Page 1
P R E L I M I N A R Y (APRIL 29, 2008) E_BT03_COVER.FM GE Healthcare Voluson® 730Expert Service Manual • Voluson® 730Expert systems with Serial number A09000 onwards • Voluson® 730Expert systems with Software version SW 3.x.x (BT03) •...
Page 3
• WARTEN SIE DIESES GERÄT NUR, WENN SIE DIE ENTSPRECHENDEN WARNUNG ANWWEISUNGEN IM KUNDENDIENST-HANDBUCH GELESEN HABEN UND (DE) NACHVOLLZIEHEN KÖNNEN. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
Page 4
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL • ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, LA TRADUCCIÓN ES RESPONSABILIDAD DEL CLIENTE.
Page 5
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL VAROITUS (FI) ΠΡΟΕΙ∆ΟΠΟΙΗΣΗ (EL) FIGYELMEZTETÉS (HU) • ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU. • EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU.
Page 6
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL VÝSTRAHA (CS) • DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK. • HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE.
Page 7
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL ĮSPĖJIMAS (LT) • DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK. • HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE.
Page 8
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL ОСТОРОЖНО! (RU) UPOZORNENIE (SK) • DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA. • OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA...
Page 9
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL (JA) (ZH-CN) (KO)
Page 10
GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
Page 11
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Healthcare. GE Healthcare - Kretztechnik may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
Page 12
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL Revision History Revision Date Reason for change September 12, 2003 Initial Release (from Software Version 3.0.x onwards) December 11, 2003 from Software Version 3.0.3 onwards March 08, 2004 from Software Version 3.1.x onwards...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL Table of Contents CHAPTER 1 Introduction Overview ............1 - 1 Purpose of Chapter 1 .
Page 14
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL CHAPTER 2 Site Preparation Overview ............2 - 1 Purpose of Chapter 2 .
Page 15
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL CHAPTER 3 Setup Instructions Overview............3 - 1 The Purpose of Chapter 3 .
Page 16
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL Scanner Power On ........3 - 22 Back-end Processor Boot Up .
Page 17
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL CHAPTER 4 Functional Checks Overview ............4 - 1 Purpose of Chapter 4 .
Page 18
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL Average Velocity (Manual Trace) ..... . 4 - 22 Average Velocity (Auto Trace) .
Page 19
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL CHAPTER 5 Components and Functions (Theory) Overview............5 - 1 Purpose of Chapter 5 .
Page 20
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL ECG Preamplifier ........5 - 18 MOD (Magneto-Optical Drive) .
Page 21
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL Mechanical Concept and Overview ......5 - 41 Major Functions of CPN6 .
Page 22
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL CHAPTER 6 Service Adjustments Overview ............6 - 1 Purpose of Chapter 6 .
Page 23
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview............7 - 1 Purpose of Chapter 7 .
Page 24
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL CHAPTER 8 Replacement Procedures Overview ............8 - 1 Purpose of Chapter 8 .
Page 25
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL Tools ..........8 - 17 Backup all Exams of the “old”...
Page 26
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL CHAPTER 9 Renewal Parts Overview ............9 - 1 Purpose of Chapter 9 .
Page 27
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL CHAPTER 10 Care & Maintenance Overview............10 - 1 Periodic Maintenance Inspections .
Page 28
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL Definition ..........10 - 16 Generic Procedure .
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the Voluson® 730Expert scanner.
Hospital’s Service Personnel • Contractors (Some parts of Chapter 2 - Pre-Installation) 1-1-4 Voluson® 730Expert Models Covered by this Manual Table 1-2 Voluson® 730Expert Model Designations GE Part Number Kretz # Description H48621N ---------- Voluson® 730Expert Main Body (BT03) 1-1-5 System History - Hardware and Software Versions This manual applies to Voluson®...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 1-2 Important Conventions 1-2-1 Conventions Used in Book Icons Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment.
Page 34
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION "Mains OFF" Indicates the power off position of the mains power Rear of system at mains switch (F1) switch.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 1-3-4 Electrical Safety 1-3-4-1 Safe Practices To minimize shock hazard, the equipment chassis must be connected to an electrical ground.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 1-3-7 Lockout/Tagout Requirements (For USA Only) Follow OSHA Lockout/Tagout requirements to protect service personnel from injuries caused by unexpected energizing or start-up of equipment during service, repair, or maintenance.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 1-4-2 Compliance The Voluson® 730Expert unit conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient requirements.
Singapore 169730 Phone: 81-426-48-2944 Japan Support Center Fax: 81-426-48-2905 1-5-2 System Manufacturer Table 1-7 System Manufacturer Manufacturer Telephone GE Medical Systems - Kretztechnik GmbH & Co OHG Tiefenbach 15 +43-7682-3800-0 +43-7682-3800-47 A-4871 Zipf Austria 1-12 Section 1-5 - Customer Assistance...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Chapter 2 Site Preparation Section 2-1 Overview 2-1-1 Purpose of Chapter 2 This chapter provides information required to plan and prepare for installation of a Voluson® 730Expert ultrasound unit.
Electrical Requirements NOTE: GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size Ground wire from the distribution panel to the Ultrasound outlet.
The Voluson® 730Expert is supplied with a main power plug, as standard. In the event that the unit arrives without a power plug, or with the wrong plug, contact your GE dealer. When necessary, the installation engineer will supply the appropriate power plug to meet the applicable local regulations.
RF gaskets touch metal. appropriate location. The interconnect cables are grounded and require ferrite beads and other shielding. Use GE- specified Cable length, material, and routing are all important; do not make any changes that do not meet harnesses and peripherals.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 2-2-4 Probe Environmental Requirements Operation: Ambient temperature 18° to 30° C Storage: -10° to 50° C NOTE: Temperature in degrees C. Conversion to degrees F = °C (9/5) + 32).
(and the accompanying electrical installations) are highly sophisticated and special engineering competence is required. All electrical work on these products must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
Mandatory Site Requirements NOTE: GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 2-3-3-1 Recommended Ultrasound Room Layout Figure 2-1 below shows a floor plan illustrating the recommended layout of the Ultrasound Room and depicting the minimal room layout requirements.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 2-3-4 Networking Pre-installation Requirements 2-3-4-1 Purpose of the DICOM Network Function DICOM (Digital Imaging and Communications in Medicine) services provide the operator with clinically useful features for moving images and patient information over a hospital network.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL This page was intentionally left blank. 2-10 Section 2-3 - Facility Needs...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Chapter 3 Setup Instructions Section 3-1 Overview 3-1-1 The Purpose of Chapter 3 This chapter contains information needed to setup the Voluson® 730Expert unit. Included are procedures to receive, unpack and configure the equipment.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-2-2 Installation Warnings 1.) Since the Voluson® 730Expert weighs approximately 136 kg (300 lbs.) without options, two people are required to unpack it. Two people are also required for installing any additional items in excess of 16 kg / 35 pounds.(e.g., Monitor).
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-2-2-4 Brake Pedal Operation REMEMBER: If the front wheels are engaged for transportation, pressing the release WARNING WARNING brake pedals (brakes on front wheels under the foot rest) once disengages the lock.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment CAUTION Transport only with forklift or stracker truck. During transport pay attention to the point of gravity (“tilt and drop” indicator)! !! CAUTION: Have two people available to unpack the Voluson®...
