Table 6-2 Equipment Needed - Honeywell UDC2300 Product Manual

Universal digital indicator
Hide thumbs Also See for UDC2300:
Table of Contents

Advertisement

Input Calibration
Equipment Needed
Table 6-2 lists the equipment you will need to calibrate the specific types of inputs that
are listed in the table. You will need a screwdriver to connect these devices to your
indicator.
Type of Input
Thermocouple
Inputs (Ice Bath)
Thermocouple
Inputs (T/C Source)
RTD (Resistance
Thermometer
Device)
Milliampere,
Millivolt, Volts, and
Radiamatic
42

Table 6-2 Equipment Needed

A calibrating device with ± 0.02 % accuracy for use as a signal
source such as a millivolt source.
Thermocouple extension wire that corresponds with the type of
thermocouple that will be used with the indicator input.
Two insulated copper leads for connecting the thermocouple
extension wire from the ice baths to the mV source.
Two containers of crushed ice.
A calibrating device with ± 0.02 % accuracy for use as a signal
source such as a millivolt source.
Thermocouple extension wire that corresponds with the type of
thermocouple that will be used with indicator input.
A decade box, with ± 0.02 % accuracy, capable of providing stepped
resistance values over a minimum range of 0 to 1400 ohms with a
resolution of 0.1 ohm.
Three insulated copper leads of equal length for connecting the
decade box to the indicator.
A calibrating device with ± 0.02 % accuracy for use as a signal
source.
Two insulated copper leads for connecting the calibrator to the
indicator.
Place current source at zero before switching ON.
Do not switch current sources OFF/ON while connected to the
UDC2300 input.
For Radiamatic inputs only, set Emissivity value to 1.0. See Section 4.5 –
Configuration Set Up prompt INPUT1, function prompt EMISS.
UDC2300 Indicator Product Manual
Equipment Needed
10/00

Advertisement

Table of Contents
loading

Table of Contents