P&M (Probe And Mux) Board Removal Procedure; Removing The Screws That Secure The P&M Board To The Rfi Board; Loosening The Screws From The P&M Board - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FL091021, R
IRECTION
8-3-1-4
P&M (Probe and MUX) Board Removal Procedure
1) Remove the Control Panel and Keyboard, as described in the
Removal Procedure" on page
2) Loosen and remove the four screws that secure the P&M Board to the RFI Board, as shown in
Figure
Figure 8-22 Removing the Screws that Secure the P&M Board to the RFI Board
3.) Loosen the two screws that fasten the P&M Board to the Bottom Assembly, as shown in
Figure 8-23
5
EVISION
8-22.
below.
Figure 8-23 Loosening the Screws from the P&M Board
Chapter 8 - Replacement Procedures
8-6.
V
S
IVID I N
ERVICE
"Control Panel and Keyboard
M
ANUAL
8-25