Network Test

Remote Control Test (Option)

RF LAN Test (option)

Revision A
Maintenance: Checkout Procedures
1. Verify that the monitor is connected to the Unity-MC (Mission
Critical) network.
NOTE
When the monitor is connected to a docking station, verify the
docking station is connected to the Unity-MC network.
NOTE
When the monitor is connected to a docking station, the monitor's
network port is disabled. Only the docking station's network port
is enabled at this time.
2. Select VIEW OTHER PATIENTS.
3. Select SELECT ANOTHER CARE UNIT.
4. Verify that you can see at least one care unit.
5. Select a care unit.
6. Select SELECT A BED TO VIEW.
7. Select a bed.
8. Verify that the patient window appears on the monitor's split-screen.
1. Verify the remote control is connected into an Autoport to Mport
adapter and is inserted into the Aux connector.
2. Select MORE MENUS > MONITOR SETUP > REVISION AND ID.
3. Select NEXT from the popup menu to display the port connectors.
4. Verify the Remote Control label appears after the appropriate port
and the software version for the remote control is 1A.
5. Press each remote control key and verify a beep tone sounds at the
monitor.
1. If the monitor has the Wireless LAN, disconnect the Ethernet cable
and verify Wireless LAN communication still exist between beds,
2. Reconnect the Ethernet Cable.
3. Return to the Main Menu.