GE Moisture Image Series 1 Startup Manual

Hygrometer
Hide thumbs Also See for Moisture Image Series 1:
Table of Contents

Advertisement

Quick Links

GE Moisture Image Series 1 Startup Manual

Advertisement

Table of Contents
loading

Summary of Contents for GE Moisture Image Series 1

  • Page 1 Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...
  • Page 2 Moisture Image Series 1 Hygrometer Startup Guide Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 3 June 2004 P r o c e s s C o n t r o l I n s t r u m e n t s M o i s t u r e I m a g e S e r i e s 1 H y g r o m e t e r S t a r t u p G u i d e 9 1 0 - 1 0 8 U B 3...
  • Page 4 June 2004 Moisture Image Series 1 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 5 GE Panametrics. Fuses and batteries are specifically excluded from any liability. This warranty is effective from the date of delivery to the original purchaser. If GE Panametrics determines that the equipment was defective, the warranty period is: •...
  • Page 6 June 2004 Table of Contents Chapter 1: Installation Introduction ..............1-1 Unpacking the Series 1 .
  • Page 7 June 2004 Table of Contents (cont.) Chapter 2: Initial Setup Introduction ..............2-1 Using the Front Panel to Access Data .
  • Page 8 Chapter 1 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 9 Installation Introduction ......... . . 1-1 Unpacking the Series 1 .
  • Page 10 April 2004 Introduction Users typically install the Moisture Image Series 1 as part of a complex process system, which includes components such as filters, pumps, and pressure regulators. In such an environment, probes and other parts of the system may be subjected to environmental hazards, such as high temperature, pressure extremes, corrosive elements and mechanical vibrations.
  • Page 11 April 2004 Checking the Delta F Before connecting the Delta F oxygen cell(s), you must check it for damage and/or leakage. Depending on the application, the oxygen Oxygen Cell for cell may have a top drain or both a top and bottom drain for the Leakage electrolyte reservoir.
  • Page 12 April 2004 Checking the Delta F 4. Let the oxygen cell stand for about 6 hours; then check for any leakage. Oxygen Cell for Leakage (cont.) 5. If there is no leakage, drain the cell completely. If the cell leaks, see the warranty information at the beginning of this manual.
  • Page 13 April 2004 Choosing a Site You should have discussed environmental and installation factors with a GE Panametrics applications engineer or field sales person by the time you receive the meter. The equipment should be suited to the application and installation site.
  • Page 14 Do not install any other components, such as filters, ahead of the probes or sample system unless instructed to do so by GE Panametrics. • Observe all normal safety precautions. Use the probes within their maximum pressure and temperature ratings.
  • Page 15 A sintered stainless steel shield is used to protect the sensor from high flow rates and particulate matter. The end cap should not be removed except upon advice from GE Panametrics. The sensor has been designed to withstand normal shock and vibration.
  • Page 16 April 2004 Static or Dynamic Use The sensor performs equally well in still air or where considerable flow occurs. Its small size makes it ideal for measuring moisture conditions within completely sealed containers or dry boxes. It will also perform well under gas flow conditions as high as 10,000 cm/sec and liquid flow conditions to 10 cm/sec.
  • Page 17 The purpose of a sample system is to condition or control a sample stream to within the specifications of a probe. The application requirements determine the design of the sample system. GE Panametrics applications engineers will make recommendations based on the following general guidelines.
  • Page 18 April 2004 Moisture Sample Systems (cont.) Sample Cell Sample Sample Outlet Inlet Vent Figure 1-3: A Typical Moisture Sample System Note: The actual sample system design is dependent on the application requirements. Installation Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 19 April 2004 Oxygen Sample Oxygen sample systems are required and can be ordered from GE Panametrics for bench or wall mounting. You can also build your Systems own sample system by using the following guidelines. IMPORTANT: The GE Panametrics warranty will be voided if the sample system does not have a relief valve.
