Figure 21: Ovr; Figure 22: Uvr; Figure 23: Heat Off Indicator - GE Lullaby Warmer Maintenance And Service Manual

Table of Contents

Advertisement

Over voltage and undervoltage indications.
Illustration

Figure 21: OVR

Figure 22: UVR

Heat OFF
Indicator

Figure 23: Heat OFF Indicator

Lullaby
Warmer
TM
2042376-001 Rev G ©2012 General Electric Company. All rights reserved. 15
Description
If the applied mains voltage exceeds the
rated voltage by 10% for atleast 1
minute, "OVR" is displayed on the set
temperature display as seen in Figure 21.
The heating shall be cut off when "OVR" is
displayed on the display
WARNING: Do not leave baby unattended
while "ovr" is displayed as heater will be
turned off.
If the applied voltage is lower than the
rated voltage by 10% for atleast 1
minute, "UVR" is displayed on the set
temperature display as seen in Figure 22.
In this condition, temperature monitoring
and all other warmer functions are
present.
Heat OFF indication:
When the hood is rotated to either of the sides
the heat level bar graph turns off and Heat
OFF indicator flashes.
NOTE: In
the Heat OFF condition, audible
alarms are absent.
In this condition, temperature monitoring and
all other warmer functions are present.