optimal performance, periodically zero the sensor in a normal atmosphere
(20.9% v/v O
) that is free of hazardous gas.
2
6. Portable safety gas detectors are life safety devices. Accuracy of ambient gas
reading(s) is dependent upon factors such as accuracy of the calibration gas
standard used for calibration and frequency of calibration. Honeywell recommends
performing a calibration at least once every 180 days (6 months).
Informations de sécurité – À lire au préalable
AVERTISSEMENT
1. Le remplacement d'un composant de l'appareil peut compromettre la sécurité
intrinsèque du détecteur.
2. Avant chaque utilisation quotidienne, Honeywell recommande d'effectuer un test
fonctionnel afin de vérifier la réaction du capteur et l'activation de l'alarme, en
exposant le détecteur à une concentration de gaz cible supérieure au seuil d'alarme
basse. Honeywell recommande également d'effectuer un test fonctionnel si le
détecteur a été soumis à un impact physique, à une immersion dans du liquide, à un
événement d'alarme Dépassement de la limite, à des changements d'utilisateur ou
chaque fois que les performances du détecteur sont mises en doute. Étalonnez le
détecteur après l'échec d'un test fonctionnel. Effectuez toujours un test fonctionnel
et étalonnez dans un environnement avec de l'air frais.
3. Détecteurs en mode veille prolongée ne sont pas fonctionnels.
BW Clip Factory Calibration Certificate
This instrument was factory inspected, tested and calibrated in accordance with the conditions
and requirements of our registered Quality System, Operating Standards and Sales Agreements.
MISE EN GARDE
1. Activez le détecteur avant la date d'activation sur l'emballage.
2. Ce produit est un détecteur de gaz, pas un dispositif de mesure.
3. Veillez à assurer que la grille du capteur est exempt de saleté, les débris, et ne soit
pas vobstrué.
4. Nettoyez l'extérieur de l'appareil à l'aide d'un chiffon doux et humide.
5. Pour des performances optimales, zéro périodiquement le capteur dans une
atmosphère normale (20,9% v/v O2) qui est exempte de gaz dangereux.
6. Les détecteurs portables de gaz de sécurité sont des dispositifs de sécurité des
personnes. La précision des valeurs de gaz ambiant dépend de divers facteurs, tels
que la précision de la norme de gaz d'étalonnage utilisée pour l'étalonnage et la
fréquence d'étalonnage. Honeywell recommande d'effectuer un étalonnage au moins
une fois tous les 180 jours (6 mois).
BW Clip
Real Time Factory Calibration Certificate
This instrument was factory inspected, tested and calibrated in accordance with the conditions
and requirements of our registered Quality System, Operating Standards and Sales Agreements.
SINGLE-GAS DETECTOR
2
QUICK REFERENCE GUIDE
Getting Started
LED
LED
visual alarms
visual alarms
LCD
LCD
Audible
Audible
alarm
alarm
Bubble label
Bubble label
with product name
with product name
and gas type
and gas type
LCD and Symbols
Button prompt. When this symbol is displayed, press and hold the
button until the symbol disappears.
BW Clip Real Time O
measured as percent by volume.
BW Clip Real Time H
concentration is measured in parts per million.
Low alarm and high alarm. These symbols are displayed when a gas
concentration that exceeds alarm setpoints is detected.
Non-compliance warning. This symbol is displayed when a
diagnostic or compliance test fails; when the operating life
countdown reaches 24 hours or less; and for up to 30 days after
operating life expiry. It is also displayed when certain notifications
are enabled through Safety Suite Device Configurator via an
IntelliDoX station.
Real-time clock. This symbol is displayed when real-time clock is
enabled through Safety Suite Device Configurator via an IntelliDoX
station.
End of operating life warning. This symbol is displayed when the
operating life countdown reaches 24 hours or less.
Peak reading. This symbol is displayed when a maximum gas
exposure event has occurred within the past 24 hours.
Bump test reminder. If the detector is configured to display a bump
test reminder, this symbol is displayed when a bump test is due.
indicates a failed bump test.
This symbol is displayed when more than 24 hours have elapsed
since the last successful zero procedure was performed (O
models) or indicates a sensor calibration is due (H
models).
indicates a failed calibration (H
failed zero procedure (O
Operating life remaining in months (MM) or days (DD).
Infrared port
Button
Button
Sensor and
Sensor and
sensor grill
sensor grill
Product label
models only. Oxygen concentration is
2
S, CO and SO
models only. Toxic gas
2
2
S, CO and SO
2
models).
2
50115903-101 EN B2
Alligator
clip
2
S, CO and SO
2
2
models) or a
2

Advertisement

loading

Summary of Contents for Honeywell BW Clip Real Time

  • Page 1 BW Clip Real Time is a member of the GasAlert family of products. It is compatible with IntelliDoX automatic test and calibration stations, as well as Safety Suite Device Configurator software. For more information, visit the product website at Alligator www.honeywellanalytics.com...
  • Page 2 An alarm is initiated when the sensor is exposed to a gas concentration that exceeds Configurator via an IntelliDoX station. BW Clip Real Time flashes, vibrates and beeps. alarm setpoints. The alarm persists until the alarm gas concentration returns to an While an event or setting is displayed, press the button to scroll to the next available acceptable range.
  • Page 3 Hib is displayed. 5. While Hib is displayed, press the button until a 5 second By default, BW Clip Real Time is configured to use the following calibration gas mixtures: countdown is displayed. When the countdown is complete, the 20-second Hibernation counter is displayed.
  • Page 4 IntelliDoX automatic test and calibration station. Liquid crystal display (LCD) Sensor Type Ratings and Certifications Electrochemical cells The BW Clip Real Time is in conformity with Battery the following standards: Lithium, non-replaceable UL 913 8th Edition Ingress Protection...