Use the steps below to change the reference values for moisture
inputs. (See Figure 2-14 on page 2-26 for a menu map.)
If you modified the supplied cables or are not using GE
Panametrics-supplied cables for moisture probes, you
may be required to adjust the moisture reference data
as described in Performing an MH Calibration/
Adjustment Test in Chapter 1 of the Service Manual.
IMPORTANT: You should record reference data on the Program
Information List supplied in Appendix A, or on a
separate sheet of paper, and keep it in a safe place.
1. At the Reference Menu (see Table 2-4 on page 2-16), press the
PROBE menu key until the Moisture Reference Table appears.
See Figure 2-8 below.
MOISTURE REFERENCE TABLE
Ch
HIGH
1
1
2
3
4
5
6
AutoCal Interval: (HH.MM) 01:00
Reference Menu
HELP
INTERVAL
Figure 2-8: Reference Menu for Moisture
2. Compare the data on the screen to the reference data printed on the
label placed on the side or front of the unit, or supplied with a
replacement channel card. If the data is incorrect, use the
procedure below to correct it. If the data is correct, do one of the
following:
continue to enter reference values for oxygen cell and pressure
inputs as described in the following sections.
press the DONE menu key until Main Menu appears on the
message line; then refer to Displaying Measurements on
page 2-21. When you exit, the Series 1 automatically calibrates
each channel.
Caution!
LOW
PROBE
DONE