Appendix A – Electromagnetic Compatibility (EMC): Electromagnetic Compatibility (EMC)
Electromagnetic Compatibility (EMC)
Guidance and Manufacturer's Declaration – Electromagnetic
Emissions
Emissions Test
RF Emissions
EN 55011
RF Emissions
EN 55011
Harmonic Emissions
EN 61000-3-2
Voltage Fluctuations/
Flicker Emissions
EN 61000-3-3
Revision A
Changes or modifications to this system not expressly approved by GE
Medical Systems can cause EMC issues with this or other equipment.
This system is designed and tested to comply with applicable regulation
regarding EMC and must be installed and put into service according to
the EMC information stated in this supplement.
WARNINGS
Use of portable phones or other radio frequency (RF)
emitting equipment near the system may cause
unexpected or adverse operation.
The equipment or system should not be used adjacent to,
or stacked with, other equipment. If adjacent or stacked
use is necessary, the equipment or system should be
tested to verify normal operation in the configuration in
which it is being used.
The Dash 3000/4000 Patient Monitor is intended for use in the
electromagnetic environment specified below. It is the responsibility of
the customer or user to assure that the Dash 3000/4000 Patient Monitor
is used in such an environment.
Compliance
Group 1
The equipment uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
Class A
Class A
The equipment is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Complies