Philips M3150B Installation And Service Manual page 318

Information center system
Table of Contents

Advertisement

Troubleshooting
Description
Source
(Event Code)
RecMgr
Error: 2515, One of the waves needed for the recording is no
longer available.
Error: 2516, The requested memory could not be allocated.
Error: 2517, Data from the pds has not arrived yet. Indicates a
long time delay in retrieving data from the pds.
Error: 2518, One or more threads in this process failed to start.
Problem is likely due to a system wide failure and may require
a reboot of the machine.
Error 250C, The recorder corba id key was missing or could
not be read or written to.
Error: 250D, A corba call failed
Error: 250E, Programmer error. Please submit a bug report
with the text from this message.
SDNConfig
The database service has not started. The application must be
terminated
sdprocess
Clinical password
entered
ServiceLog
Winsock connection failed
EntryApp
Could not initialize and/or update the application. The
application was terminated
All devices on the network are unavailable. Application
terminated.
Cannot connect to database. Application terminated.
EMS Service Log Database is corrupt.
Application could not be started.
ServiceLog
Winsock connection failed
ViewerApp
Could not initialize and/or update the application. The
application was terminated
All devices on the network are unavailable. Application
terminated.
Cannot connect to database. Application terminated.
7-32
Maintenance, Troubleshooting, and Repair
Table 7-7. Application Events
Explanation
Clinical password was used in Unit
Settings password window
Corrective Action
If this occurs around a patient
discharge or transfer or when a
bedside monitor module is pulled
out, no action required.
If occurs at other times, reboot system.
Reboot system. If this persists, reload
Philips software.
If this occurs around a patient
discharge or transfer or when a
bedside monitor module is pulled
out, no action required.
If occurs at other times, reboot system.
Reboot system. If this persists, reload
Philips software.
Reinstall Philips software.
Reboot system. If this persists, reinstall
Philips software
Report text in Event Log to Response
Center.
Reboot system.
Reboot system.
Verify proper user access
Reboot system. If this persists, reinstall OS
software
Reinstall Philips software
Check SDN cable and connections, test
SCC
Reboot system. If this persists, reinstall
Philips software
Shut down the Service Log and Contact
the Response Center
Reboot system. If this persists, reinstall
Philips software
Reboot system. If this persists, reinstall OS
software
Reinstall Philips software
Check SDN cable and connections. Test
SCC
If this persists, reinstall Philips software