  • Page 20 April 2004 Oxygen Sample Systems (cont.) Sample Outlet Sample Inlet Figure 1-4: A Typical Oxygen Cell Sample System Note: The actual sample system design is dependent on the application requirements. Installation 1-11 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 21 The sample system is normally fastened to a metal plate that has four System mounting holes. GE Panametrics also provides the sample system in an enclosure if requested. In either case, fasten the sample system plate or enclosure with four bolts–one in each corner. If you ordered sample system outline and dimension drawings, they will be included in your shipment.
  • Page 22 Screw the probe in, making sure not to cross thread it. Tighten down securely. Figure 1-5 below shows a typical moisture probe orientation with the probe mounted in a GE Panametrics sample cell. Install moisture probes with different fittings in the appropriate manner.
  • Page 23 The Series 1 uses any type of 0/4 to 20-mA or 0 to 2-V pressure transducer or transmitter. GE Panametrics offers two types of pressure transmitters: the P40 and P40X. The P40 has a 1/4-inch...
  • Page 24 April 2004 Delta F Oxygen Cell Although the Series 1 accepts other oxygen devices as auxiliary inputs, it is designed to accept oxygen inputs directly from the Delta F Oxygen Cell. There are three steps for installing the Delta F Oxygen Cell: preparing the oxygen cell for operation, calibrating the oxygen cell, and then connecting the cell to the gas line.
  • Page 25 April 2004 Preparing the Oxygen To prepare the oxygen cell for operation, you must fill it with the Cell electrolyte that has been supplied in a plastic bottle. !WARNING! The electrolyte contains potassium hydroxide that is harmful if it comes in contact with eyes or skin. Consult your company safety personnel for proper procedures for handling the electrolyte.
  • Page 26 April 2004 Preparing the Oxygen 4. Replace the top of the oxygen cell. Cell (cont.) Note: Once you add the electrolyte, DO NOT add additional electrolyte to the reservoir. If the level falls below the minimum level, refer to Checking and Replenishing the Electrolyte in the Delta F Oxygen Cell in Chapter 2 of the Service Manual to replenish the cell.
  • Page 27 Use the correct cable type for each probe and make sure the cables are not damaged during installation. If the cable being used is not a GE Panametrics-supplied cable or is a modified cable, see Installing Optional Features in Chapter 1 of the Service Manual.
  • Page 28 Code Article 500 applicable sections which pertain to the hazardous environment classification in which the electronics will be used. Note: The power line is the main disconnect device. However, GE Panametrics does not provide power supply cords with CSA Div. 2 hygrometers.
  • Page 29 April 2004 Connecting Moisture GE Panametrics manufactures a variety of moisture probes for the Series 1. The most commonly used are the M Series, TF Series, and Probes Moisture Image Series probes. If you ordered an M and/or TF Series probe(s), GE Panametrics has entered the necessary probe setup data on a preassigned channel.
  • Page 30 (approximately 1/8 inch of a turn). If you are not using a GE Panametrics-supplied cable, refer to Figure 1-10 below to make proper pin connections to a bayonet-type connector.
  • Page 31 April 2004 M Series Probes (cont.) If you connect a probe to the wrong channel, you can either reconnect the probe to the assigned channel, or reconfigure the current channel as described in Verifying and Entering Setup Data in Chapter 2. NOTICE FOR BASEEFA CERTIFICATION The M Series probe may not be capable of withstanding the 500 V insulation test required by clause 5.7 of...
  • Page 32 April 2004 TF Series Probes Using the special GE Panametrics cable, connect the TF Series Probe to the designated terminal blocks on the back panel of the Series 1. You can locate the TF Series probes up to 600 meters (2000 feet) from the meter if you are measuring moisture and temperature only.
  • Page 33 April 2004 TF Series Probes (cont.) If you do connect a probe to the wrong channel, you can either reconnect the probe to the assigned channel, or reconfigure the channel (where the probe is connected) as described in Verifying and Entering Setup Data in Chapter 2 of this manual.
