Philips M3150B Installation And Service Manual page 247

Information center system
Table of Contents

Advertisement

Field
Warning
The height, weight, and gas units of measure configured on the Information Center must match the units of
measure at the bedside. Failure to do so may result in incorrect hemodynamic calculations and patient
documentation.
Multi-lead ECG Layout
Patient Name in
Overview
Allow patient
navigation for Web
Web Overview Port
Warning
If Hemodynamic Pressures and/or Gas Pressures are changed, the entire patient
database for the Server -- Alarms, Waves, Trends, and Parameters -- will be erased.
Local General Config Settings set at the individual Information Center or Client are only
applicable to that particular device.
Field
Patient Type
All Arrhythmia Alarms
Off (for SDN Bedsides
Only)
No Data From Bed
INOP - alerting
Arrhythmia Capability
Note
The Arrhythmia Capability parameter configuration (Enhanced or Basic) must have the same configuration
setting enabled in the Patient Monitor (if arrhythmia is on).
Description
Sets several initial presentation parameters for the display in the 12-lead review applications. The
first set of selections sets the type of layout of ST segment review data - International or
Cabrera. The Layout group selects the layout for the presentation of the 12-lead waveforms, as
either 3 rows by 4 columns, 6 rows by 2 columns, or 12 rows by 1 column. The Speed
group allows for selecting the chart speed used in presenting the waveforms, and can be either 25
mm/sec or 50 mm/sec.
If yes, the patient name for all overviewed beds is displayed in the alarm reflector, overview bed
list, overview window, and alarm overview window. If no, the name is not displayed. This setting
is for M2/M3/M4/IntelliVue patient monitors monitored by the Information Centers connected to
this Database Server. This setting does not apply to SDN beds.
Enabled only if the Patient Data/Transfer parameter was enabled in the Purchased Options
window.
Note—This option must be disabled if automated web access is used.
Port of the TCP/IP socket connection for web access. The default is 9000.
Description
Default setting for the type of patient being monitored -- Adult, Pediatric, Neonate -- for SDN
Bedside monitors. This can be changed for specific patients in the Admit window.
Arrhythmia alarms from telemetry monitors can be turned off in the Arrhythmia Alarms Setup
menu for SDN Bedside monitors. Enabled means they can be turned off. Disabled means
they cannot be turned off.
INOP sound is to be sounded when there is a NoDataFromBed INOP. The NoSound selection
disables the sound alert while the Sound selection will play the alert tone
Sets whether the Enhanced or Basic set of arrhythmia measurements are to be used for SDN
Bedside monitors. Enhanced has 22 arrhythmia alarms. Basic has 10 arrhythmia alarms
Software Installation & Configuration
Configuration
6-21