Pre-Service Test Procedures; Spo Checks (Pro 300 & 400) - GE Dinamap PRO Series Manual

Hide thumbs Also See for Dinamap PRO Series:
Table of Contents

Advertisement

1.6 PRE-SERVICE TEST PROCEDURES

1.6.1 SpO
14
DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual
off. The Save button must then be selected for this to be a permanent
change.
Cancel
Selection of this button takes the user back to the Service Mode
System Menu. Changes to the Asset ID tag will not be retained if the
Save button has not been selected before leaving the menu.
Save
Selection of this button enters the Asset ID Tag number being created
in eeprom. None of the operations of the Digit, Clear, or Back
buttons are permanent until the Save button is selected. On power-
up, this saved number will be displayed on start-up screen.
The following procedures can be used to check the primary functions
of the PRO Monitor before releasing the unit for clinical service.
The Monitor is tested by applying various stimuli to the sensor
interfaces or by measuring of specific parameters. The test
procedures employ features of the operational software and test
modes of the service menus.
A guide to the Monitor controls, indicators, and connectors is in
Section 2 of the PRO Monitor service manual.
Test Equipment
1. Digital Pressure Gauge 0-375mmHg, 0.2% accuracy, Digitron
P200L or similar.
2. NELLCOR
®
SpO
SpO
or equivalent.
2
3. Biotek SpO
Simulator for Masimo
2
4. NIBP Calibration Kit P/N 320246, or similar.
5. IVAC
®
Probe Simulator TE 1811.
Test Conditions
Testing shall be conducted with an ambient temperature of 25 °C ± 5
°C (77 °F ± 9 °F).
Checks (PRO 300 & 400)
2
For Monitors equipped with Nellcor SpO
On occasion when testing the integrity of the Nellcor oximetry system,
abnormal results may occur when introducing large changes in the
pulse rate and/or pulse amplitude. Extreme changes in rate sent to
the Nellcor sensor by the SpO
algorithm to completely miss finding the pulse rate.
2018554-001
Simulator Model SRC-MAX or similar for Nellcor
2
SpO
®
2
simulator may cause the SpO
2
or equivalent.
2
2
Revision A