  • Page 34 April 2004 TF Series Probes (cont.) STD/TF Probe Terminal Block Figure 1-13: STD/TF Probe Terminal Block Location Moisture Image Series Connect the Moisture Image Series probes to the Series 1 using a Probes standard twisted-pair cable with a bayonet-type connector. You can locate the Moisture Image Series Probe up to 915 meters (3000 feet) from the electronics unit.
  • Page 35 April 2004 Moisture Image Series Once the probe is assembled, connect the cable to the terminal block Probes (cont.) labeled on the back panel of the electronics unit (refer to MIS PRB Figure 1-15 and Table 1-4 below). IMPORTANT: DO NOT connect the cable to the Moisture Image Series Probe until you make the proper connections to the back of the meter.
  • Page 36 (approximately 1/8 inch of a turn). If you are not using a GE Panametrics-supplied cable, see Figure 1-16 below to make proper pin connections to a bayonet-type connector. Once you complete connecting the Moisture Image Series Probe(s), you must activate the probe on the installed channel as described in Activating and Changing Probes in Chapter 2.
  • Page 37 (see Figure 1-17 on page 1-29). For proper operation, make connections to each set of electrodes using a four-wire shielded cable. GE Panametrics provides the “O” type cable with 22 AWG leads for the Delta F oxygen cells. IMPORTANT: To maintain good contact at each terminal block and...
  • Page 38 June 2004 Connecting the Delta F Note: Cable with 16 AWG wire is the largest cable size that can be easily installed into the instrument terminal blocks and the Oxygen Cell (cont.) Delta F sensor terminal posts. Instructions for connecting each type of cell are described in the following sections.
  • Page 39 June 2004 Standard Delta F Figure 1-18 below shows the standard oxygen cell and identifies the Oxygen Cells sensing and secondary electrodes. Make oxygen cell connections from the electrodes on the oxygen cell to the OXYGEN terminal block on the back of the electronics unit. Use Table 1-6 below to make oxygen cell connections.
  • Page 40 June 2004 Weatherproof Delta F Figure 1-19 below shows the weatherproof oxygen cell. The oxygen Oxygen Cells cell has a set of sensing and secondary electrodes that are wired to a terminal strip in the weatherproof enclosure. Connect the weatherproof oxygen cell using a four-wire shielded cable with a mating bayonet-type connector.
  • Page 41 June 2004 Explosion-proof Delta F Figure 1-20 below shows the explosion-proof oxygen cell. The Oxygen Cells oxygen cell has a set of sensing and secondary electrodes that are wired to a terminal strip in the explosion-proof enclosure. Connect the explosion-proof oxygen cell using a four-wire shielded cable. Connect one end of the cable to the terminal block on the OXYGEN...
  • Page 42 April 2004 Establishing a Gas Caution! Establish a gas sample flow before you power up or Flow Through the damage may occur to the oxygen cell. Oxygen Cell If you are using an oxygen cell, you must establish a gas flow through the cell before powering up.
  • Page 43 April 2004 Establishing a Gas Slowly open the flow control valve until the recommended flow of 2 to 2.5 SCFH is established on the flow meter. Flow Through the Oxygen Cell (cont.) When the proper flow is achieved, make sure the relief valve is closed by placing an object (e.g., your finger, if the gas is non-corrosive) over the relief valve vent.
  • Page 44 June 2003 Installation 1-35 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 45 June 2003 Installation 1-36 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 46 June 2003 Installation 1-37 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 47 June 2003 Installation 1-38 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 48 June 2003 Partial Cut Away View with I/O Connections To wire the power line to the disconnect device: 1. Make sure the disconnect device is not energized, by placing the switch in the OFF position and cutting the main power source. 2.
  • Page 49 June 2003 Installation 1-40 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 50 June 2003 Installation 1-41 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 51 June 2003 Installation 1-42 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 52 Chapter 2 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 53 Initial Setup Introduction ......... . . 2-1 Using the Front Panel to Access Data .
  • Page 54 June 2003 Introduction The Moisture Image Series 1 is a highly advanced and versatile instrument. Because of this, operation ranges from basic to advanced functions. This chapter is designed to provide a step-by-step guide to getting the meter up and running as quickly as possible. Additional...