Page 55
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 3-4 Unpacking Procedure Step Task Open front panel by removing the screws from the side panels. Remove probes and accessories from the wooden box.
Section 3-4-2 "System Voltage Settings" on page 3-7 NOTE: Report any items that are missing, back-ordered, or damaged, to your GE Healthcare - Kretztechnik sales representative. The contact address is shown in Contact Information on page 1-11. Section 3-4 - Preparing for Set Up...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-4-2 System Voltage Settings Verify that the scanner is set to the correct voltage. The Voltage settings for the Voluson® 730Expert Scanner is found on the identification plate, on the rear of the system.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 3-5 Connection of Auxiliary Devices 3-5-1 Preparations NOTE: Normally the auxiliary devices and peripherals come already installed with the system.
Page 59
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-5-1 Preparations (cont’d) Table 3-6 below outlines Voluson® 730Expert hardware installation procedures described in the sub- sections. Table 3-6 Connection Procedures...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-5-5 Global Modem Connection Figure 3-9 Global Modem Connection Scheme Never install the telephone wiring, or use the modem during an electric storm;...
Mitsubishi HS-MD3000 NOTE: There are two types of Mitsubishi HS-MD3000 VCR (PAL and NTSC) approved by GE Healthcare - Kretztechnik. Use the standard VCR type (PAL or NTSC) that is suited to your region. ONLY the specific GE - Kretztechnik Version of this video recorder type may be used in...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-5-6-2 Sony SVO-9500MD Figure 3-11 Sony - Video Recorder Connection Scheme Set DIP Switches on back of the VCR as shown:...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-5-7 B/W Video Printer Connection Figure 3-13 B/W Video Printer Connection Scheme Set DIP Switches on back of the Black/White printer as shown in the image (all DOWN).
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-5-8 Line Printer Connection Figure 3-14 Line Printer Connection Scheme Please observe that the Line Printer has to be located outside of the patient environment CAUTION (acc.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-5-9 Digital Color Printer Connection Figure 3-15 Digital Color Printer Connection Scheme Pay attention to lateral distances. See Instruction Manual of the printer!
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-5-10 Bluetooth Printer Connection NOTE: Please refer to connection scheme belonging to the used B/W Printer type: • Section 3-5-10-1 "HP 5600/5900 Series" on page 3-18 •...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-5-10-2 Canon Pixma MP600 / MP610 Canon Bluetooth Printer Adapter Figure 3-17 Canon Bluetooth Printer - Connection Scheme Please observe that the complete Bluetooth Printer Assembly has to be located outside of the CAUTION patient environment (acc.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-5-11 External USB-Devices Do not connect or disconnect any external USB-devices to or from the system while WARNING WARNING scanning a patient! The appearing dialogs could distract you from the scan!
Page 71
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-5-11-2 External USB-Devices - Disconnection (cont’d) The „Unplug or Eject Hardware“ dialog shows all USB-devices that are connected to the system.
The connection of the Voluson® 730Expert ultrasound unit to a power source should be performed by a qualified person who has completed basic Voluson® 730Expert System User Training. Use only the power cords, cables and plugs provided by or designated by GE Healthcare - Kretztechnik to connect the unit to the power source.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-6-2-2 Back-end Processor Boot Up Press the ON/OFF Standby switch left below the control panel. ON/OFF Standby switch Figure 3-23 ON/OFF Standby Switch...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-6-3 Power Off / Shutdown After turning off a system, wait at least 10 seconds before turning it on again.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-6-4 Transducer Connection Connect a transducer to one of the three rightmost transducer receptacle as follows: 1.) Inspect the probe and probe socket to verify that it is free of debris.
Section 3-7-4 "Adjustment of Printer Settings" on page 3-36 The Bluetooth Printer Connection set as well as the Color Deskjet printer MUST NOT be installed CAUTION by the user! !! CAUTION: For installation please contact your local distributor or GE service representative. 3-26 Section 3-7 - Printer Installation...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-7-1 Installing Line Printer HP 990cxi or HP 995c 1.) Power off/Shutdown the system as described in: Section 3-6-3 on page 3-24.
Page 78
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-7-1-1 Install the HP printer software/driver (cont’d) 2.) Click the NEXT buttons to start the Hardware Wizard and to locate the driver files.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-7-2 Installing Digital Color Printer Sony UP-D21MD or UP-D23MD 1.) Power off/Shutdown the system as described in: Section 3-6-3 on page 3-24.
Page 80
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-7-2-1 Install the UP-D21MD / UP-D23MD printer software/driver (cont’d) 3.) Use the BROWSE button to search the following path (see:...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-7-3 Printer Installation manually 1.) On the Touch Panel, press UTILITIES. 2.) In the Utilities menu, touch SYSTEM SETUP to invoke the setup desktop on the screen.
Page 82
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-7-3 Printer Installation manually (cont’d) 6.) Click the ADD PRINTER button. A warning message appears: Please read this message carefully and click YES if you have skills to do this.
Page 83
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-7-3 Printer Installation manually (cont’d) 9.) Select the corresponding Printer Port (e.g., Figure 3-36 = USB001) and click NEXT. Figure 3-36 Select Printer Port 10.)In the following window select the HAVE DISK button.
Page 84
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-7-3 Printer Installation manually (cont’d) Figure 3-38 Select Printer Driver path (C:\Utilities\PrinterDriver\..) 12.)Click OPEN, select the “xxx.inf” file and click OPEN again.
Page 85
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-7-3 Printer Installation manually (cont’d) 15.)Assign a name, decide if the printer should be used as default printer and confirm with NEXT.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-7-4 Adjustment of Printer Settings 1.) After system restart, touch the UTILITIES key, and then SYSTEM SETUP on the Touch Panel.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-7-4-1 HP 990cxi / HP 995c - Printer Settings 1.) Call up the ‘Printer Properties’; operation see: Section 3-7-4 "Adjustment of Printer Settings"...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-7-4-2 UP-D21MD / UP-D23MD - Printer Settings 1.) Call up the ‘Printer Properties’; operation see: Section 3-7-4 "Adjustment of Printer Settings"...
Page 89
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-7-4-2 UP-D21MD / UP-D23MD - Printer Settings (cont’d) 3.) Select the page. From the “Color Adjust” pop-up menu select: GRAPHICS a.) Color Balance: Cyan = 0;...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-7-5 Remote Control Selection To assign an auxiliary device (e.g., printer) to the remote keys PRINT A and/or PRINT B, or to adjust Foot switch, etc.:...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 3-8 System Configuration 3-8-1 System Setup Modifications of system parameters are supported by diverse dialog pages and windows on the system setup desktop: •...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-8-1-2 How to enter Date and Time Select the page in the System Setup see: Figure 3-48. GENERAL 1.) Select the “Date Format” (only one can be active).
