Product Maintenance - Omega OM-CP-RFCURRENT2000A User Manual

Wireless current data logger
Table of Contents

Advertisement

OM-CP-RFCURRENT2000A

Product Maintenance

Battery Replacement
Materials: OM-CP-BAT103
1. On the bottom of the enclosure, open the battery compartment by pulling in on the cover tab.
2. Remove the battery by pulling it from the compartment.
3. Install the new battery, taking note of the polarity.
4. Push the cover closed until it clicks.
Recalibration
Standard recalibration for the OM-CP-RFCURRENT2000A-3A are two points, 0A and 1.00A. For the
OM-CP-RFCURRENT2000A-20mA, the standard calibration points are 0mA and 25V. For the OM-CP-
RFCURRENT2000A-160mA the standard calibration points are 0mA and 100mA.
Recalibration is recommended annually for any Omega data logger; a reminder is automatically displayed in
the software when the device is due.