  • Page 55 June 2003 Powering Up The Series 1 has a universal power supply that automatically adjusts to line voltages from 90 to 260 VAC. After making electrical connections (including grounding) as described in Chapter 1, Installation, press the power switch to power up the meter. Note: The AC power cord is the main disconnect device.
  • Page 56 (see Figure 2-2 below). Screen Keypad MOISTURE IMAGE SERIES 1 P OWE R Menu Keys Figure 2-2: The Moisture Image Series 1 Front Panel Initial Setup Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 57 June 2003 Screen and Key The screen displays measurements, up to four menu options, system messages, and a pointer. The pointer acts as a measurement Functions indicator, as well as a guide during data entry. While the meter displays measurements in the matrix format, the pointer moves sequentially from box to box to indicate measurement updates.
  • Page 58 June 2003 Screen and Key The keypad to the side of the screen consists of 16 keys, including the [.] and [-] symbols, two arrow keys, and two response keys: [YES] Functions (cont.) and [NO]. The numeric keys are for numeric entry only; however, the arrow and response keys have more than one function.
  • Page 59 Verifying and Entering Before the meter can make measurements, it must have the proper setup data entered into its memory. GE Panametrics has entered all or Setup Data most of the setup data for each probe before delivery; however, you should verify that all the data is correct and entered into your unit.
  • Page 60 April 2004 Activating and Changing Although the probes are physically connected to the back of the electronics unit, you must “tell” the electronics what type of Probes measurements the probe is capable of making. In addition, if you want to use a constant value rather than a live input, you must “tell” the electronics that you are using a constant value.
  • Page 61 April 2004 Verifying Probe 2. Verify that the correct probe types have been activated for each Configuration Data channel. To do this, make sure a probe is activated for each measurement it is capable of taking. For example, if a Moisture (cont.) Image Series Probe with the temperature and pressure functions is connected to channel 1, “MIS”...
  • Page 62 April 2004 Entering or Changing See Figure 2-14 on page 2-26 for a menu map. Probe Configuration 1. From the Probe Configuration Menu (Table 2-1 on page 2-7), Data move the pointer to the channel and probe you want to select. The arrow will only move to installed channels.
  • Page 63 April 2004 Entering or Changing Table 2-2: Probe Types Probe Configuration Data Measurement Mode Probe Type – (None) (Percent) OXY (Oxygen) ppM (Parts per Million) ppB (Parts per Billion) – (None) MIS (MIS Probe) HYG (Hygrometry) TF (Three-Function Probe) Mxx (M Series Probe) Kh (Constant Dew Point) –...
  • Page 64 Delta F Oxygen Cell, page 2-14 Be sure you have the Calibration Data Sheets that are supplied with each GE Panametrics probe. Each Calibration Data Sheet consists of a list of data points that you will need to enter or verify. Each Calibration Data Sheet lists its corresponding probe serial number, as well as the preassigned channel number.
  • Page 65 Calibration Data is not necessary to enter calibration data for the Moisture Image Series Probe unless you send the probe back to GE Panametrics for calibration without its electronics module. If this is the case, enter the calibration data for the recalibrated probe as described in this section.
  • Page 66 April 2004 Entering Moisture Probe 1. At the System Calibration Menu (see Table 2-3 on page 2-11), Calibration Data (cont.) press the PROBE menu key until the Moisture Probe Calibration screen appears. See Figure 2-6. 2. Use the CHANNEL menu key to cycle to the desired channel. The channel number is indicated in the top right-hand corner of the screen.The screen will only display installed channels.
  • Page 67 IMPORTANT: Staple Calibration Data Sheets to the Program Information List (Appendix A) and store them in a safe place. If they are lost, contact GE Panametrics for a duplicate. If you alter any of the calibration data, make sure you make the change on the Calibration Data Sheet.