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-8-2 On-Board Optional Peripherals Mains outlets: Mains socket ST1, ST2, ST3, ST4, ST5 for accessories. All mains outlets are co-switched by the unit’s mains switch via built-in isolation transformer.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 3-13 COMP VIDEO OUT Connector Pin No Signal Description 1 (Center Pin) Signal 2 (Coax Pin) Signal GND Table 3-14...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 3-11 Network IP Address Configuration NOTE: Following Information must be provided by customer or hospital engineer before you can start: A Station name, AE Title, IP address and Port Number for the Voluson®...
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 3-11-1 Map Network Drive 1.) Select the MAP NETWORK DRIVE button (in the System Setup - Network page, see: Figure 3-50) to open a dialog where the system can be connected to a shared network drive of another server.
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 3-12 Connectivity Setup Worksheet Site System Information Comments: Floor: Site: Dept: Room: Type: Serial #: REV: CONTACT INFORMATION Name...
Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. Chapter 3 - Setup Instructions...
Page 102
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL This page was intentionally left blank. 3-52 Section 3-13 - Paperwork...
Page 103
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the Voluson® 730Expert scanner diagnostics by using the built-in service software, and power supply adjustments.
Page 104
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place.
Page 105
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-3-1-1 Scanner Power On (cont’d) 5.) Press the ON/OFF Standby switch left below the Control Panel. ON/OFF Standby switch Figure 4-2 ON/OFF Standby Switch...
Page 106
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-3-3 System Features 4-3-3-1 Control Panel Figure 4-3 Control Panel Tour 1.) Touch Panel screen 2.) Touch Panel digipot and toggle switch controls 3.) Power-, Audio Volume- toggle switch controls...
Page 107
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-3-3-2 Touch Panel Figure 4-4 Touch Panel - Main Menu 1.) Main menu key: to change from one Sub menu to another.
Page 108
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-3-3-3 Monitor Display 9 10 User program 7.5-5.0 Pwr 85% Gn -8 C6 / M7 P4 / E2 STATUS BAR AREA...
Page 109
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 4-4 Functional Checks For a basic functional check of the system’s different modes, following pages will familiarize you with image optimization for: •...
Page 110
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-1 2D Mode Checks Figure 4-6 2D Main and 2D Sub Menu Table 4-3 2D Mode Functions Step Task Expected Results...
Page 111
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 4-3 2D Mode Functions Step Task Expected Results Advantage of the Trapezoid Mode: The scan area is very increased in...
Page 112
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-2 Additional (optional) Operating Modes 4-4-2-1 B-Flow Check B-Flow On/Off switch Press the BF key to activate/deactivate the B-Flow mode.
Page 113
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-3 M Mode Checks Figure 4-9 M Main and M Sub Menu Table 4-4 M Mode Functions Step Task Expected Results Cursor Position Adjust the M Cursor position with the TRACKBALL in the 2D Single image.
Page 114
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-4 Spectral Doppler Mode Checks NOTE: Different menus are displayed depending on which Spectral Doppler Mode (PW or CW) is selected.
Page 115
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 4-5 Spectral Doppler Mode (PW, CW) Functions Step Task Expected Results SCALE To select the displayed measuring unit (in relation to the zero-line).
Page 116
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 4-6 Color Doppler Mode (CFM, PD, TD) Functions Step Task Expected Results Used to eliminate Doppler “noise” that is caused by vessel wall or cardiac wall motion.
Page 117
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-6 Volume Mode Checks Real Time 4D, RT_4D_Biopsy, DiagnoSTIC, VCI and VOCAL are Options. NOTICE If these options are not part of the system configuration, the appendant checks can be omitted.
Page 118
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 4-7 Pre-Volume Mode Functions Step Task Expected Results AQUISITION: • 3D STATIC 3D Volume Mode - Static volume acquisition (also in combination with PD, HD-Flow or CFM) •...
Page 119
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-6-2 Functions after the 3D Acquisition Figure 4-13 3D Static - Sectional Planes and Image Rendering Table 4-8 Functions after the 3D Acquisition...
Page 120
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-6-3 Sub Menus Figure 4-14 3D/4D Sub Menu Table 4-9 Sub Menus Step Task Expected Results To select the desired Render View Direction.
Page 121
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-7 Using Cine 4-4-7-1 Activating Cine Press FREEZE, then roll the TRACKBALL to display the images of the stored sequence one by one.
Page 122
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-8 Basic Measurements NOTE: Different menus are displayed depending on which Mode is selected. General remarks to perform Basic Measurements: •...
Page 123
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-8 Basic Measurements (cont’d) 4-4-8-2 Circumference/Area (using Ellipse or Trace) Measurements 1.) Press the CALIPER key once. 2.) Touch the corresponding item on the Touch Panel. An active cursor displays.
Page 124
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-8 Basic Measurements (cont’d) 4-4-8-4 Velocity Measurements (Spectral Doppler Mode) NOTE: The Spectral Doppler image is displayed based on time (X-axis) and velocity (Y-axis).
Page 125
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-9 Calculations NOTE: Confirm that the patient information is correct and the probe and application are selected properly. General remarks to perform Calculations: •...
Page 126
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-10 Probe/Connectors Usage 4-4-10-1 Connecting a probe 1.) Place the probe's carrying case on a stable surface and open the case.
Page 127
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-11 Image Management (Sonoview) For Sonoview - Image Management functionality refer to Chapter 15 in the Basic User Manual of Voluson®...
Page 128
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-12 Using the optional MOD (Magneto-Optical Drive) The MOD (Magneto-Optical Drive) is an Option. NOTICE !! NOTICE: The optional 3.5 inch Magneto-Optical disk drive supports the following medias: 1.3GB;...
Page 129
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-12-1 Formatting Media To erase/format the backup media, DVD/CD+(R)W or MOD, press the SONOVIEW key on the Control panel. The Sonoview screen appears on the monitor; see: Figure 4-15 on page 4-25.
Page 130
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-4-12-1 Formatting Media (cont’d) By selecting DVD/CD the unit displays the “Erase DVD/CD” window as shown in Figure 4-19. 1.) Select the “Erase Mode” and click the [OK] button to start the process.
Page 131
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 4-5 Backup and Restore Database, Preset Configurations and Images From Software Version 3.1.x onwards, it is also possible to save/load/delete “Full Backup” data via the...
Page 132
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-5-1 Save User Settings Only (Application Settings) The User Settings contains: • Application Settings • User Programs • Auto Text •...
Page 133
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-5-2 Load User Settings Only (Application Settings) The loading procedure overwrites the existing application settings on the local hard drive. Make sure to CAUTION insert the correct DVD/CD or MO.
Page 134
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-5-2 Load User Settings Only (Application Settings) (cont’d) 7.) Select the Complete Backup (marked blue; see: Figure 4-25) and click the >> button to copy the Complete Backup into the Load Data field.
Page 135
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-5-3 Save Full Backup (Presets, Configurations & Application Settings) A full backup always contains the following data • Patient demographic and exam data (database containing the patient data and measurements) •...
Page 136
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-5-3 Save Full Backup (Presets, Configurations & Application Settings) (cont’d) When the Full Backup is stored on a network drive (to map a network drive see: Section 3-11-1 "Map...
Page 137
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-5-4 Load Full Backup (Presets, Configurations & Application Settings) Always backup any data before an upgrade; see: Section 4-5-3 on page 4-33.
