Gantry Anchor Location - GE BrightSpeed Elite Preinstallation Manual

Table of Contents

Advertisement

GE H
EALTHCARE
D
5141177-100, R
IRECTION
EVISION
Anchor
11-7/8"
(302.2)
Figure 6-2 Gantry
Anchor Pad
loaded to
910 lbf
Back
Left
Front
Anchor Pad
loaded to
1040 lbf
Figure 6-3 Gantry Anchor Location
14
40.1"
(1018)
6-11/16"
(170)
Anchor
24/13-16"
(630)
RIGHT SIDE
Anchor
Location #2
Redundant
Anchor Location
Cable Entrance
Redundant
Anchor Location
Anchor
Location #1
N ote :
Adjusters are used at each anchor location. Anchor hole ID is 1" (2.5 cm).
Void between adjuster and anchor must be filled according to local building
codes for siesmic application.
Chapter 6 - Floor Loading and Weights
B
S
RIGHT
PEED
81"
(2057)
Anchor
FLOOR
34.75"
(878)
FRONT SIDE
Anchor
Location #3
Redundant
Anchor Location
Redundant
Anchor Location
Anchor
Location #4
E
, E
, E
: P
-I
LITE
DGE
XCEL
RE
NSTALLATION
Anchor
34.75"
(878)
Anchor Pad
loaded to
960 lbf
Right
Anchor Pad
loaded to
1090 lbf
Page 107