  • Page 68 April 2004 Entering Delta F Cell µ 5. Move the pointer to the Zero A line and press [YES]. Calibration Data (cont.) µ 6. Enter the microamp ( A) value and press [YES]. 7. Move the pointer to the Zero ppm (or %) line and press [YES]. 8.
  • Page 69 April 2004 Entering High and Low The last step in setup is entering high and low reference values. The Series 1 requires reference values for its measurement circuitry for Reference Values moisture and oxygen inputs. The references are factory calibration values that are specific to each channel card.
  • Page 70 Reference Values inputs. (See Figure 2-14 on page 2-26 for a menu map.) Caution! If you modified the supplied cables or are not using GE Panametrics-supplied cables for moisture probes, you may be required to adjust the moisture reference data as described in Performing an MH Calibration/ Adjustment Test in Chapter 1 of the Service Manual.
  • Page 71 April 2004 Entering Moisture HIGH 1. Move the pointer to for the desired channel, and press Reference Values (cont.) [YES]. 2. Enter the high reference value for that channel and press [YES]. 3. The pointer automatically moves to for the same channel. Press [YES].
  • Page 72 April 2004 Entering the Delta F IMPORTANT: The instructions in this section apply to entering Oxygen Cell Reference reference data for the Delta F oxygen cell only. Values Use the steps below to change the reference values for oxygen cell inputs.
  • Page 73 April 2004 Entering Delta F Oxygen Zero 1. Move the pointer to the line for the high reference for the Cell Reference Values desired channel and press [YES]. (cont.) 2. Enter the zero value for the high reference for that channel and Span press [YES].
  • Page 74 June 2003 Displaying The Series 1 can display measurements in a matrix format or a line graph. The first time you power it up, the screen will display Measurements measurements in the matrix format. The matrix format has six pages and each page consists of six boxes as shown in Figure 2-10.
  • Page 75 June 2003 Displaying Measurements (cont.) Channel Measurement Units Number Mode 5 Pressure 27.10 PSIG Indicators X Unassigned Value Unassigned Figure 2-11: Assigned and Unassigned Boxes The channel number is from 1 to 6, depending on how many channel cards are installed in the unit. Additional channel cards can be purchased separately and installed later (see Installing a Channel Card on Chapter 2, page 2-22 of the Service Manual).
  • Page 76 April 2004 Table 2-5: Measurement Modes and Units for the Series 1 Selected Displayed Displayed Meas. Mode Description of Units Meas. Mode Units %= Percent Oxygen default Oxygen /ppM = Parts Per Million Oxygen /ppB = Parts Per Billion Oxygen Oxygen /µA = Microamps (Diagnostic Mode) Oxygen...
  • Page 77 Your passcode is 2719. Please remove this page and put it in a safe place for future reference. Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 78 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 79 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 80 Chapter 3 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 81 Specifications Overall Specifications ........3-1 Moisture Measurement ........3-2 Pressure Measurement .
  • Page 82 April 2004 Overall Specifications General Motorola 68332-based, one- to six-channel, aluminum oxide, absolute humidity/oxygen analyzer system. Power Universal power supply adjusts automatically from 90 to 260 VAC, 50/60 Hz, 50 Watts maximum power. Note: Power cord is the main disconnect device. Fuses 1/2 Amp, Slo-Blo 3AG.
  • Page 83 Pounds per Million Standard Cubic Feet of Natural Gas or Ideal Gas (#/MMSCF) Vapor Pressure (mmHg) Vapor Pressure (Pascals) Contact GE Panametrics for the availability of other special measuring modes. Specifications Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 84 ±0.5 at -30°C (0.9°F at -22°F) Note: All temperature outputs available in °C or °F. Oxygen Measurement Type Delta F, non-depleting electrochemical cell (optional). Also compatible with other GE Panametrics’ oxygen analyzers such as TMO2, TMO2D, and Series 350. Range 0-500 PPB 0-10,000 PPM...