Page 138
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-5-4 Load Full Backup (Presets, Configurations & Application Settings) (cont’d) 1.) On the Touch Panel, touch UTILITIES. 2.) In the Utilities menu, touch SYSTEM SETUP to invoke the setup desktop on the screen.
Page 139
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-5-4 Load Full Backup (Presets, Configurations & Application Settings) (cont’d) 8.) Click the NEXT button and then select YES to start, or NO to cancel the restore procedure.
Page 140
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-5-6 Archiving Images 1.) Press the SONOVIEW key on the control panel. 2.) Insert the DVD/CD+(R)W or MOD into drive.
Page 141
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 4-6 Software Configuration Checks Touch the UTILITIES key, then touch SYSTEM SETUP. The System Setup desktop offers different pages to check:...
Page 142
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 4-11 Measurement Setup Checks Step Task Expected Result(s) OB: Check “Show Author´s Name at Measure Menu” setting assigned as desired by the customer Cardiac: Check “2D Circumference and Area Method”...
Page 143
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 4-8 Mechanical Function Checks 4-8-1 Rotation of the Control Console locking lever for locking and unlocking the control console...
Page 144
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 4-9 Site Log 4-9-1 Site Log - System (Service Database) 1.) On the Touch Panel, press UTILITIES. 2.) In the Utilities menu, touch SYSTEM SETUP to invoke the setup desktop on the screen.
Page 145
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4-9-2 Site Log - Paper Documentation Table 4-16 Site Log Date Service person Problem Comments Chapter 4 - Functional Checks...
Page 146
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL This page was intentionally left blank. 4-44 Section 4-9 - Site Log...
Page 147
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains Voluson® 730Expert’s system concepts, component arrangement, and subsystem function.
Page 148
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-2 General Information Voluson® 730Expert is a digital beamforming curved-, linear- and phased array ultrasound imaging system. It has provisions for analog input sources like ECG and Phono.
Page 149
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-2 General Information (cont’d) Monitor without GEU Figure 5-2 Major System Components Major System Components: • GEF: Main Board Chassis:...
Page 150
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-2 General Information (cont’d) Monitor User-Interface-Module Trolley (housing, mechanical framework). GEM- External IO ECG+Diskdrive The following components Module are part of GW:...
Page 151
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-2 General Information (cont’d) The Voluson® 730Expert used digital beamforming technology which provides high resolution and high penetration performance. It is a general purpose, mobile, software controlled diagnostic ultrasound scanner.
Page 152
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-2-1 Description of Voluson® 730Expert Operating Modes 5-2-1-1 B-Mode or 2D-Mode B-Mode or 2D-mode is a two-dimensional image of the amplitude of the echo signal. It is used for location and measurement of anatomical structures and for spatial orientation during operation of other modes.
Page 153
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-2-1-3-2 Power Doppler A real-time two dimensional cross-section of blood flow is displayed. The 2D cross-section may be presented as a rectangle, parallelogram, trapezoid, sector, or a full circle, depending on the particular transducer used.
Page 154
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-2-2-1 3D Data Collection and Reconstruction (cont’d) By changing the position of the cross-section, a variety of views of the underlying structure are obtained, and these views can be used to understand a 3-dimensional structure in the body.
Page 155
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-2-3 Block diagram Voluson® 730Expert Figure 5-4 Top Level Architecture Chapter 5 - Components and Functions (Theory)
Page 156
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-2-3 Block diagram Voluson® 730Expert (cont’d) Figure 5-5 Voluson® 730Expert - Block diagram 5-10 Section 5-2 - General Information...
Page 157
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-2-4 Data Flow Control Description This section describes the functions of the Voluson® 730Expert Boards vs different Operation Modes. •...
Page 158
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-2-4-1-1 Special B-Mode Techniques a.) THI = HI (Tissue Harmonic Imaging incl. Coded Harmonics) In one method of THI the RX-Frequency is doubled, so that the radial resolution is increased due to the higher RX-Frequency.
Page 159
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-2-4-3 D-Mode (Pulsed Wave- and Continuous Wave Doppler) 1.) CRS PRF-generator; see: 5-2-4-1 B-Mode After DC-cancelling the signal is mixed with RX-Frequency and brought to LF-Spectrum, where the LowPassFilter cuts HF.
Page 160
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-2-4-5 CFM-Mode (Color Flow Mode) 1.) CRS PRF-generator; see: 5-2-4-1 B-Mode After DC-cancelling the signal is mixed with RX-Frequency and brought to LF-Spectrum, where the LowPassFilter cuts HF.
Page 161
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-2-4-11 Sonoview write mode (store Image to Sonoview) 1.) CRS Not used 2.) CPR Not used 3.) CKV - DMA section Not used 4.) CKV - Video section...
Page 162
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-2-5 Description of Software Options To activate the software options, see Section 8-8 "Replacement or Activation of Options" on page 8-15.
Page 163
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-2-5-4 VOCAL - Virtual Organ Computer-aided Analysis Diagnosis and therapy of cancer is one of the most important issues in medical care.
Page 164
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-2-6 Description of Hardware Options Table 5-3 Hardware Options HW-Options Description CW-Doppler 5-2-6-1 CW - Continuous Wave Doppler ECG Digital Module...
Page 165
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-2-6-5-1 Location in the Unit Figure 5-7 Modem (placed on the GEF-box) 5-2-6-5-2 LEDs The Modem has 10 LEDs on its front.
Page 166
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-3 Main board Chassis GEF Module The GEF Module contains the Front End processor and the Back End processor and the Secondary Power supply for the full GEF Chassis.
Page 167
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-4 FrontEnd Processor Figure 5-10 Front End Processor - Block diagram Chapter 5 - Components and Functions (Theory) 5-21...
Page 168
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-4-1 FrontEnd - Board Descriptions 5-4-1-1 CPU - Probe Connector Board • 1 CW-Probe Connector • 3 Probe-Connectors 260pin •...
Page 169
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-4-1-3 CPD - Beamformer Sub-board 1x : Beamformer-ASIC generates TX-Pulses + TX-Focus +TX-Apodization, Rx Focus Delay and Summation 16x : TX-Power Amplifier...
Page 170
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-4-1-4 CRW - CW-Doppler Board (optional) • Separate Clock Oscillator + Clock generator PRF, FN • TxDelayMatrix • Transmitter, •...
Page 171
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-4-1-7 CRS - Signal Processor Board A.) MID-Processor: 1.) For B, C, D modes: Multi Beam-DeInterleave Subtraction Filter (for THI)
Page 172
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-4-1-8 CPP - Power Supply Secondary Board + Motor Power stage used for Supply of both FrontEnd and BackEnd DC-DC-Converter: 59VDC to following output voltages: + 3.3V, +/-5V, +12V, +/-15V,...
Page 173
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-5 BackEnd Processor Figure 5-14 BackEnd Processor - Block diagram Chapter 5 - Components and Functions (Theory) 5-27...
Page 174
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-5-1 Block diagram CKV DMA Controller PCI Bus 32 Bit / 33MHz PLX PCI 9054 DMA 0 DMA 1 Host PC...
