GE Vivid T8 Basic Service Manual page 68

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

Site Preparations
2-2-3-6 Power stability requirement
IEC 61000-4-11
Voltage dips, short
interruptions and
voltage variations
on mains supply
2-8
Table 2-4:
Power stability requirement
< 5%T (> 95% dip) for
0.5 cycle;
40%T (60% dip) for 5
cycles;
70%T (30 dip) for 25
cycles;
< 5%T (>95% dip) for 5
sec
< 5%T (> 95% dip) for
0.5 cycle;
40%T (60% dip) for 5
cycles;
70%T (30 dip) for 25
cycles;
< 5%T (>95% dip) for 5
sec
Vivid T8/Vivid T8 Pro – Basic Service Manual
5718826-100 English Rev. 10
Mains power
quality should be
that of a typical
commercial or
hospital
environment.