Site circuit breaker
CAUTION
CAUTION
Site power outlets
Unit power plug
LOGIQ V2/LOGIQ V1 – Basic Service Manual
5610739-100 English Rev.10
General Ultrasound system requirements
Power outage may occur. The LOGIQ V2/LOGIQ V1 requires a
dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you do not
have any other equipment operating on the same circuit.
It is recommended that the branch circuit breaker for the
Ultrasound system be readily accessible.
POWER OUTAGE MAY OCCURE.
The LOGIQ V2/LOGIQ V1 requires a dedicated single branch
circuit. To avoid circuit overload and possible loss of critical
care equipment, make sure you DO NOT have any other
equipment operating on the same circuit.
A dedicated AC power outlet must be within reach of the
Ultrasound system without extension cords. Other outlets
adequate for the external peripherals, medical and test
equipment needed to support this Ultrasound system must also
be present within 1 m (3.2 ft.) of the Ultrasound system.
Electrical installation must meet all current local, state, and
national electrical codes.
If the Ultrasound system arrives without a power plug, or with
the wrong plug, you must contact your GE dealer or the
installation engineer must supply what is locally required.
2-7