Page 175
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-5-2 BackEnd - Board Descriptions 5-5-2-1 SBC - Single Board Computer Built in or external Components: AGP VGA 100 Mbit LAN USB 2.0...
Page 176
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-5-2-7 CPP - Power Supply Secondary Board + Motor Power stage used for Supply of both FrontEnd and BackEnd DC-DC-Converter: 59VDC to following output voltages: + 3.3V, +/-5V, +12V, +/-15V,...
Page 177
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-6 Internal I/O StdbySwitch (User PowerON/OFF) GES - External IO Board GEM - Disk Drive Module SUPPLY DVD +...
Page 178
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-6 Internal I/O (cont’d) Figure 5-17 Connectors on the Main Module of the GEF Item Connector Name Description VGA1...
Page 179
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-6 Internal I/O (cont’d) Figure 5-18 Connectors on PC-part of the GEF Item Connector Name Description USB-GEM Connector for Disk Drive Module GEM...
Page 180
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-7 Top Console The Voluson® 730Expert Operator Control Panel (OCP) consists of the following electronic subassemblies and/or functional components: •...
Page 181
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-7 Top Console (cont’d) Figure 5-20 Voluson® 730Expert - Control Console Chapter 5 - Components and Functions (Theory) 5-35...
Page 182
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 5-5 Voluson® 730Expert - key codes key code Description Voluson® 730Expert - Functionality call-up of the patient data entry menu...
Page 183
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 5-5 Voluson® 730Expert - key codes key code Description Voluson® 730Expert - Functionality left trackball key Trball_b upper trackball key...
Page 184
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-8 Monitor RIGHT LEFT KEY FUNCTION a. MENU KEY This button will enable the On Screen Display. This button is also used to select the function in the Main Menu or to save the setting in the Sub Menu.
Page 185
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-9 External I/O Figure 5-22 External I/O Panel Connectors Table 5-6 External I/O Connector Description Item Connector Name Description...
Page 186
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-10 Peripherals 5-10-1 General Information - GEM (Disk Drive Module) The GEM Module contains the DVD/CD+(R)W Drive. Additionally the Magneto Optical Drive and the ECG-preamplifier (MAN6) can be installed as an Option.
Page 187
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-11 Power Distribution Power Distribution depends on the currently installed CPN - Primary Power Module. NOTICE • Section 5-11-1 "CPN6 - Primary Power Module" on page 5-41 !! NOTICE: •...
Page 188
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-11-1-3 Fuses on Rear Panel (CPN6) F1: The main Input voltage is fused by a magnetic Circuit breaker (Rated current 16A) built in the Main Power switch labelled F1 F2:The AC Output voltage (115/230V) is fused by F2 (magnetic Circuit breaker 2.5/5A)
Page 189
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-11-2 CPN80-81 - Primary Power Module Figure 5-26 Primary Power Module - CPN80-81 (old version) Note: The new version of CPN80-81 has different print layout! It shows an electrical hazard icon on the right side (instead of the “Caution”...
Page 190
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-11-2-3 Fuses on Rear Panel (CPN80-81) F1 + F2: The main Input voltage is fused by two 16 Ampere fuses (rated current 16A) labeled F1 and F2.
Page 191
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-12 Mechanical Descriptions 5-12-1 Physical Dimensions 560,5 607,6 Figure 5-29 Physical Dimensions Chapter 5 - Components and Functions (Theory)
Page 192
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-12-2 Monitor • Tilt: 11° forwards and backwards • Swivel: +/-90° rotation. 5-12-3 Top Console Positioning 30° Figure 5-30 Top Console Positioning...
Page 193
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-12-5 Assembly Drawing GW & GEU & Monitor Figure 5-32 Assembly Drawing Voluson® 730Expert Chapter 5 - Components and Functions (Theory)
Page 194
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-13 Air Flow Control 5-13-1 Air Flow Distribution The fans at the right side of the Main Board Chassis draw air into the scanner, through the filter grid, and pushes it through the card rack.
Page 195
Introduction The Service Platform will increase service productivity and reduce training and service costs. This web-enabled technology provides linkage to e-Services, e-Commerce, and the iCenter, making GE’s scanners more e-enabled than ever. The Service Platform contains: • specific software/hardware test modules, system setup, update, etc. for Voluson systems; see: Section 5-15 "Service Page"...
Page 196
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-14-3 iLinq Interactive Platform Many of the services of the Service Platform come from its integration with iLinq. If an iLinq checkout was performed on the Voluson® 730Expert, the ILINQ button is available in the “Utilities”...
Page 197
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 5-15 Service Page 5-15-1 Introduction The Service Page contains specific software/hardware test modules, system setup, update, etc. for Voluson systems only.
Page 198
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 4.) Enter the password and click the ACCEPT button to display the Service Tools window. Auto Tester Monitor Test Printer...
Page 199
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-15-3-2 Service Viewer Provides common information about System Temperature, Probes, Working hours of system components and probes. 1.) Select the SERVICE VIEWER button to get access to the E-Service page.
Page 200
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 1.) Click the MAINTENANCE REPORT button. The following message box will be displayed. Figure 5-40 Maintenance Report 2.) Fill in the requested information and click OK.
Page 201
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 5-15-3-6 Monitor Test 1.) To perform the Monitor test, select the MONITOR TEST button. The following message appears on the screen.
Page 202
3-7-3 "Printer Installation manually" on page 3-31. Only accessories explicitly recognized and released by the system manufacturer WARNING WARNING GE Healthcare - Kretztechnik may be used in connection with the system. !! WARNING: 5-15-3-8 Update 5-15-3-8-1 Reserved for loading Native Language Support. Not for use in the field.
Page 203
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of Chapter 6 This chapter describes how to test and adjust the mechanical capabilities of a scanner that may be out of specification.
Page 204
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 6-3 Monitor Adjustment The Monitor can be tilted or rotated. • Tilt angle: up/down 11° • Rotation angle: right/left 90°...
Page 205
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 6-3-1-1 Degauss Degaussing refers to the process of removing magnetic-field effects from the monitor. Operation of the monitor within a magnetic field may adversely effect color purity. Degaussing can be used to correct the problem.
Page 206
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 6-3-1-3 Contrast and Brightness Adjusting the monitor´s contrast and brightness is one of the most important factors for proper image quality.
Page 207
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 6-4 Control Console, Transport Lock 6-4-1 Control Console The control console can be rotated 30° to the right. When rotating the control console grasp it only the front grip.
Page 208
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 6-5 Trackball Adjustment Adjustment of the mechanical movement may be necessary to ensure smooth running of the trackball. 1.) Remove the outer fixing ring by turning it counterclockwise.
Page 209
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 6-6 Daylight Saving Time (DST) - New Dates Issue: The automatic Daylight Saving Time feature in Microsoft Windows products may no longer run on the correct date, causing the system’s time to be incorrect.
Page 210
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL This page was intentionally left blank. Section 6-6 - Daylight Saving Time (DST) - New Dates...
Page 211
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
Page 212
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 7-2 Collect Vital System Information The following information is necessary in order to properly analyze data or images being reported as a...
