Astart/Bstart (Starting Points Of The Gates); Awidth/Bwidth (Width Of The Gates); Athrsh/Bthrsh (Response And Measurement Threshold Of The Gates) - GE Krautkramer USM 35X Operator's Manual

Table of Contents

Advertisement

Operation
aSTART/bSTART
(Starting points of the gates)
You can fix the starting point of the gates A or B within
the adjustment range of 0 ... 9999 mm/250".
– Select the function aSTART or bSTART.
– Uuse the right-hand rotary knob to adjust the required
setting.

aWIDTH/bWIDTH (Width of the gates)

You can determine the gate width within the range of
0.2 ... 9999 mm/0.008 ... 250".
– Select the function aWIDTH or bWIDTH.
– Use the right-hand rotary knob to adjust the required
value.
5-16
Setting the gates (function groups aGAT and bGAT)
aTHRSH/bTHRSH (Response and mea-
surement threshold of the gates)
You can determine the threshold value of the gates within
the range of 10 to 90 % screen height for triggering the
LED alarm if this value is exceeded or not reached, de-
pending on the setting of the aLOGIC/bLOGIC function.
In the RF mode, the threshold can be additionally set
from –90 % to –10 %.
– Select the function aTHRSH or bTHRSH.
– Set the required value using the right-hand rotary
knob.