Honeywell MultiPro Manual page 7

Multi-sensor gas detector
Table of Contents

Advertisement

4.
E91-LR6, Eveready EN91 batteries in the alkaline battery pack. Substitution of
batteries may impair intrinsic safety.
5
.
batteries with new batteries and do not mix batteries from different manufacturers.
6.
appropriate MultiPro charger. Standard versions of the MultiPro must be charged
with the UL/CSA-approved charger, which is part number 54-49-103-1. European
versions of the MultiPro must be charged with the ATEX-approved charger, which is
Honeywell Analytics part number 54-49-103-5.
7.
during the charging cycle. Charging the MultiPro in a hazardous location may
impair intrinsic safety.
8.
Panasonic CGA103450 Lithium-Ion batteries. These batteries are not replaceable by
the user. The rechargeable pack must be obtained from Honeywell Analytics and
replaced as an assembly.
9.
with known concentration calibration gas. Failure to check accuracy can lead to
inaccurate and potentially dangerous readings. (The Canadian Standards
Association (CSA) requires an accuracy check using known concentration
calibration gas prior to each day's use.)
10.
atmosphere that is known to contain 20.9% oxygen, 0.0% LEL and 0 PPM toxic gas.
11.
following any known exposure to contaminants by testing with known concentration
test gas before further use. Failure to check accuracy can lead to inaccurate and
potentially dangerous readings.
12.
following any physical shock. Failure to check accuracy can lead to inaccurate and
potentially dangerous readings.
13.
tolerance should be replaced immediately. An instrument that fails calibration may
not be used until testing with known concentration test gas determines that
accuracy has been restored, and the instrument is once again fit for use.
14.
using a calibration gas concentration that differs from the one that is normally
supplied by Honeywell Analytics for use in calibrating the MultiPro.
Customers are strongly urged to use only calibration materials when calibrating the
MultiPro. Use of non-standard calibration gas and/or calibration kit components can
lead to dangerously inaccurate readings and may void the standard Honeywell
Analytics warranty.
15.
components when calibrating the MultiPro can lead to inaccurate and potentially
dangerous readings and may void the standard Honeywell Analytics warranty.
Honeywell Analytics offers calibration kits and long-lasting cylinders of test gas
specifically developed for easy MultiPro calibration. Customers are strongly
Use only Duracell MN1500 or Ultra MX1500, Eveready Energizer
To reduce the risk of explosion, do not mix old or used
Do not charge the MultiPro with any charger other than the
The MultiPro must be located in a non-hazardous location
MultiPro rechargeable battery packs are supplied with
The accuracy of the MultiPro should be checked periodically
Fresh air/zero calibrations may only be performed in an
The accuracy of the MultiPro should be checked immediately
The accuracy of the MultiPro should be checked immediately
A sensor that cannot be calibrated or is found to be out of
Do not reset the calibration gas concentration unless you are
Use of non-standard calibration gas and/or calibration kit