Gas chromatograph
Table of Contents

Advertisement

Connection
Connection
Connection
Connection
Sample gas streams 1 – 5
Inlet pressure1-4 barg (15-57 PSIg)
CAL Calibration Gas
1-4 barg (15-57 PSIg)
Carrier 1: HE
Helium
5-6 barg (71-86 PSIg)
Recommended pressure 5.5 barg)
Sample and blocking vent
(SBV)
Carrier 2:
AR
Argon
5-6 barg (71-86 PSIg)
Recommended pressure 5.5 barg
Purge vent
(PV)
In the standard version, they take the form shown in the figure below.
Figure 6.21: Layout of the gas connections
Each gas inlet consists of the coupling (a 1/8" Swagelok pipe fitting) and an adapter
with an internal 2
filter.
µ
the connector ) to a 1/16" pipe and connected to the internal sample gas system. This
pipe is used as a flame arrester and must not be reduced in size. It must be more than
25 cm (19") in length and its maximum internal diameter may be 0.015".
EnCal 3000 Hardware Manual
Position on the device/Note
Position on the device/Note
Position on the device/Note
Position on the device/Note
Right-hand sideIt is possible to measure up to 5 different
gases in sequence.
Left-hand side, first connection
Used to maintain measurement accuracy.
Left-hand side, second connection
For transporting the measurement gas inside the device;
waste gas is released into the housing.
Left-hand side, third connection The unmeasured sample
gas is released at ambient pressure.
Left-hand side, penultimate connection
Second carrier gas only needed to analyse biogas or
hydrogen in natural gas. In standard applications this
connection is not used and closed without piping inside
Left-hand side last connection
Flushing or bypass gas is released at a high flow rate.
Inside the device, the gas connections are each soldered (at
73022344 o