Page 213
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 7-2-1 Needed Data - in case of System crash or Errors In case of a system crash, or errors that are probably software bugs, provide as much detail information NOTICE as possible before contacting the OLC (On-Line Center).
Page 214
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 7-3 Check Points Voltages 7-3-1 How to check power Open the locking, flap the service lid on the back of theVoluson® 730Expert upwards, and check the status of the Diagnostic LED´s.
Page 215
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 7-4 Screen Captures and Logs There may be times when the customer or field engineer will want to capture a presentation on the screen so it may be recovered by the OnLine Center.
Page 216
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 7-4-2-1 Export System Data Every time an error message like Figure 7-4 is produced, a dump-file and a text file containing the error dump and the error message are created in D:\export.
Page 217
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 7-5 How to use the Auto Tester program 1.) Touch UTILITIES and SYSTEM SETUP on the Touch Panel. 2.) Select the page on the screen.The “password window”...
Page 218
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 7-5 How to use the Auto Tester program (cont’d) The following message box appears. Pause key on the right upper corner...
Page 219
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 7-5 How to use the Auto Tester program (cont’d) The following screen will appear. “DVD/CD Burn” icon Figure 7-8 Auto Tester Finished 10.)Insert an empty DVD/CD+(R)W in the Drive and select the “DVD/CD Burn”...
Page 220
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 7-6 Minimum Configuration to Boot/Scan 7-6-1 Minimum Configuration to Scan Following cables must be connected to scan; see: Figure 7-11 a.) PS2 (connector for Mouse/Keyboard)
Page 221
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 7-7 Troubleshooting Trees and Instructions 7-7-1 System does not Power On / Boot Up Is the scanner plugged Make sure the...
Page 222
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 7-7-2 Noise disturbs the Image Noise is shown in Image. Is there any electrical devices that affect the Scanner? Keep the Scanner away from those electrical devices.
Page 223
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 7-7-3 Trackball The trackball is low sensitive! Does the trackball move at all? Remove the ball from the Trackball unit, and clean it.
Page 224
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 7-7-4 System Does Not Power Off / Shutdown System has been switched off system switched off Primary Power (CPY System starts with the automatically in approx.
Page 225
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 7-7-5 Monitor Troubleshooting Fault symptom Check these items Check the power cord is properly connected. CPN6 only: Check the Power Switch of Peripherals (F2) on back of system is set to the “ON”...
Page 226
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 7-7-6 Unable to Record to VCR Unable to record to VCR! Is the VCR properly connected? Check that all signal, and power supply cable connections to the VCR unit are OK.
Page 227
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 7-7-7 Printer Troubleshooting Unable to use Printer(s)! Is the printer(s) properly connected? Check that all signal, and power supply cable connections between Printer and Scanner are OK.
Page 228
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 7-7-7 Printer Troubleshooting (cont’d) Still unable to use the Printer(s)! Is printing unavailable even by manual operation? Check the signal cables between GEF and Backpanel.
Page 229
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 7-7-8 DVD/CD+(R)W Troubleshooting (DVD/CD Drive) 1.) Insert an empty DVD/CD+(R)W into the Drive. 2.) Enter “Sonoview” by pressing the SONOVIEW key on the control panel; see Figure 7-19.
Page 230
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 7-7-10 Network Troubleshooting 7-7-10-1 No Connection to the Network at All 1.) Check that the network cable between the scanner and the wall network is connected and well seated in both ends.
Page 231
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 7-8 Error Messages If the problem (error message) still exists after performing the described actions, call technical support. NOTICE...
Page 232
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Error Messages Actions B-DynContrast Out Of Range press ok and save this user-setting once again BF: can't set BM RxApod restart the system BF: can´t set C Rx Apod...
Page 233
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Error Messages Actions CPS_InterfaceToHW::vSet_VOL_MODE -- incorrect value for restart the system Mode_4D CtrlMvGrpRepresentations::isInWriteMode not supported for restart the system mode nr xx...
Page 234
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Error Messages Actions Enhance Out Of Range press ok and save this user-setting once again Ensemble Out Of Range press ok and save this user-setting once again Error in CreateCineImage try to store again;...
Page 235
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Error Messages Actions File Pos load volume files from other storage medium File Read Error load volume files from other storage medium...
Page 236
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Error Messages Actions M_Gain Out Of Range press ok and save this user-setting once again M_Reject Out Of Range press ok and save this user-setting once again...
Page 237
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Error Messages Actions Overflow restart the system Persistance Out Of Range press ok and save this user-setting once again Persistence coeff page index too big...
Page 238
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Error Messages Actions RepresentationManager::vCreate: Nr. Dim 1 incorrect, see restart the system vSetReplayParams() call RepresentationManager::vCreate: Nr. Dim 2 incorrect, see...
Page 239
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Error Messages Actions Start ET too small restart the system Start frequency too small restart the system StoragePool: Not sufficient replay storage for B & CF Mode...
Page 240
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Error Messages Actions ViewerConMngr::vAssembleCF_DFE: attempt to get ECG_Consumer_2D- or ECG_Draw_2D-Block from restart the system ECGViewer Objects failed!! ViewerConMngr::vAssembleM_DFE: attempt to get...
Page 241
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter contains replacement procedures for different modules and their subsystems.
Page 242
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 8-1-2 Returning/Shipping Probes and Repair Parts Equipment being returned must be clean and free of blood and other infectious substances.
Page 243
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 8-6-3 Preparations (cont’d) 1.) Touch the UTILITIES key on the Touch Panel once to display the Utilities menu. 2.) Touch SYSTEM SETUP in the Utilities menu.
Page 244
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 8-6-3 Preparations (cont’d) 5.) Select the page on the System Setup desktop screen. NETWORK a.) Click on the DICOM / SONOVIEW CONFIGURATION button in the page.
Page 245
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 8-2-5 Software - Installation Procedure (via Service Page) 1.) Insert the System DVD or Upgrade-CD into the drive. 2.) Touch UTILITIES and SYSTEM SETUP on the Touch Panel.
Page 246
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 8-2-5 Software - Installation Procedure (via Service Page) (cont’d) 6.) Confirm the following message with YES. Figure 8-7 Update UIS message 7.) A new window pops up on the screen and the InstallShield starts extracting files.
Page 247
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 8-2-5 Software - Installation Procedure (via Service Page) (cont’d) 9.) Complete the Software installation with the FINISH key. Figure 8-10 V730 Software Setup The system restarts automatically.
Page 248
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 8-3 User Settings Only (Application Settings) Loading Procedure 8-3-1 Introduction The User Settings contains: • Application Settings • User Programs •...
Page 249
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 8-5 Service Platform (SSW) Upgrade Procedure 8-5-1 Manpower One Person, 10 min. 8-5-2 Tools System DVD or Upgrade CD...
Page 250
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 8-5 Service Platform (SSW) Upgrade Procedure (cont’d) 8.) Select the DVD/CD DRIVE button 9.) Browse for the install.bat file and click OK.
Page 251
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 8-6 Electronic User Manual (EUM) Upgrade Procedure 8-6-1 Manpower One Person, 15 min. 8-6-2 Tools System DVD or Upgrade CD...