  • Page 85 April 2004 Electronic Specifications Graphic/text Liquid Crystal Display (LCD) with backlight 256 × 128 Display pixels. Display Functions Display up to six channel/parameter combinations simultaneously using text and/or graphics. Operating Temperature 0°C to 60°C (32°F to 140°F) Storage Temperature -30°C to 70°C (-22°F to 158°F) Memory Retention Moisture, temperature, and pressure calibration data is stored in a 512-KB battery-backed SRAM.
  • Page 86 April 2004 Output Specifications Analog Two per channel Internally optically isolated 12 bit (0.025% Resolution) • Standard Switch- 0 to 2 V, 10K ohm minimum load resistance Selectable Outputs • 0 to 20 mA, 400 ohm maximum series resistance • 4 to 20 mA, 400 ohm maximum series resistance Each output can correspond to any one parameter on that channel.
  • Page 87 Transmitters, Temperature Transmitters, etc. Note: Contact the factory for a higher voltage input. Resolution 16 bits Moisture Sensor Probes GE Panametrics types: M Series, TF Series, and Moisture Image Series. Temperature Sensor Thermistor (optionally supplied as part of the moisture probe assembly).
  • Page 88 April 2004 Probe Specifications Moisture Image Series Probe Type Aluminum oxide moisture sensor probe and electronics module. Calibration Each sensor is individually computer calibrated against known moisture concentrations, traceable to National Institute of Standards and Technology (NIST) or National Physical Lab, U.K. (NPL) as approved by Irish Laboratory Accreditation Board (ILAB).
  • Page 89 April 2004 Moisture Image Series Probe (cont.) Built-in Pressure Sensor Type: (optional) Solid state/piezoresistive Available Ranges: 30 to 300 PSIG 50 to 500 PSIG 100 to 1000 PSIG 300 to 3000 PSIG 500 to 5000 PSIG Accuracy: ±1% of span Specifications Artisan Technology Group - Quality Instrumentation ...
  • Page 90 TF Probe/Analyzer Up to 600 meters (2000 feet) for moisture and temperature Separation (Consult GE Panametrics for distances up to 1200 meters.) Up to 152 meters (500 feet) for pressure Consult GE Panametrics about longer lengths. Probe/Analyzer Cable...
  • Page 91 April 2004 TF Series Probe (cont.) Built-in Temperature Type: Sensor (optional) Thermistor network Operating Range: -30°C to 70°C (-22°F to 158°F) Accuracy: ±0.5°C overall Response Time: 1 second in well-stirred oil 10 seconds in still air for a 63% step change in increasing or decreasing temperatures.
  • Page 92 April 2004 M Series Probe Type Aluminum oxide moisture sensor probe (patented). Impedance Range 50 kΩ to 2 MΩ at 77 Hz (depending on vapor pressure of water). Calibration Each sensor is individually computer-calibrated against known moisture concentrations. Calibrations are traceable to the National Institute of Standards and Technology (NIST).
  • Page 93 Non-depleting electrolytic oxygen sensing cell. Available Cells Range: L: 0 to 500 ppb /5 ppm /50 ppm Ranges for each cell are software selectable in GE Panametrics analyzers. Ranges: A: 0 to 1/10/100 ppm B: 0 to 10/100/1000 ppm C: 0 to 100/1000/10,000 ppm...
  • Page 94 April 2004 Delta F Oxygen Cell (cont.) Sample Requirements Temperature: -18 to 66°C Inlet Pressure: <-0.5 psig (use compressor) -0.5 psig to 0.2 psig (use pump) 0.2 to 1.0 psig (standard range) 1.0 to 60 psig (use valve or regulator) >60 psig (use pressure regulator) Flow Rate: 0.5 to 1.5 liters per minute...
  • Page 95 April 2004 External Pressure Transmitter (optional) General purpose P40X For Class I, Group D, Division 1 locations. Transducer P40 - solid state piezoresistive-silicon sensor in stainless steel housing; on-board zero and span trim. P40X - capacitive-sensor in explosion-proof housing; on-board zero and span trim.
