7-4-3

Radio Frequency Interface (RFI) Diagnostic Tests

7-4-3-1
Accessing the RFI Diagnostic Test Options
Access the Diagnostic Test window, as described in the Accessing the Front End Diagnostic
Options section, on page 7-25.
7-4-3-2
Full RFI Test
NOTE:
For a description of this test, refer to the information provided in
IMPORTANT In order to access an individual board test, it is necessary to insert the service dongle from
NOTICE
system boot-up and enter the Service Password.
NOTE:
When performing this test, no probe should be connected to the system and nothing should be touching
the probe connectors!
1.) In the Diagnostic Test window, trackball to Full RFI Test in the list of RFI tests available for selection
from the Diagnostic Test Tree (see
displayed under Execute; the relevant system component to be tested is highlighted light blue in the
Schematic View (RFI), as shown below:
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
1
EVISION
Figure 7-11 Full RFI Test
Chapter 7 - Diagnostics/Troubleshooting
V
IVID I N AND
below) and press Set. The Full RFI Test name is
Figure 7-11
V
N S
IVID Q
ERVICE
Full RFI Test on page 7 -
M
ANUAL
11.
7-31