Calibration Due Date (optional)

9. The LCD displays the number of days remaining until the
next calibration is due.
If the Calibration Interval field is set to 0 in Fleet Manager II, it
disables the calibration due date function and this test is
bypassed. Refer to
Calibration Interval
If any sensor is past the calibration due date, the detector
beeps, flashes, and vibrates while the LCD displays the
following screen.
Press C to acknowledge the warning. If C is not
pressed within 2 minutes, the detector automatically
deactivates.
Note
in the User Options.

Force Calibration Enabled (optional)

10. If the Force Calibration option is enabled in Fleet
Manager II and a sensor is past due, the sensor(s) must
be calibrated to continue and enter normal operation.
Press C to acknowledge. Refer to
to begin calibration. If C is not pressed to enter calibra-
tion within 2 minutes, the detector automatically deacti-
vates.
GasAlertMaxXT II
Startup Tests
Calibration Procedure
17