Set Date; Set Alarms; Set Cal Gas; Set Cal Due - Honeywell MultiPro Manual

Multi-sensor gas detector
Table of Contents

Advertisement

Click the MODE button once to change the
selection from hours to minutes (or vice
versa).
Hold the MODE button down to advance the
figure that is currently displayed (hours or
minutes).
Once the time has been correctly set, click the
MODE button three times in rapid succession
to return to the Main Menu.
6.4

Set date

The set date submenu is divided into year,
month and day screens. The MultiPro will
automatically scroll through the Year, Month
and Day screens at 5-second intervals.
Click the MODE button once to advance
through the year, month and day screens.
Hold the MODE button down to advance the
year, month or day while the specific setting is
shown.
Once the date has been correctly entered,
click the MODE button three times in rapid
succession to return to the Main Menu.
6.5

Set alarms

Once the "Set Alarms"
submenu has been reached,
the MultiPro will display the
Default Alarms screen and
then begin scrolling through
the various alarm levels.
Click the MODE button once to advance
through the screens.
When the alarm that you want to change is
shown, hold the MODE button down to
advance the counter until the appropriate level
is shown. If the alarm level required is lower
than the current setting, allow the counter to
advance past the maximum possible alarm
level for the specific channel, and continue
back to the lower settings.
To restore the default alarms click the MODE
button three times in rapid succession with
"Default Alarms" displayed. "Def Set" will be
shown to indicate that the default alarm
settings have been restored.
Once the alarm values have been set
correctly, click the MODE button three times
rapidly to return to the Main Menu.
Note: To enable the H
of 1.6 PPM, first enable the H
as described above in section 6.2. The 1.6
PPM alarm value will then appear as an
option in the TWA alarm setting for H
6.6

Set cal gas

Once the set cal gas submenu has been
reached, the MultiPro will scroll through the
various calibration gas values screens and the
set default cal gas option.
Click the MODE button once to advance
through the screens.
Hold the MODE button down to advance the
counter for a specific calibration gas.
To set default calibration gas
levels click the MODE button
three times in rapid
succession with "Default Cal
Gas" displayed.
Once the calibration gas values have been set
correctly, click the MODE button three times in
rapid succession to return to the main menu.
6.7

Set cal due

Once the set cal due
submenu is reached, the cal
due screen will be shown.
Hold the MODE button down
to advance the counter. The number of days
shown is the number of days that must pass
following a successful calibration before the
calibration due warning will be shown. The
maximum length of time between calibrations
is 180 days.
To disable the calibration due warning, set the
interval to 0 days (OFF).
Once the calibration due interval has been set
correctly, click the MODE button three times
rapidly to return to the Main Menu.
Note: Honeywell Analytics recommends
regular verification of accuracy with
calibration gas. See Appendix B for
details.
6.8

Set bump test due

Once the bump due submenu is reached, the
bump due screen will be shown. The number
of days shown is the number of days that
must pass following a successful calibration
before the bump due warning will be shown.
Hold the MODE button down to advance the
counter in days. OFF will be followed by
DAILY, which will then be followed by 2 days
and so on. Once the interval passes 7 days, it
29
S TWA alarm setting
2
S decimal point
2
S.
2