Page 252
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 8-6-4 EUM - Upgrade Procedure 1.) Click the EUM button. The “Update Software” dialog appears. 2.) Select the DVD/CD DRIVE button 3.) Browse for the EUMSetup_en_V730Expert.exe file and click OK.
Page 253
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 8-6-4 EUM - Upgrade Procedure (cont’d) 7.) Click the EUM button again to install the new EUM. 8.) Select the MO DRIVE or the DVD/CD DRIVE button, depending on the storage medium you use 9.) Browse for the EUMSetup_en_V730Expert.exe file again and click OK.
Page 254
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 8-7 Software and Functional Checks after the Upgrade 1.) Press the UTILITIES key on the Touch Panel once to display the Utilities menu.
Page 255
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 8-8 Replacement or Activation of Options Following Options are available: • Real Time 4D • DICOM • RT_4D_Biopsy •...
Page 256
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 8-9 Transfer of Patient Database and Images from System-to-System 8-9-1 Introduction From Software Version 3.1.x onwards, it is possible: NOTICE •...
Page 257
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 8-9-2 Transfer of Patient Database and Images via Sonoview 8-9-2-1 Introduction This section describes how to transfer the complete patient database and images from one system (= “old”...
Page 258
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 8-9-2-4 Backup all Exams of the “old” system (cont’d) 7.) Click on the close button and then select the SETTINGS icon on the left side of the screen.
Page 259
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 8-9-2-5 Restore all Exams (of the “old” system) to the “new” system (cont’d) The folder which is selected to be renamed (e.g., A09008, that contains the backup data from a different...
Page 260
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 8-9-2-5 Restore all Exams (of the “old” system) to the “new” system (cont’d) 11.)Select the desired action: The exam on the hard disk is replaced with the exam in the backup.
Page 261
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 8-10 Replacement of the Monitor Task Lamp 8-10-1 Manpower One person, 15 min. 8-10-2 Tools Philips screwdriver 1 and 2, stubby (length ~30mm)
Page 262
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 8-11 Replacement of the Probe Holder Replacement of the probe holder depends on the used trolley (GW). NOTICE • GW130, not to be replaced by the user •...
Page 263
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 8-12 Replacement of the Trackball top fixation ring 8-12-1 Manpower One person, 5 min. 8-12-2 Trackball top fixation ring - Replacement Procedure 1.) Remove the fixation ring by turning it counterclockwise.
Page 264
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL This page was intentionally left blank. 8-24 Section 8-13 - Replacement of Digipots and TGC Slider controls...
Page 265
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Spare Parts available for the Voluson® 730Expert (BT03).
Page 266
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 9-2 List of Abbreviations • CKV - Video Converter Board • CPD - Sub-Board on Beamformer (CPR) • CPE - Motherboard Extension (Backpanel I/O-Card) •...
Page 267
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 9-3 Parts List Groups Figure 9-1 Console Views Table 9-2 Mechanical and user accessible parts Item Part Group Name...
Page 268
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 9-4 Housing (GW) and additional Console Hardware GW130 GW140 Figure 9-2 Housing (GW) Section 9-4 - Housing (GW) and additional Console Hardware...
Page 269
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 9-3 Housing (GW) Item Part Name Part Number Description Housing with wheels, backpanel with connectors KTZ154695 and cables GW130 Trolley Voluson® 730Expert...
Page 270
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 9-4 Housing (GW) and additional Console Hardware (cont’d) 8 pieces each Figure 9-3 Additional Console Hardware Table 9-4 Additional Console Hardware...
Page 271
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 9-5 User Interface (GEU Top Console) 203 204 212 213 210 211 Figure 9-4 User Interface (GEU) Chapter 9 - Renewal Parts...
Page 272
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 9-5 User Interface (GEU) Item Part Name Part Number Description Voluson® 730Expert Front Handle KTZ220039 Front Handle User Interface (only used with GEU80 - KTZ154693) Voluson®...
Page 273
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 9-6 Monitor + Monitor Replacement Parts Figure 9-5 Monitor + Monitor replacement parts Table 9-6 Monitor + Monitor replacement parts...
Page 274
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 9-7 Disk Drives (GEM) 301 302 305 306 307 308 Figure 9-6 Disk Drives (GEM) Table 9-7 Disk Drives (GEM)
Page 275
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 9-8 Main Power Module (CPN) 414 415 410 411 412 Figure 9-7 Main Power Module (CPN) Chapter 9 - Renewal Parts...
Page 276
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 9-8 Main Power Module (CPN) Item Part Name Part Number Description CPN6 Power Supply Module KTZ195905 CPN6 Power Supply Module...
Page 277
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 9-9 Main Board Module (GEF) Figure 9-8 Main Board Module (GEF) Table 9-9 Main Board Module (GEF) Item Part Name...
Page 278
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 9-9-1 FrontEnd (US-Part) Figure 9-9 FrontEnd (US-Part) Table 9-10 FrontEnd (US-Part) Item Part Name Part Number Description KTZ195929 Beam former Sub-board (CPR81.P11 only) CPD81.P17 Beam former SUB-Board...
Page 279
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 9-9-1 FrontEnd (US-Part) (cont’d) Figure 9-10 FrontEnd (US-Part) cont´d Table 9-11 FrontEnd (US-Part) cont´d Item Part Name Part Number Description...
Page 280
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 9-9-2 BackEnd Processor (PC-Part) 575 576 Figure 9-11 Back Processor (PC-Part) 9-16 Section 9-9 - Main Board Module (GEF)
Page 281
CPE80.P5 Motherboard - Extension electrical Signal- and Supply-Connection between all KTZ195699 CPM3.P2 Motherboard PC-Plug-In-Boards Hard disk IDE 80GB; Voluson 730Expert/Pro (BT03) Hard disk IDE 80GB V730Expert/Pro KTZ208800 (obsolete - replaced by Universal HDD KTZ196003) (BT03) Universal Hard disk (can replace KTZ208800)
Page 282
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 9-10 Options and Upgrades Table 9-13 Options and Upgrades Item Part Name Part Number Description bootable DVD for System HDD recovery (SW3.1.6) System/Boot DVD (SW3.1.6)
Page 283
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 9-11 Miscellaneous Cables Table 9-14 Miscellaneous Cables Item Part Name Part Number Description Set contains two adapters: • Audio Adapter Line-IN/OUT...
Page 284
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 9-14 Miscellaneous Cables Item Part Name Part Number Description 12V/5V-Power Supply-Distributor-Cable (2x) for PC-Component (yellow-black, red-black) KTZ212426 Cable HDD Y-Power...
Page 285
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 9-14 Miscellaneous Cables Item Part Name Part Number Description External Rear Panel with electrical Signal- and Supply- Connection-Cables to the Voluson® 730Expert Main- Unit (internal) Rear-Panel.
Page 286
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 9-14 Miscellaneous Cables Item Part Name Part Number Description Data Cable for (optional) Global Modem KTZ212199 Modem Data Cable...
Page 287
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 9-14 Miscellaneous Cables Item Part Name Part Number Description Power Cord United Kingdom 240V KTZ212441 Power Cord UK Power Cord USA Hospital Grade KTZ212402 Power Cord USA (Hosp.grade)
Page 288
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 9-14 Miscellaneous Cables Item Part Name Part Number Description Set contains two parts: USB for PC-Slot, Connector on Backpanel.
