Parameter Test Procedures

MANDREL
2 TUBE
BP CUFF
2037106-001 B
Maintenance: Parameter Level Functional Testing
Complete the Test Record at the end of this section as tests are performed.
NOTES
This test is written so that a knowledgeable technician who is familiar with
the V100 Monitor and the test equipment and will be able to follow the test
procedure.
To enter Service Mode press and hold Cycle button while pressing the On/
Off button.
Setup
1.
Connect manometer to unit as shown.
2.
'T' an inflation bulb into the pneumatic setup.
3.
Consult the following diagram for pneumatic setup guidelines.
CARESCAPE V100 Vital Signs Monitor