  • Page 96 Appendix A Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 97 Program Information List Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
  • Page 98 Program Information List This list provides a space to enter program data that is stored in the Moisture Image Series 1 memory. If for any reason data is lost, use this list to re-program your unit. Store this sheet and any other related documents in a safe place for future reference.
  • Page 99 June 2003 Reference Values Reference values are printed on a sticker on the side of Series 1 chassis. Record these reference values in the spaces provided below.. (page 2-16) MOISTURE REFERENCE TABLE PRESS HIGH SETTINGS SERVICE REFS PROBE PRESSURE REFERENCE TABLE HIGH OXYGEN CELL REFERENCE TABLE HIGH...
  • Page 100 June 2003 Recorder Setup (page 3-1, Programming RECORDER OUTPUT SETTINGS Manual) RECORDER A RECORDER B Mode: Mode: PRESS Range: Range: SETTINGS Zero: Zero: Span: Span: OUTPUTS Mode: Mode: RECORDER Range: Range: Zero: Zero: Span: Span: Mode: Mode: Range: Range: Zero: Zero: Span: Span:...
  • Page 101 June 2003 User Constants (page 3-9, Programming USER CONSTANT TABLE Manual) PRESS: OPTIONS USER Saturation Constants Compound:________Compound:_______ Compound:________ Channel #:_________ Channel #:________ Channel#:_________ (page 3-9, Programming Temp Temp Temp Manual) PRESS: OPTIONS USER Compound:________Compound:_______ Compound:________ Channel #:_________ Channel #:________ Channel#:_________ Temp Temp Temp...
  • Page 102 April 2004 Index Activating and Changing Probes ..2-7 Gas Flow Arrow Keys Establishing for Oxygen Cell ..1-33 Functions ......2-5 Auxiliary Inputs Activating .
  • Page 103 April 2004 Index (cont.) M Series Probe Output Connecting ......1-20 Specifications ..... . 3-5 Installing into Sample System .
  • Page 104 April 2004 Index (cont.) Safety Verifying Setup Data....2-6 Initial Checkout ....1-2, 1-3 Sample System Connecting to Gas Line for Oxygen .
  • Page 105 Waltham, MA 02453 U.S.A. as the manufacturer, declare under our sole responsibility that the product Moisture Image Series 1 Analyzer to which this document relates, in accordance with the provisions of ATEX Directive 94/9/EC Annex II, meets the following specifications: II (1) G [EEx ia] IIC (-20°C to +50°C)
  • Page 106 Shannon Industrial Estate Shannon, Co. Clare Ireland declare under our sole responsibility that the Moisture Image Series 1 Analyzer Moisture Image Series 2 Analyzer Moisture Monitor Series 3 Analyzer to which this declaration relates, are in conformity with the following standards: •...
  • Page 107 Shannon Industrial Estate Shannon, Co. Clare Ireland déclarons sous notre propre responsabilité que les Moisture Image Series 1 Analyzer Moisture Image Series 2 Analyzer Moisture Monitor Series 3 Analyzer rélatif á cette déclaration, sont en conformité avec les documents suivants: •...
  • Page 108 GE Panametrics Wir, Shannon Industrial Estate Shannon, Co. Clare Ireland erklären, in alleiniger Verantwortung, daß die Produkte Moisture Image Series 1 Analyzer Moisture Image Series 2 Analyzer Moisture Monitor Series 3 Analyzer folgende Normen erfüllen: • EN 50014:1997+A1+A2:1999 • EN 50020:1994 •...
  • Page 109 U S A G E P a n a m e t r i c s 2 2 1 C r e s c e n t S t r e e t , S u i t e 1 W a l t h a m , M A 0 2 4 5 3 - 3 4 9 7 T e l e p h o n e : ( 7 8 1 ) 8 9 9 - 2 7 1 9 T o l l - f r e e : ( 8 0 0 ) 8 3 3 - 9 4 3 8...
  • Page 110 Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT • FAST SHIPPING AND DELIVERY Experienced engineers and technicians on staff Sell your excess, underutilized, and idle used equipment at our full-service, in-house repair center We also offer credit for buy-backs and trade-ins •...