Page 289
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 9-14 Miscellaneous Cables Item Part Name Part Number Description VGA cable HD15-HD15 KTZ212275 VGA cable HD15-HD15 Set contains two cables:...
Page 290
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 9-12 Optional Peripherals and Accessories 806 807 814 815 816 817 Figure 9-12 Optional Peripherals and Accessories 9-26 Section 9-12 - Optional Peripherals and Accessories...
Page 291
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 9-15 Optional Peripherals and Accessories Item Part Name Part Number Description KTZ211332 Analog B/W Video Printer Analog B/W Video Printer NTSC/PAL...
Page 292
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 9-12 Optional Peripherals and Accessories (cont’d) Figure 9-13 Optional Peripherals and Accessories cont´d Table 9-16 Optional Peripherals and Accessories cont´d...
Page 293
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 9-17 System Manuals Item Part Name Part Number Kretz # Description KTZ105899 105899 Voluson® 730Expert Service Manual System User Manuals Basic User Manual, Voluson®...
Page 294
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 9-13 Probes 9-13-1 2D-Probes Figure 9-14 2D curved array Transducers Table 9-18 2D curved array Transducers Item Part Name...
Page 295
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 9-13-1 2D-Probes (cont’d) Figure 9-15 2D linear- and phased array Transducers Table 9-19 2D linear- and phased array Transducers Item...
Page 296
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 9-13-2 Real-Time 4D Volume Probes Figure 9-16 Real-Time 4D Volume Probes Table 9-20 Real-Time 4D Volume Probes Item Part Name...
Page 297
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 9-13-3 CW-Pencil Probes Figure 9-17 CW- Pencil Probes Table 9-21 CW- Pencil Probes Item Part Name Part Number Description single element Continuous Wave (CW) Doppler pencil probe with a KTZ195540 PCW4.0...
Page 298
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 9-14 Biopsy Needle Guides Figure 9-18 Biopsy Needle Guides 9-34 Section 9-14 - Biopsy Needle Guides...
Page 299
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 9-22 Biopsy Needle Guides Item Part Name Part Number Description reusable Biopsy needle guide for probe IC5-9 PEC42 H46721F needle diameter: <...
Page 300
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL This page was intentionally left blank. 9-36 Section 9-14 - Biopsy Needle Guides...
Page 301
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Chapter 10 Care & Maintenance Section 10-1 Overview 10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your Voluson® 730Expert system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory.
Page 302
Testing results, corrective action and the effects of corrective action must be documented and maintained on the site. Your GE service representative can help you with establishing, performing and maintaining records for a quality assurance program. Please contact us for coverage information and/or price for service.
Page 303
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 10-3-1 How often should care & maintenance tasks be performed? (cont’d) Table 10-2 Customer Care Schedule Item Service at Indicated Time...
Page 304
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Table 10-2 Customer Care Schedule Item Service at Indicated Time Notes Also after corrective • Console Leakage maintenance or as required by Current Checks your facilities QA program.
Page 305
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 10-4 Tools Required 10-4-1 Special Tools, Supplies and Equipment 10-4-1-1 Specific Requirements for Care & Maintenance Table 10-3 Overview of Requirements for Care & Maintenance...
Page 306
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary.
Page 307
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 10-5-2 Functional Checks (see also Chapter 4) The functional checks take about 60 minutes to perform. Refer to the system user documentation whenever necessary.
Page 308
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 10-5-2-2 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted.
Page 309
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 10-5-5 Physical Inspection Table 10-9 Physical Checks Step Item Description Labeling Verify that all system labeling is present and in readable condition.
Page 310
The Basic User Manual provides a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. Review the Basic User Manual of Voluson®...
Page 311
To guarantee that the sensitive probe head does not hit the bottom of Endocavity soaking cup kit RIC-Holder is available at GE Accessory the container, place the probe in the RIC Holder during the disinfection (Order Nr. : E8385TA ) procedure.
Page 312
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 10-6 Using a Phantom Refer to the User Manual of the Phantom for information on using a phantom and quality assurance tests.
Page 313
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 10-7-2 GEMS Leakage Current Limits The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards.
Page 314
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 10-7-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
Page 315
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 10-7-4 Grounding Continuity Electric Shock Hazard. The patient must not be contacted to the equipment during this test CAUTION !! CAUTION: Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case.
Page 316
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 10-7-5 Chassis Leakage Current Test 10-7-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break.
Page 317
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 10-7-5-3 Data Sheet for Chassis Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-11.
Page 318
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 10-7-6-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit Off and On. For each combination, the operating controls such as the lead switch should be operated to find the worst case condition.
Page 319
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 10-7-8 Isolated Patient Lead (Sink) Leakage-Isolation Test Reference the procedure in the IEC 60601-1. When using the Dale 600, switch the meter’s function selector to the LEAD-ISO.
Page 320
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 10-7-8-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less than the value shown in the table below.
Page 321
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 10-7-9 Probe Leakage Current Test 10-7-9-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
Page 322
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 10-7-9-3 No Meter Probe Adapter Procedure Figure 10-6 Check Without Probe Adapter Follow these steps to test each transducer for leakage current.
Page 323
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL 10-7-9-4 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than values shown in...
Page 324
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL Section 10-8 When There's Too Much Leakage Current... 10-8-1 Chassis Fails Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent.
Page 325
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL ULTRASOUND INSPECTION CERTIFICATE Customer Name: System ID: Dispatch Number / Date Performed: Warranty/Contract/HBS System Type Model Number: Serial Number: Manufacture Date:...
Page 326
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R DRAFT (A 29, 2008) IRECTION EVISION PRIL ERVICE ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source Leakage Current - Probe...
Page 327
Control Description, 5-11 Top Level Architecture, 5-9 Data needed Voluson 730Expert, 5-10 in case of System crash or Errors, 7-3 Voluson 730Expert Basic, 5-4 Daylight Saving Time (DST), 6-7 Bluetooth Printer Delete Connection Scheme, 3-18 Full Backup (Presets, Config. & Appl. Settings), 4-37...
Page 328
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL Digital Color Printer 2D Mode, 4-8 Connection Scheme, 3-17 Additional Operating Modes, 4-10 Installation UP-D21MD, 3-29 Basic Measurements, 4-20 Installation UP-D23MD, 3-29 Brakes, 4-41 Disconnection...
Page 329
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL Physical Dimensions, 2-5 Power Distribution, 5-41 Label Locations, 1-9 CPN6 - Primary Power Supply, 5-41 Line Printer CPN80-81 - Primary Power Supply, 5-43 Connection Scheme, 3-16...
Page 330
GE H ® 730E (BT03) EALTHCARE RETZTECHNIK OLUSON XPERT 105899, R IRECTION EVISION ERVICE ANUAL S-VHS Video Recorder Specifications, 3-46 Connection Scheme, 3-13 Voltage Settings, 3-7 System Components, 5-3 Volume Modes System Configuration Functional Checks, 4-15 How to change Language, 3-42...